Download Moisture Image Series 1 Hygrometer
Transcript
June 2004 Standard Delta F Oxygen Cells Figure 1-18 below shows the standard oxygen cell and identifies the sensing and secondary electrodes. Make oxygen cell connections from the electrodes on the oxygen cell to the OXYGEN terminal block on the back of the electronics unit. Use Table 1-6 below to make oxygen cell connections. IMPORTANT: To maintain good contact at each terminal block and to avoid damaging the pins on the connector, pull the connector straight off (not at an angle), make cable connections while the connector is away from the unit, and push the connector straight on (not at an angle) when the wiring is complete. Table 1-6: Standard Delta F Oxygen Cell Connections Connect: To Delta F Oxygen Cell: To Series 1 OXYGEN Terminal Block red wire + sensing electrode pin 1 green wire – sensing electrode pin 2 white wire + secondary electrode pin 3 black wire – secondary electrode pin 4 shield --- pin 5 Secondary Electrodes Sensing Electrodes Figure 1-18: Standard Delta F Oxygen Cell 1-30 Installation