S traightpoint (UK) Limited World Leaders In Load Monitoring Technology Workshop Manual For: Loadlink plus Miniweigher plus Loadblock plus PCB Snap Offs Loadlink plus, Miniweigher plus 2t & 5t, Loadblock plus 2t & 5t snap the PCB along the GREEN LINES Miniweigher plus 100kg, 250kg, 500kg &1000kg Loadblock plus 100kg, 250kg, 500kg & 1000kg Handheld plus snap the PCB along the RED LINES. Connections Sig+ SigExEx+ Screen +ve to sounder -ve to sounder to pin 1 of binder socket to pin 2 of binder socket +ve to battery -ve to battery Computer Set Up First Select ‘loadlink’ on the rotary switch. Plug the comms cable from the PC into the loadcell. Now double click ‘loadlink’ icon on PC - you will see this screen: Select ‘LLD’ here When the ‘instrument setup’ screen appears select ‘DTR Line Hi’ Comms traffic window Select ‘DTR Line Hi’ Errors window If you have sucessfully connected to the loadcell the ‘comms trafic’ screen will show communication and no errors will be reported in the ‘errors’ window. select the ‘open parameter’ button double click on the model you wish to set up - in this case LLP10T Calibration Now select ‘wizards from the menu and choose ‘user calibration’ and linearisation. Follow the calibration wizard as instructed on screen. Overload Reset Lastly, before disconnecting it is important to reset the ‘overload counter’ to zero before despatch. 1) Click on the ‘+’ sign next to ‘Commands’ 2) Click here to execute the ‘RSTOVRLOD’ (reset overload) command. Shut down the software and disconnect comms cable, the loadcell is fully programmed and calibrated.