Download PROVECTA 70 - Air Techniques
Transcript
SYSTEM ERROR MESSAGES ERROR CODE E25 LLL HHH Greater than 4.00 seconds or 240 pulses Flashing Exposure Time AUDIBLE SIGNAL None None None None None SOURCE OF ANOMALY ACTION REQUIRED The X-ray emission exceeds the programmed time and is terminated by the hardware timer at nominally 5.5 seconds. Test with Tubehead Simulator. Reset power by turning the Main Power Switch OFF (0) and ON (I). If E25 error repeats, replace the X-ray Control Panel. Press the INCREASE () or The AC input line voltage is less DECREASE () buttons to reset the than 106 Volts when the display. Exposure Switch is pressed. If LLL error repeats, check the AC line X-RAY GENERATION IS INHIBITED. voltage at the input terminals to the Xray Control Panel The AC input line voltage is greater than 134 Volts when the Exposure Switch is pressed. Press the INCREASE () or DECREASE () buttons to reset the display. X-RAY GENERATION If HHH error repeats, check the AC line voltage at the input terminals to the X-ray Control Panel IS INHIBITED. The combination of low AC input Press the INCREASE () or line voltage and long selected DECREASE () buttons to reset the exposure time when the display. Exposure Switch is pressed Reduce the selected exposure time X-RAY GENERATION IS INHIBITED. such that the corrected time is less than 4.00 seconds or 240 pulses. The Exposure Switch was released prior to the completion of the programmed X-ray emission time. X-RAY GENERATION WAS PREMATURELY TERMINATED. 34 Press the INCREASE () or DECREASE () buttons to reset the display.