Download HYDRIM M2
Transcript
1. Cycle Description P1 - Regular Wash cycle P2 - Wash Disinfect 80ºC P3 - Wash Disinfect 90ºC Description Use for instruments and cassettes when disinfection is not required Use for instruments and cassettes when disinfection (minimum A0 of 600) is required Use for instruments and cassettes when disinfection (minimum A0 of 3000) is required Cold Prewash <30ºC <30ºC <30ºC Wash Rinse/Disinfect 50ºC 9 minutes 60ºC 50ºC 9 minutes 80ºC for 10 minutes 50ºC 9 minutes 90ºC for 5 minutes Dry* 20 min. standard 30 min. intensive 15 min. standard 20 min. intensive 15 min. standard 20 min. intensive 31 minutes 50 minutes 51 minutes 40 L with or without drying 40 L without drying 70-80L with drying 40 L without drying 70-80L with drying Total Time** without Drying Water Consumption * Cassettes may be of a complex design that does not permit full dryness of the cassette inner surfaces ** Cycle times depend on the temperature of incoming water, mass of the load and mains voltage. Specifications: Height, freestanding: Height, built-in: Width: Depth: Depth with door open: Weight: Running Noise: Hot and cold water connections: Inlet water pressure: Incoming hot water temperature: Drain: Water softener: Drying System: Electrical Connection: Equipment pollution degree: Equipment installation category: Maximum relative humidity: Operating temperature range: Max. Altitude: 850 mm 830 mm 600 mm 600 mm 1200 mm 80 kg 60 dB(A) G ” 2-5 bar max. 60-84°C G ” 1 kg salt capacity Blower 230-240 V 50 Hz 13 A Pollution Degree 2 Installation category II 80% for temp up to 31°C 50% for temp up to 40°C +5°C to 40°C 2000m HYDRIM M2 Service Manual 96-111286 Rev. 1.0 3