Configure Filters’s Parameters
• At the Configuration Screen tap the Filters Button to set the parameters.
• Enter the parameters of the next date to autoclave the sterile water filter (A).
• Enter the parameters of the next date to replace sterile water filter (B).
• Enter the parameters of the next date to replace pre filter (C).
Note
 The sterile water filter must be autoclaved before first use and there after
once per week.
 After it has been autoclaved 10 times or if it has been used for a
maximum of 4 weeks without processing, the sterile water filter /
Pall filter must be replaced.
 The SCOPE-CLEANER S20 will alert the user that the date to autoclave or replace
the sterile water filter or to replace pre filter have expired showing an alert
messenger in the main screen (see Appendix IV: Error and warning messages).