Download Instruction Manual - Cole
Transcript
Model 2655-00 Flame Photometer Installation 2.2 continued Site Conditions WARNING Under no circumstances install the instrument beneath overhanging cupboards. There must be at least 1 metre of clear space above the chimney. For optimum performance, this instrument should be installed in accordance with the following conditions: 1. The environment must be clean and free from dust. 2. The instrument must be placed on a strong, level worktop, free from vibration. The Model 2655-00 requires approximately 500mm x 500mm of bench space, which includes an area in front for solutions and clearance at the rear for fuel and air tubing, with clear access to the mains supply switch. 3. Avoid sites that expose the instrument to direct sunlight or draughts. 4. To meet the specification the ambient temperature must be within the range +10¡C to +35¡C and the maximum relative humidity must not be more than 85%, noncondensing. 5. Make sure that the correct voltage is selected for the local a.c. supply, as shown in the following table. Do not attempt to connect the power supply lead until the voltage is correctly set. VOLTAGE SELECTED 100V 120V 220V 240V VOLTAGE RANGE 90V to 110V 100V to 132V 198V to 229V 230V to 264V