EXERCISE / DIAGNOSTICS TROUBLESHOOTING
Symptom or
Error Displayed
Error 28
Possible Cause
Customer Remedy
Service Remedy
A Reagent Strip that
was detected, as
being placed on the
platform, was not
detected at the first
readhead.
A strip was never placed or
was removed after being
placed: Check your printout
of results, or the Results
Error Report displayed at the
end of the run, to determine
if a result set is missing and,
if so, retest the specimen.
Be sure you do not place
your hand or other objects
on the strip loading station,
as these can be mistaken for
a Reagent Strip. If the error
occurs repeatedly, turn the
instrument power off, wait
several seconds, then turn it
back on to recalibrate the
strip sensor.
If a strip was present,
remove and clean the
Moving Table, Fixed Table,
Holddown (Reference
operating manual section 5,
“Daily Cleaning” and
“Weekly Cleaning”).
Refer to section 7-2-2-1.
Confirm strip detector
operates properly using
instrument test card
(Refer to section 7-3-3-4).
Moving Table is not
installed.
Rev.1199H
Verify that the cable coming
from the strip detector has its
connector properly seated
into the mother PCB.
Replace the strip detector
(Refer to section 9-14).
Check for proper readhead
performance by running
reflectance test with
instrument test card.
(Refer to section 7-3-5-4)
Replace Pre-amp PCB
(Refer to section 9-16-2).
Remove the Fixed Table and
check for the presence of
the Moving Table. If missing
install.
CLINITEK 500 Service Manual Page 7.17