16) Imaging: Video camera option for ACT, it also gives a picture of the
measurement area on the PC monitor. Standard camera monochrome GigE
CCD camera with variable exposure time and gain control ≥ 1080 × 720
pixel, ≥ 12 bits, max ≥ 30 frames per second with 10, 20 and 50 x objectives.
Allowing real time images at variable angle of incidence (< 700).
17) Spectroscopic Measurements: Automatic setting up of optical components
and spectral display in standard energy (cm-1, eV) and wavelength ( , nm)
units for spectroscopic measurements.
18) Power requirement: should be maintained at 230 VAC single phase or 440
V three phase, 50 Hz.
19) Fitting options: Fast regressive fitting approach/algorithm with option
for Fit error, correlation matrix calculation with simultaneous graphical
display of experimental and calculated data.
B) Computers & Software:
1) Computer system (I7, minimum 8 GB RAM, 1 TB hard disk, 24” LED
screen with genuine widows 7 OS) with image analysis software [with pre
defined libraries/models recipes for ease of use with wide material based
on dispersion relations such as SiO2, Si3N4, TiO2, ZrO2, HfO2, Al2O3, ZrO2,
Er2O3, SiOXNY, GeO2, GeOX, etc. 2D materials (Graphene, MoS2, WS2, Boron
nitirde), organic dielectrics, polymers (PDMS, PVP, etc), photoresists, on
Si/GaAs/Ge substrates].
2) The system must have the ability to be controlled remotely via an internal
network along with the facility to connect with internet and to troubleshoot
remotely.
3) Software should also be capable of analysis of ellipsometric, transmission
and reflection data, database
and list of
bibliographic reference
extendable by user, data and graphs easily transferred to Windows
applications, Import/export package functions for high flexibility in file
manipulation, Advanced mathematical fitting algorithms.
C) Surface Plasmon Resonance (SPR) add on of Sensing and Imaging:
1) Angle scan measurement method (sensing) and Intensity interrogation
(imaging) with 2-3 channels for sample flow and array sensing point.
2) LED of wavelength (sensing) 780 nm and wavelength (imaging) 760 nm as
a light source with dove prism for micro fluidic flow cell with 2D CMOS
detector (better preferable).
3) Sensing prism material BK7.
4) Imaging prism material BK & LaF2 or user modified.
5) Sample injection volume in the range of 2-7 µl (better preferable).
6) Data processing via software USB port.
7) Detector 2D Charge Peltier Cooled Coupled Device (CCD).
D) Optional Items for Ellipsometry
1) Electrochemical cell for the study of electrochemical processes using
ellipsometer.
2) Laser Driven Lamp source with high lifetime ≥ 10,000 hours.
3) Non contact sheet resistance tester.
General Requirement :
i) Warranty and maintenance
a. The complete instrument should be under warranty at least for a period of one
year from the date of installation. Additional cost, if any for extended warranty
of three years may be quoted.