Download AAMI EC11 - Electrical, Computer & Biomedical Engineering
Transcript
3.2.8 3.2.9 3.2.10 3.2.11 3.2.12 3.2.13 3.2.14 3.2.15 3.2.16 displacement slope error in lead weighting factors hysteresis after 15-mm deflection from baseline Standardizing voltage:1) nominal value rise time decay time amplitude error Input impedance at 10Hz (each lead) DC current (any input lead) DC current (any other patient electrode) Common mode rejection: allowable noise with 20 V, 60 Hz & ± 300 mV dc & 51-kilohm imbalance System noise: RTI, p-p multichannel crosstalk Baseline control and stability: retun time 10 s after reset return time after lead switch Baseline stability: baseline drift rate RTI total baseline drift RTI (2-min period) Overload protection: No damage from differential voltage, 60-Hz, 1 - V p-p, 10-s application No damage from sumulated defibrillator discharges: overvoltage energy recovery time energy reduction by defibrillator shunting transfer of charge through defibrillator chassis ECG display in presence of pacemaker pulses: amplitude pulse duration rise time frequency Risk current (isolated patient connection) Auxillary output (if provided): no damage from short circuit risk current (isolated patient connection) max mV/s 0.30 max % 5 max mm 0.5 NA max min max mV ms s % 1.0 1 100 ±5 min max megohms µA 2.5 0.1 max µA 1.0 max max mm mV 10 1 max max µV % 30 2 max s 3 max s 1 max µV/s 10 max µV 500 min V 1 NA NA max V J s 5000 360 8 max % 10 max µC 100 range range max max mV ms µs pulses/min 2 to 250 0.1 to 2.02) 100 100 as per applicable document 2.1.1 as per applicable document 2.1.1 © 2000 Association for the Advancement of Medical Instrumentation