3.7. Pre calibration checks Radiation survey meters should be calibrated with a radioactive source. Electronic calibration alone is not acceptable [3]. The following items should be observed before exposing the instrument to a source for adjustment and calibration: the instrument should be free of significant radioactive contamination; the meter should be adjusted to zero or a point specified by the manufacturer using adjustments provided; the batteries or power supply should comply with manufacturers specification for the instrument; the instrument should be turned on and allowed to warm up for the period specified by the manufacturer; electronic adjustments such as high voltage should be set, as applicable, to the manufacturer's specifications; geotropism should be known for orientation of the instrument in the three mutually perpendicular planes, and this effect should be considered during calibration and performance testing; the performance of any internal sampling time base in digital readout instruments should be verified as being within the manufacturer's specification. 3.8. Test instruments The test procedures for radiation survey meters should consider the use of the following instruments: oscilloscope, pulse generator, frequency meter, power supplies (bench and HV power supplies), digital multimeter and counter/timer. 3.9. Test conditions Test conditions should consider the background radiation, temperature and radioactive source employed. 4. TEST CIRCUITS Manufacturer specifications and recommendations must always be followed. If not available some test circuits can be adopted for use in testing different types of survey meters. Refer to standard ANSI/IEEE test procedures [4] or NPL guide 14 (1999) for specific radiation detectors. 4.1. Instruments based in ionization chamber detectors Radiation detectors consist of a chamber filled with air or gas, in which an electric field inside the detector is applied for the collection of charges associated with ions and electrons produced in the measuring volume of the detector by the ionizing radiation. An electrometer is used to measure very small electrical currents (in the range from 10-8 A to 10-15 A) or small electrical charges (in the range from 10-12 C to 10-15 C). Sensitivity in ionization chamber detectors is the ratio between the current produced by an ionization chamber and the exposure rate, given for a radiation source. The isotope employed must be specified. Fig. 2 shows some of the setups for testing ionization chambers [5]. 38