Download - Frank`s Hospital Workshop
Transcript
7 Troubleshooting Error Log Entry Alarm Text Condition (Basic info) Priority Source Enabling Criteria Action/Troubleshooting VENT ADC VREF FAIL Vent Fail. Monitoring Only VSIB ADC3.200V ref voltage out of range <3.179 or >3.221 Vdc High AC, Vent Vent +12.5 V (10 VA) is OK In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from Board Supplies: If “Vent Int Bd 10VA Voltage” reads “OK, and 3.2 Vdc reads “Fail”, replace the VIB. If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd reads “OK”, Check cabling between ACB and VIB. VENT AIRWAY OVERPRESS SIGNAL Inspiration stopped High airway overpressure signal set. Medium AC, Vent Mechanical Ventilation On No Service Action. Reboot system. If problem continues, check Airway Pressure signal in Service Mode. VENT AIRWAY OVERPRESS SIGNAL FAIL Vent Fail. Monitoring Only Ventilator SIB indicates the High Airway overpressure signal was set and Paw < 90 cmH2O and Pmanifold <80 cm H2O. Medium AC, Vent Mechanical Ventilation On No Service Action. Reboot system. If problem continues, check Airway Pressure signal in Service Mode. VENT FLOW VALVE FAIL CURRENT Vent Fail. Monitoring Only Incorrect current feedback for Medium 7 consecutive readings. AC, Vent Reboot System. If problem continues, in the Service Software / Vent Flow & Pressure Diagnosis, increase the Flow Valve counts and view the Flow Valve Current mA and Counts. VENT SUSTAINED PAW SDOWN Vent Fail. Monitoring Only Paw > 100 cmH2O for 10 seconds. Medium AC, Vent In-range Paw data available No Service Action. Reboot system. If problem continues, check Airway Pressure signal in Service Mode. VENT VALVE POWER FAIL Vent Fail. Monitoring Only Nominal 12.5V <11.3 V or >13.13Vdc Medium AC, Vent In the Service Software / Vent Interface Bd Power Diagnosis, view the “Vent Int Bd 10VA Voltage” from Board Supplies: If “Vent Int Bd 10VA Voltage” reads “OK, and the Vent Valve 10VA Volts reads “Fail”, disconnect the VIB to Pan connector harness. If the Vent Valve 10VA Volts continues to read “Fail”, replace the VIB. If “Vent Int Bd 10VA Voltage” reads “Fail” and the “Vent Int Bd 10VA Voltage” from the Anes Cntrl Bd reads “OK”, Check cabling between ACB and VIB. M1046983 05/05 7-67