Download TDA-110P - Air Techniques International

Transcript
OPERATION &
MAINTENANCE
MANUAL
TDA-110P
Aerosol Penetrometer
Air Techniques International
Division of Hamilton Associates, Inc
11403 Cronridge Drive
Owings Mills, MD 21117 USA
TEL 410-363-9696
FAX 410-363-9695
www.atitest.com
TDA-110P
Air Techniques International
Page Intentionally Left Blank
ii
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
TDA-110P
Table of Contents
1.
2.
Utility Requirements....................................................................................................... 1
Specifications ................................................................................................................. 2
2.1.
2.2.
2.3.
2.4.
2.5.
2.6.
2.7.
2.8.
3.
4.
Control Location and Functions ................................................................................... 4
Features and Options................................................................................................... 10
4.1.
4.2.
4.3.
4.4.
4.5.
4.6.
5.
Proper Fill Level ................................................................................................................................... 13
Generator Regulator Pressure ............................................................................................................. 13
Generator Fill ....................................................................................................................................... 13
Machine Startup Sequence.......................................................................................... 14
8.1.
8.2.
8.3.
8.4.
8.5.
8.6.
8.7.
8.8.
9.
Input Air ................................................................................................................................................ 11
Exhaust Connection ............................................................................................................................. 11
Communications Port ........................................................................................................................... 11
User Interface Port ............................................................................................................................... 11
Initial Setup ................................................................................................................... 12
Filling Methods ............................................................................................................. 13
7.1.
7.2.
7.3.
8.
Test Types ........................................................................................................................................... 10
Alarm Parameters ................................................................................................................................ 10
Alarm (audible) ..................................................................................................................................... 10
Usage Timer......................................................................................................................................... 10
Data Output .......................................................................................................................................... 10
User Interface Port ............................................................................................................................... 10
Installation .................................................................................................................... 11
5.1.
5.2.
5.3.
5.4.
6.
7.
Physical Characteristics ......................................................................................................................... 2
Operational Requirements ..................................................................................................................... 2
Aerosol Generation ................................................................................................................................ 2
Aerosol Detection ................................................................................................................................... 2
Filter Test Flow Measurement ............................................................................................................... 2
Pressure Measurement .......................................................................................................................... 2
Efficiency Measurement ......................................................................................................................... 2
Communications Port ............................................................................................................................. 3
Power On ............................................................................................................................................. 14
System Initialization ............................................................................................................................. 14
System Version .................................................................................................................................... 14
Input Air ................................................................................................................................................ 14
System Vacuum ................................................................................................................................... 14
Check Liquid Level ............................................................................................................................... 14
System Warm up ................................................................................................................................. 14
Initial Calibration .................................................................................................................................. 14
Test Menu...................................................................................................................... 15
9.1.
Start Cycle............................................................................................................................................ 15
9.2.
Test Loading ........................................................................................................................................ 15
9.3.
Testing ................................................................................................................................................. 15
9.4.
Test End ............................................................................................................................................... 15
9.5.
Failed Check ........................................................................................................................................ 15
9.6.
Calibration Prompt ............................................................................................................................... 15
9.7. .................................................................................................................................................................... 15
10. Standby Mode ............................................................................................................... 16
11. Calibration Sequence ................................................................................................... 17
11.1.
11.2.
11.3.
Test Flow Setup Menu ......................................................................................................................... 17
Test Flow Setting ................................................................................................................................. 17
Penetration Setup Menu ...................................................................................................................... 17
12. Setup Menu ................................................................................................................... 19
12.1.
12.2.
12.3.
Select Pressure Units .......................................................................................................................... 19
Maximum Percent Penetration Variable (Alarm Point) ........................................................................ 19
Minimum Percent Penetration Variable (Alarm Point) ......................................................................... 19
iii
TDA-110P
12.4.
12.5.
12.6.
12.7.
12.8.
12.9.
12.10.
12.11.
12.12.
12.13.
Air Techniques International
1800103 Rev C
Maximum Resistance Variable (Alarm point) ....................................................................................... 19
Minimum Resistance Variable (Alarm point) ........................................................................................ 20
Test Flow Tolerance Variable (Alarm Point) ........................................................................................ 20
Load Time Variable .............................................................................................................................. 20
Extended Load Time Variable.............................................................................................................. 20
Sample Time Variable .......................................................................................................................... 20
Open Time Variable ......................................................................................................................... 21
Cycles to Calibration ........................................................................................................................ 21
Concentration Reference Alarm ...................................................................................................... 21
Set Date and Time ........................................................................................................................... 21
13. Maintenance.................................................................................................................. 22
13.1.
13.2.
13.3.
13.4.
Initial Setup .......................................................................................................................................... 22
Definitions and Features ...................................................................................................................... 22
Scheduled Maintenance ...................................................................................................................... 22
Spare Parts List ................................................................................................................................... 22
14. Maintenance Menu Structure ...................................................................................... 23
14.1.
14.2.
14.3.
14.4.
14.5.
14.6.
14.7.
14.8.
14.9.
14.10.
14.11.
14.12.
14.13.
14.14.
Maintenance Menus ............................................................................................................................. 23
Gravimetric ........................................................................................................................................... 25
Res Offset ............................................................................................................................................ 26
Flow Balance........................................................................................................................................ 27
LSC Check ........................................................................................................................................... 27
LSC Offset............................................................................................................................................ 28
View Sensor ......................................................................................................................................... 29
Cal Sensor ........................................................................................................................................... 31
Primary (HIGH) Flow Sensor Setup ..................................................................................................... 33
Auxiliary (LOW) Flow Sensor Setup ................................................................................................ 35
ID ..................................................................................................................................................... 37
Console ............................................................................................................................................ 37
Security ............................................................................................................................................ 37
Setting the Machine ID .................................................................................................................... 38
15. Console Commands ..................................................................................................... 39
16. LSC Maintenance ......................................................................................................... 40
16.1.
16.2.
16.3.
Photomultiplier Tube Replacement ...................................................................................................... 40
Cleaning ............................................................................................................................................... 40
Aligning ................................................................................................................................................ 41
Appendix A.
Appendix B.
Appendix C.
Appendix D.
Appendix E.
Appendix F.
Appendix G.
Appendix H.
Appendix I.
Appendix J.
Appendix K.
Dip Switch Settings .................................................................................... 42
TDA-110P Operating Parameters .............................................................. 43
Default Settings .......................................................................................... 44
Fault and Error Messages .......................................................................... 45
Remote Mode Operation ............................................................................ 47
USER INTERFACE Circuit Example .......................................................... 48
4B Lite Aerosol Generator ......................................................................... 49
PAO-4 Safety Data Sheet ........................................................................... 50
Blower Motor............................................................................................... 56
Flow Sensor Operation and Service Manual ............................................ 57
Manual Revision ......................................................................................... 58
iv
1800103 Rev C
Air Techniques Incorporated
Page Intentionally Left Blank
v
TDA-110P
1800103 Rev C
Air Techniques Incorporated
TDA-110P
1. Utility Requirements
Control Unit

2
Compressed Air - Clean, dry, compressed airflow rate of 10 SCFM at 80 psig (283 slpm @ 5.6 kg/cm )
NOTE: Maintain the compressed air temperature between 60 and 90 degrees Fahrenheit (70 to 75 degrees is
recommended). It is highly recommended that a coalescing filter and pressure regulator be installed between the
plant airline and the air inlet.

Power –
Control Unit - Stable electrical power at 115 VAC, 50/60 Hz (a power line conditioner and surge
protection is recommended). 230 VAC, 50/60 Hz is optional.
NOTE: A dedicated power circuit is recommended for stable machine operation. Consult the manufacturer.

Ambient Conditions - Temperature controlled environment: 68 to 85 degrees Fahrenheit
recommended. Humidity controlled environment: 45 – 75% relative humidity recommended.
Aerosol Unit

2
Compressed Air - Clean, dry, compressed airflow rate of 10 SCFM at 80 psig (283 slpm @ 5.6 kg/cm ).
Three separate lines required.
NOTE: Maintain the compressed air temperature between 60 and 90 degrees Fahrenheit (70 to 75 degrees is
recommended). It is highly recommended that a coalescing filter and pressure regulator be installed between the
plant airline and the air inlet.

Blower Motor Power - Stable electrical power at 208-230/460 VAC, 3-Phase, 60 Hz.
NOTE: A dedicated power line is recommended for stable machine operation. Consult the manufacturer.

Ventilation - Exhaust the air with a duct capable or directing 500 cubic feet per minute at a resistance of
≤1 inch of water (see Section 5.2 - Exhaust Connection for details).

Ambient Conditions - Temperature controlled environment: 68 to 85 degrees Fahrenheit
recommended. Humidity controlled environment: 45 – 75% relative humidity recommended.
NOTE: High ambient temperatures may create particle instability.
1
TDA-110P
Air Techniques International
1800103 Rev C
2. Specifications
2.1. Physical Characteristics
Control Unit Size
Weight
24 in wide x 30 in deep x 72 in high
210 lbs
Aerosol Unit Size
Weight
120 in wide x 44 deep in x 54 in high
750 lbs
System Blower Size
Weight
22 in wide x 28 in deep x 17 in high
150 lbs
2.2. Operational Requirements
Control Unit Power
Compressed Air Input
System Blower Power
System Exhaust
115 VAC, 60 Hz, 2.5A
220 VAC, 50 HZ, 1.25A (Optional)
2
10 SCFM @ 80 psig (283 slpm @ 5.6 kg/cm )
208-230 VAC-3 Phase, 50/60 Hz or 460 VAC – 3 Phase, 50/60 Hz
Aerosol is exhausted from the system, while not operating in the Test Mode, at a rate
of approximately 400 cfm (cubic feet per minute)
2.3. Aerosol Generation
Technique
Compressed Air Input
Reagent
Diameter
Geo Std Deviation
Concentration
Pneumatic nebulization with impaction using “Cold” 4 centistokes PAO
2
10 SCFM @ 80 psig (283 slpm @ 5.6 kg/cm ), 2 lines required
PAO-4 (poly alpha olefin – 4 centistokes) (See Appendix I)
0.185 ± .020 micrometer CMD (meets 42 CFR Part 84)
< 1.6 (meets 42 CFR Part 84)
3
20 milligrams per cubic meter (mg/m ) minimum.
2.4. Aerosol Detection
Technique
Dynamic Range
Accuracy
Sample Flow Rate
Near forward light scattering
3
0.0001 to 200 milligrams per cubic meter (mg/m )
± 1%
Partial flow through detector, 45 LPM (liters per minute)
2.5. Filter Test Flow Measurement
Technique
Accuracy
Range
NIST traceable flow meter
± 1.0%
10 to 330 cfm (cubic feet per minute) @ 21.1°C, 760 mm Hg
2.6. Pressure Measurement
Technique
Accuracy
Range
Electronic pressure transducer
± 1.0% of full scale
0 to 100 mm water column (0 to 3.94 inches water column)
2.7. Efficiency Measurement
Media Test Flow
Operating Range
10 – 330 cfm (cubic feet per minute) @ 21.1°C, 760 mm Hg
Efficiencies to 99.9995%
2
1800103 Rev C
Air Techniques Incorporated
2.8. Communications Port
Serial Port
Data Bits
Baud Rate
Parity
Stop Bit
Flow Control
Cable required
RS-232
8
9600
None
1
Xon/Xoff
Null modem configuration (shown below)
3
TDA-110P
TDA-110P
Air Techniques International
1800103 Rev C
3. Control Location and Functions
Before operating this unit, familiarize yourself with its features and functions. Each number in the photo is keyed
to the list below.
8 READY Indicator light
11 % Penetration window
12 Resistance window
9 FAULT Indicator light
13 Test Flow window
10 FAIL Indicator light
7 VF Display window
6 Function keys
5 Chuck Close Switches
4
1800103 Rev C
Air Techniques Incorporated
TDA-110P
17 Exhaust Damper meter
1 Power ON
2 Power indicator
3 Power OFF
4 Emergency Stop
15 Test Flow Rate meter
16 Exhaust Damper adjustment
14 Test Flow Rate adjustment
19 Aerosol Generator ON/OFF
Switch
Pneumatic Tubing Connections-Enclosure Lower Right-side
(39) Generator Pilot
(34) Chuck Pilot
(38) Damper Close
(35) Aerosol UP Stream
(32) Resistance
UP Stream
(31) Chuck Supply
(37) Damper Open
(36) Aerosol DWN Stream
(33) Resistance
DWN Stream
5
TDA-110P
Air Techniques International
1800103 Rev C
20 AC Power Input
28 RS232 Communication Port
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
29 User Interface port
Power ON – Green push button turns system power and air on
Power Indicator – Indicates that the power is on when lit
Power OFF – Red push button turns system power and air off
Emergency Stop – Push to turn off system power and air. Rotate clockwise to re-set.
Chuck Close Switches – Used to actuate test chuck movement. Press simultaneously to close the test
chuck. Pressing either switch will stop the function being performed and open the Test Chuck
Function Keys – Four function keys, labeled A through D, used to select or change options. The display
window shows the function of each key.
VF Display Window – Displays system information and function key options
Ready Indicator – Green light indicates the unit is ready to start a new test
Fault Indicator – Yellow light indicates a pre-calibration condition or system fault (see Table 3 – System
Faults)
Fail Indicator – Red light indicates a failed test parameter
% Penetration Window – Displays the % penetration readings during testing and displays set points
when in Setup mode
Resistance Window – Displays the resistance (pressure drop) during testing in units of inches H 2O
Flow Rate Window – Displays the flow rate during testing in cfm
Test Flow Damper Adjustment – pneumatic control of flow control damper. Used in conjunction with
the exhaust damper for flow rate adjustment
Test Flow Damper Meter – displays the position of the flow control damper in 0-100 % open.
Exhaust Damper Adjustment – controls opening and closing of the exhaust damper. Used in with the
test flow damper for flow rate adjustment
6
1800103 Rev C
Air Techniques Incorporated
TDA-110P
17. Exhaust Damper % Open Meter – displays position of exhaust damper in 0 -100 % open
19. Aerosol Generator ON/OFF switch – Turns on the aerosol generator pilot air pressure to enable the
aerosol generators.
20. AC Power Input – Recessed 3-prong, AC power inlet connector for control unit
21. Generator compressed air supply – Regulated compressed air supply for system aerosol generator
30 Main compressed air inlet port
24 Main Power Relay
22 Electronic Enclosure
25 AC line Filter
23 Main PCBA
7
TDA-110P
Air Techniques International
1800103 Rev C
26 Upstream Pressure
Transducer port - P2
27 Downstream
Pressure transducer
port - P1
22. Electronics Enclosure – contains the control relays and PCBA (printed circuit board assemblies)
23. Main PCBA – microprocessor circuit board. Location of operating program E-Prom and 8 bit
microprocessor
24. Main Power Relay (CR1) – maintains power to the control unit circuits, relays and valves after actuation
by the main Power ON switch
25. AC Line Filter –power input line conditioner to remove transient voltage fluctuations
26. Upstream Pressure Transducer Port P2 - Used to determine the filter pressure differential
27. Downstream Pressure Transducer Port P1 - Used to determine the filter pressure differential
28. Communication Port – A DB-9 female connector for RS-232 computer interface used for connection to
a PC serial port
29. User Interface Port - Used to interface with external control equipment
30. Main Pressure Input Port - Input for system compressed air supply
31. Chuck Supply– Compressed air used to raise test chuck to its open position
32. Resistance UP Stream - Used to determine the filter pressure differential
33. Resistance DWN Stream - Used to determine the filter pressure differential
34. Chuck Pilot - Compressed air used to lower test chuck to its closed position
35. Aerosol UP Stream – Sample line for measuring 100% aerosol challenge used for testing
36. Aerosol DWN Stream – Sample line for measuring aerosol challenge level penetrating through the filter
under test
8
1800103 Rev C
Air Techniques Incorporated
TDA-110P
21 Generator Supply
41 Virtual Impactor
40 Aerosol Generator
Damper Open – Opens the aerosol damper.
Damper Close – Closes the aerosol damper.
Generator Pilot – Provides air pilot signal to turn on aerosol generators.
Aerosol Generator - Cold nebulization aerosol generator. Two units included.
Virtual Impactor – Impactor to control aerosol particle size output by removing large particles. Two units
included.
42. Flow Sensor Control Unit – Contains the control electronic and adjustments for the flow sensor signal to
the control unit.
43. Flow Sensor – Measures the air speed for the flow measurement.
37.
38.
39.
40.
41.
43 Flow Sensor
42 Flow Sensor Control Unit
9
TDA-110P
Air Techniques International
4. Features and Options
4.1. Test Types



Percent Penetration
Resistance (available units of measure are mm
water column, inches of water, Pascals, or
kilopascals)
Test Flow Rate (CFM)
4.2. Alarm Parameters
Alarm points can be set and enabled by the user (see
Setup Menu beginning at Section 12.5 for details). If
an enabled alarm point is reached, the test fails, the
display flashes, and the audible alarm sounds.
4.3. Alarm (audible)
The alarm is used for prompting the operator and
indicating system actions and faults. The time
interval during which the audible alarm functions is
set by adjusting the “Open Time” variable.
4.4. Usage Timer
When the unit is first turned on, the penetration
display shows the total time, in hours, that the unit
has been in an energized state.
4.5. Data Output
Outputs test data through the RS-232
communications port. Port settings are 9600, N, 8, 1
4.6. User Interface Port
The port allows the user to interface with ancillary
equipment for controlling or monitoring the unit. (See
Section 5.4 – User Interface Port & Appendix F –
Remote Mode Operation)
10
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
Table 1 & Appendix G & H for connection
details).
5. Installation
5.1. Input Air
Ready (Output) – Indicates that the unit is ready to
start a new test
Before connecting the compressed air line,
check the line for any fluids or unwanted
contaminants.
Pass (Output) – Indicates a passed test
NOTE: Pre-filtering is recommended.
Connect the compressed air sample line to the
Main Pressure Input Port shown in the Figure
on page 7, Item No. 30 and the two 4Blite
generators.
Fail (Output) – Indicates a failed test
Fault (Output) – Indicates a system fault (See
Appendix D –Fault & Error Messages)
In Cal (Output) – Indicates when the unit is in
penetration calibration
CAUTION: For the user’s safety, make sure all the
connections are secure.
External Start (Input) – The signal used to start a
test cycle from an external source. A pulse of 524V DC for a minimum 0.100 seconds will start a
test cycle.
5.2. Exhaust Connection
Connect each of the two 6–inch exhaust hoses
(user supplied) to the system exhaust ducts.
Exhaust draw should be as minimal as
possible.
NOTE: This port is isolated from the unit to protect
against external damage.
WARNING: Excessive exhaust draw will cause
unstable aerosol concentration resulting in
faulty and/or inconsistent results.
Table 1 – Connector Pins & Functions
Connector Pin
1
2
3
4
5
6
7
8
9
5.3. Communications Port
The communications port is a standard 9-pin
male (DTE configuration).
Using a standard RS-232 null modem cable,
connect the unit to your PC serial port.
Configure your serial port to the following
settings:
Baud
Data Bits
Parity
Stop Bit
Flow Control
TDA-110P
9600
8
None
1
Xon/Xoff
5.4. User Interface Port
The outputs are an open collector type capable
of sinking 50 mA for a non-inductive load (see
11
Function
External Start
NC
Fault
Pass
NC
Ground
Ready
Fail
In Cal / Busy
TDA-110P
Air Techniques International
1800103 Rev C
6. Initial Setup
1) Verify that all the input compressed air supplies are connected.
2) Plug the unit into the proper power source.
3) Fill both the 4Blite generators to the fill line (middle) with the oil specified in the
Factory Settings sheet (see Section 7 - Filling Methods).
NOTE: Over or under filling the generator will result in an unstable aerosol
source.
WARNING: If any liquids are used other than that which is specified on the
Factory Settings sheet, the generator will not function to specification.
4) Follow the Machine Startup Sequence (Section 8) until the Test menu (Section 9) is
reached. At this point the TDA-4Blite generators should be on and each of the
regulators located at the generators adjusted to 30 psi.
5) Check the system settings against the Factory Settings sheet included with this
manual.
6) Power the unit off by pressing the red push button.
7) Perform the test set up at this time (Section 12 – Setup Menu) to adjust user
specific testing parameters.
12
1800103 Rev C
Air Techniques Incorporated
7. Filling Methods
7.1. Proper Fill Level
PAO Moil must always be used with this equipment.
ATI 4Blite Laskin aerosol generators tanks are
marked with a fill line at the center point of the
reservoir. The PAO oil level should always be
maintained within ± 0.25 inches of the fill line.
WARNING: Do not fill generator more than 0.50
inches from the fill line.
7.2. Generator Regulator Pressure
The 4B lite generator regulators must be set at 30 psi
when the regulators are functional. Switch the
Aerosol Generators switch to ON at the control panel
and check to ensure the regulators are set to 30 psi.
7.3. Generator Fill
Remove the generator fill caps from each generator
and fill with PAO oil noting the fill line on the sight
glass. Replace the fill cap.
WARNING: NEVER operate the equipment
without the fill caps in place.
.
13
TDA-110P
TDA-110P
Air Techniques International
1800103 Rev C
8. Machine Startup Sequence
In this manual, the following conventions are
used:
8.4. Input Air
Awaiting Air
<Function> indicates the function key that is
pressed. This usually corresponds to
the switch label on the bottom line of
the display. If no label is present, the
switches are designated “A”, “B”, “C”,
and “D”, from left to right.
xx
indicates a variable displayed value.
[P]
Indicates the penetration display.
[R]
Indicates the resistance display.
[F]
Indicates the test flow display.
<CLOSE>
Indicates that both of the chuck close
switches should be pressed
simultaneously.
<OPEN>
Indicates that pressing either chuck
close switch should open the chuck.
The unit is waiting for the input air.
8.5. System Vacuum
Awaiting Vacuum
The unit is waiting for the system vacuum.
8.6. Check Liquid Level
Check the liquid level in generators and fill if
necessary. Remove any liquid in the drip jars
before continuing.
NOTE: Pressing either chuck close switch cancels
the function being performed and opens the chuck.
Check Liquid Level
CONT
8.1. Power On
Turn on the unit power by pressing the green button
on the control unit then turn on the blower and
aerosol generator using the switches.
<CONT> Continue with Startup
8.7. System Warm up
The system will now enter an eight minute
warm-up to allow the aerosol output to stabilize.
The standard warm-up period is eight minutes.
8.2. System Initialization
Immediately after power on, the system will
enter an initialization sequence lasting
approximately 15 seconds. The progress of the
initialization is indicated by a series of periods on
the second line of the display.
SYSTEM WARMUP
BYPASS
<BYPASS> Continue with Startup
Initializing System
.........
8.8. Initial Calibration
After a system start the system must perform
the penetration calibration sequence (Section 11).
After this initial calibration is completed, the
machine will be ready for use and will display the
screen shown in Section 9.
8.3. System Version
ATI 110P
Ver
n.n
UNIT ID 9999
n.n
9999
[P]
Software Version
The unit id assigned to this unit. (See
Section 15.21 for information on
setting the unit id.)
Hours the unit has been in operation
14
1800103 Rev C
Air Techniques Incorporated
TDA-110P
9. Test Menu
After successful initialization, the TDA-110P will
be ready to test samples.
indicated by the operator panel status lights. The
displayed values are:
[P]
[R]
[F]
9.1. Start Cycle
At this point the machine is idle and can either
initiate a test, setup operating parameters (limits
and timings), calibrate the LSC, or enter a standby
state.
If all three of the test parameters are within the
limits specified for the test the sample passes and
the machine will return to the Start Cycle menu.
START CYCLE
STBY
<SETUP>
<CAL>
<STBY>
<CLOSE>
CAL
Test % Penetration reading
Test Resistance reading
Test Flow reading
If any of the test parameters are outside the
specified limits, the sample will be failed, the red fail
indicator will be illuminated, the display for the
failing parameter(s) will flash, and system will
display the Failed Check screen (See Section 9.5) if
the Failed Check option is enabled.
SETUP
Enters the Setup Menu (Section 12).
Enters the Calibration Sequence.
(Section 11)
Enters into Standby Mode
Close chuck to begin testing
Results of a test will remain on the displays until
the next operation is started.
9.2. Test Loading
9.5. Failed Check
The system is waiting the designated load time
prior to testing.
The Failed Check is used to confirm a failed
test. This check is optional and may be bypassed if
the Failed Lockout dipswitch is set or if the unit is in
the Automatic Testing mode. (See Appendix A for
setting locations.)
** LOADING **
9.3. Testing
** FAILED **
Test readings are displayed for the total sample
time in approximately one-second intervals. The
chuck will automatically open at the end of the test
cycle if the sample time has been set.
CONT
<CONT>
** TESTING **
[P]
[R]
[F]
Returns to Start Cycle
9.6. Calibration Prompt
This prompt will appear if the “Cycles to
Calibration” setup option has been selected
(Section 12.14) and the specified number of test
cycles has been reached.
Test % Penetration reading
Test Resistance reading
Test Flow reading
Cal Penetration?
9.3.1.1.1.
NOTE: If the
sample time is disabled,
the test continues until
either chuck switch is
pressed to end the test.
YES
<YES>
<NO>
9.4. Test End
At the completion of the test the last sample
parameter readings will be displayed on the
operator panel and the success or failure of the test
9.7.
15
NO
Enter the penetration calibration
process (See Section 11.7)
Skip the calibration and return to the
Start Cycle menu. If the calibration is
skipped, the prompt will not appear
until the number of cycles is reached
again.
TDA-110P
Air Techniques International
10. Standby Mode
The Standby Mode allows the user to turn off the
aerosol generator when the system is idle. If desired
the blower may be disabled using the switch on the
control panel.
STANDBY
CONT
<CONT>
Reactivates the aerosol generator. You
must also ensure that the blower and
aerosol generator switches are on. If
the machine has been in standby mode
for more than 10 minutes the System
Warmup screen (see Section 8.7) will
be displayed. If the system has been in
standby for 10 minutes or less, the
following screen (“Wait”) will be
displayed.
WAIT
This screen indicates that the system is waiting for
the aerosol generator to stabilize. The screen will be
displayed for approximately 30 seconds after which
the system will return to the system start screen (see
Section 9) and will be ready for further testing
16
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
TDA-110P
11. Calibration Sequence
NOTE: Any time the chuck is closed, pressing
either chuck switch will abort the function being
performed.
11.1. Test Flow Setup Menu
This menu is used to adjust the filter test flow.
This screen will only be displayed if the flow has
already been set. If the machine has just been
started, this screen will be bypassed and you will
continue with the screen in section 11.6.
* CALIBRATING *
* Loading *
SET FILTER TEST FLOW
This screen is displayed while loading chuck
with aerosol.
YES
NO
<Yes>
<No>
Enters into Test Flow Setting menu
Skips to Penetration Setup menu
* CALIBRATING *
11.2. Test Flow Setting
100% Setting
Insert the test filter into the chuck.
INSERT FILTER;
This screen is displayed while measuring the
100% concentration level.
CLOSE
EXIT
<CLOSE>
<EXIT>
11.3.2. Display Concentration Level
Close chuck to continue
Returns to Test Flow Setup (Section
11.5)
This value is only displayed if the Concentration
Reference Alarm option is set in the Setup menu.
NOTE: Any time the chuck is closed; pressing
either chuck switch will abort the function being
performed and open the chuck.
Adjust the Exhaust Damper and Test Flow
valve controls until the desired flow is reached.
Conc = nnn.nnnn
CONT
<CONT>
Flow = 9.9999
11.3.3. Resistance Tare Setting
The unit establishes the resistance tare value.
(This is the inherent resistance of the air path with
no filter in place.)
ENTER
9.9999
Test Flow (cfm)
NOTE: Allow the flow reading to stabilize for 10-15
seconds before accepting the setting.
<ENTER> Accepts flow setting, opens chuck, and
continues to Penetration Setup menu
Clear Chuck & Close
11.3. Penetration Setup Menu
<CLOSE>
This procedure establishes the response of the
LSC to zero and 100% levels of aerosol.
Close chuck to proceed
NOTE: Any time the chuck is closed, pressing
either chuck switch will abort the function being
performed.
11.3.1. 100% Setting
The unit establishes the 100% value.
Insert Filter
and Close
<CLOSE>
Continues with the calibration process.
Close chuck to proceed
17
TDA-110P
Air Techniques International
11.3.4. Zero Setting
* CALIBRATING *
0 SETTING
This screen is displayed while the system
establishes the zero reference.
11.3.5. Complete Calibration
WAIT
This screen is displayed while the system
completes the calibration process. The system will
then return to the screen from which the calibration
was initiated.
18
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
TDA-110P
12. Setup Menu
The Setup Menu is a series of screens that are
used to establish the operating parameters of the
machine.
<UP>
<DWN>
<DSBL>
General Notes:
 A number of the parameters are numeric.
These parameters are entered by
incrementing or decrementing a starting
value. If the increment or decrement buttons
are held down, the rate of the change
increases and will continue to increase as
long as the button is held. To return to the
slower rate of increment or decrement,
release the button and press it again.

<ENBL>
<ENTER>
Some parameters, such as alarm limits, may
be disabled. If a limit is disabled, no alarm
will be raised if the parameter exceeds the
specified limits. If it is possible to disable an
alarm or function an additional button,
ENBL/DSBL (Enable/Disable), is available.
When the button is labeled ENBL, its function
is to enable the parameter. When the button
is labeled DSBL, its function is to disable the
parameter. Disabled parameters are shown
as all dashes (“-“).
12.3. Minimum Percent Penetration
Variable (Alarm Point)
Pen Min = 999.9999 %
UP
999.9999
12.1. Select Pressure Units
<UP>
<DWN>
<DSBL>
The TDA-110P can display pressure in one of
four units: inches of H2O (inH2O), millimeters of H2O
(mmH2O), Pa, and kPa. Use this choice to select the
desired unit.
Pres. Unit now
mmH20
NEXT
<NEXT>
<CONT>
<ENBL>
CONT
<ENTER>
Selects the next pressure unit. The top
line will change to reflect the new unit.
Saves the current unit choice and
proceeds with the setup process.
UP
999.9
<UP>
<DWN>
<DSBL>
Pen Max = 999.9999 %
999.9999
DSBL
DSBL
ENTER
The alarm will be activated at readings
less than this set point (Range: -0.0005
to 99.9). If the digits are replaced by
dashes (“-“), the alarm has been
disabled.
Increases the setpoint.
Decreases the setpoint.
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
Res Max =
12.2. Maximum Percent Penetration
Variable (Alarm Point)
DWN
DWN
12.4. Maximum Resistance Variable
(Alarm point)
NOTE: The units in subsequent pressure displays will
reflect the selected unit choice made here. For
the purposes of this manual, we will assume
that inH20 is the selected unit.
UP
0.0001 to 99.9). If the digits are
replaced by dashes (“-“), the alarm has
been disabled.
Increases the setpoint.
Decreases the setpoint.
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
ENTER
The alarm will be activated at readings
greater than this set point (Range:
19
999.9 mmH2O
DWN
DSBL
ENTER
Alarm activated at readings greater than
this set point (Range: 0 – 250 mmH2O)
Increases the variable
Decreases the variable
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
TDA-110P
<ENBL>
<ENTER>
Air Techniques International
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
<ENTER>
1800103 Rev C
Accepts the setting and enters the next
Setup menu screen
12.6. Test Flow Tolerance Variable
(Alarm Point)
Flow Tol =
12.5. Minimum Resistance Variable
(Alarm point)
Res Min =
UP
999.9
<UP>
<DWN>
<DSBL>
<ENBL>
UP
99
999.9 mmH2O
DWN
DSBL
ENTER
<UP>
<DWN>
<DSBL>
Alarm activated at readings less than
this set point (Range: 0 – 250 mmH2O)
Increases the variable
Decreases the variable
Disable the alarm function.
When pressed, “DSBL” changes to
“ENBL” and the digits are replaced by
dashes (“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
<ENBL>
<ENTER>
<ENTER>
12.7. Load Time Variable
The load time variable is the time from when the
chuck is closed to when the unit begins to measure
the properties of the sample. This time is used to
establish a stabile aerosol flow.
Load Time =
UP
99
<UP>
<DWN>
ENTER>
UP
99
<UP>
<DWN>
<DSBL>
NOTE: This variable must be greater than the
normal Load Time variable.
<ENBL>
99 sec
<ENTER>
99
<UP>
<DWN>
Accepts the setting and enters the next
Setup menu screen
Sample Time =
Sets the time from when the chuck is closed to
when the test starts to sample. The extended load
time is used if the previous test failed or if a
calibration has just been performed.
DWN
Alarm activated at test flow readings +/this value (Range: 1 – 10)
Increases the variable
Decreases the variable
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
NOTE: If the sample time is disabled, the test
continues until either chuck switch is pressed to end
the test.
ENTER
12.8. Extended Load Time Variable
UP
ENTER
The sample time variable is the length of time the
test is sampled in one-second intervals. The unit
displays the test results at approximately one second
intervals during the testing.
Load time in seconds. (Range 1 – 60)
Increases the variable
Decreases the variable
Accepts the setting and enters the next
Setup menu screen
Ext Ld Time =
DSBL
12.9. Sample Time Variable
99 sec
DWN
DWN
± 99 cfm
ENTER
Extended Load Time in seconds.
(Range the value set for the load time to
100)
Increases the variable
Decreases the variable
20
DWN
99 sec
DSBL
ENTER
Time in seconds. (Range: 1 – 3600)
Increases the variable
Decreases the variable
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
1800103 Rev C
Air Techniques Incorporated
12.10. Open Time Variable
Concentration is
The open time variable is used to delay the start
of the next test cycle
Open Time =
UP
99.9
<UP>
<DWN>
<ENTER>
99.9
ENTER
<DSBL>
<SET >
<ENTER>
9999
<DSBL>
<UP>
<DWN>
<DSBL>
<ENBL>
<ENTER>
DWN
DSBL
ENTER
Disables the alarm function
Activates the alarm function
Accepts the setting and enters the next
Setup menu screen.
12.13. Set Date and Time
Set/Enter screen
This screen allows the user to select whether
they want to change the date/time recorded in the
machine or proceed to the next screen.
Sets the number of tests that must be completed
between prompts for system recalibration (See
Section 9.6). When the specified number of tests has
been completed, the system will prompt the operator
to recalibrate the penetration. This same interval is
used to update the temperature and barometric
pressure if the ”Auto” selection is made for flow
compensation is selected. (Section 12)
UP
DSBL
SET
12.11. Cycles to Calibration
9999
ENTER
Concentration is
Open Time in seconds. (Range 0 to 60)
Increases the variable.
Decreases the variable.
Accepts the setting and enters the next
Setup menu screen.
Cycles to Cal =
SET
DSBL
sec
DWN
TDA-110P
mm/dd/yy hh:mm
SET
ENTER
mm
dd
yy
Month (1 to 12)
Day of Month (1 to last day of month)
Last two digits of year. (The first two
digits are assumed to be “20”.)
hh
Hour of the day using the 24 hour clock.
(0 to 23)
mm
Minutes in the hour. (0 to 59)
<SET>
Enters the date/time setting mode
below.
<ENTER> Accepts the setting and enters the next
Setup menu screen.
Date/Time Setting Mode
In the date/time set mode, each group of digits
(month, day, year, etc.) is incremented or
decremented in turn by the <UP> and <DWN>
buttons. Pressing the <CONT> button will advance
to the next group of digits.
CTS
ENTER
(Range 1 – 9999)
disables the function
Increases the variable
Decreases the variable
Disable the alarm function. When
pressed, “DSBL” changes to “ENBL”
and the digits are replaced by dashes
(“-“).
Enable the alarm function. When
pressed, “ENBL” changes to “DSBL”
and the digits of the limit are displayed.
Accepts the setting and enters the next
Setup menu screen
mm/dd/yy hh:mm
12.12. Concentration Reference Alarm
UP
After the Penetration Setup menu, the test
3
aerosol concentration in mg/m is displayed. A
concentration error will occur if the concentration falls
outside of the concentration alarm point window.
(See Section 15.15 – Gravimetric Setup in the
Maintenance section)
<UP>
<DWN>
<CONT>
21
DWN
CONT
Increases the variable
Decreases the variable
Advance to the next group of digits.
After the minutes group, returns to the
Set/Enter Screen.
TDA-110P
Air Techniques International
1800103 Rev C
13. Maintenance
Weekly
13.1. Initial Setup
The following maintenance routines should be run
to familiarize the user with these functions as well
as set up the unit for operation in the user’s
environment: Flow Balance, Pressure Check (not
pressure setup), and Gravimetric Test.

Check the chuck for debris and clean as
needed.

Check pressure differential gauges on blower
pre-filter and HEPA filter. Replace if pressure
exceeds 0.5in.
Monthly
13.2. Definitions and Features
Gravimetric – Used to measure and set the system
aerosol concentration level and alarm points (See
Section 15.15 - Gravimetric Setup)

Pressure Check and Setup – Used to check or
recalibrate the system pressure sensor to an
external source (See Sections 15.18 - Pressure
Check and Pressure Setup)
Biannually
Security – Used to prevent changes to the alarm
values or critical load and sample times (See
Section 15.13 - Security)
LSC Check (Light Scattering Chamber) – Used to
adjust the LSC after maintenance and for general
trouble shooting (See Section 15.14 - LSC Check)
Check the system tubing for oil or particle
buildup and replace if necessary.

Check the pressure sensor against an external
source and calibrate if necessary.

Replace the Test Flow After-filter (carbon/HEPA
filter).

Replace the Reference (HEPA) filter.
Annually
Dip Switches – Used for initial system setup (See
Appendix A - Dip Switch Settings)
13.3. Scheduled Maintenance

Clean the LSC (see Section 7 - LSC
Maintenance).

Have the mass flow meter calibrated.
13.4. Spare Parts List
The following provides a one-year supply.
NOTE: Scheduled maintenance is based on an 8hour operating day. If operating time is increased,
the frequency of maintenance operations should be
increased as well.
1 ea.
1 ea.
12 ft.
12 ft.
30 ft.
1 ea
Daily

Check the liquid level in the aerosol generator
and refill as needed.

Check the liquid level in the drain jar and empty
if necessary.

Check the compressed air input filter for liquid
and drain if necessary.
1 ea

Check blower inlet for obstructions and clean as
necessary.
2 ea
22
HEPA -Cap 75 mm, w/barbs Part # 5500128
HEPA –Cap 75 mm, Part # 5500122
Clear ½” Tubing, Part # 5200155
Clear 1/4” Tubing, ATI Part # 5200138
Black 3/8” Tubing, ATI Part # 5200140
MERV8 2 inch Pre-filter, Flanders,
Precisionaire, Pre Pleat 62R, Part #
2242BPFTPR
24 x 24 x 12 HEPA Filter, Flanders
Precisionaire, Superflow 2K, Part # SF24-8G2-GG
Sanitary Clamp Silicone Gasket, ATI Part #
4300127.
1800103 Rev C
Air Techniques Incorporated
TDA-110P
Exit Maintenance?
14. Maintenance Menu Structure
14.1. Maintenance Menus
The menu structure within the Maintenance
portion of the TDA-100P operating program is laid
out in a loop. Using the <Next> and/or <Prev> keys
allows navigation through the loop until arriving
again at the entry point. Selecting <Cont> enters
the feature displayed in the topmost row.
The top level Maintenance features are
“Gravimetric”, “Res Offset”, “Flow Balance”, “LSC
Check”, “LSC Offset”, “View Sensor”, “Cal Sensor”,
“ID”, “Console” & “Security”. After “Security” the
menu loops back to “Gravimetric”.
Yes
No
A
B
<Yes>
<No>
C
D
Exits Maintenance and restarts the unit.
Returns to Maintenance entry menu
Maint: Res Offset
Prev
Next
Cont
Back
A
B
C
D
<Prev> Returns to Gravimetric selection
The 100P will enter maintenance mode in three
circumstances:
<Next> Proceeds to Flow Balance selection
1. In response to an error such as an LSC failure
that requires use of maintenance mode to test
and recalibrate the system.
<Cont> Enters Res Offset
<Back> Prompts for Maintenance exit
2. If the internal memory becomes corrupted. This
will require that all the maintenance calibration
routines be performed to restore the machine to
an operational state.
Maint: Flow Balance
3. After a restart of the machine.
To enter
maintenance mode after a restart, wait until the
System Version is displayed, and then press
and hold the leftmost function key, ‘A’, until the
display changes or a short beep is heard.
Prev
Next
Cont
Back
A
B
C
D
<Prev> Returns to Res Offset selection
<Next> Proceeds to LSC Check selection
<Cont> Enters Flow Balance
4. To re-initialize ALL values stored in the units
battery-backed RAM substitute the rightmost
function key, ‘D’, for the ‘A’ key called out in
step 3 above.
<Back> Prompts for Maintenance exit
Maint: LSC Check
* Please note that once this initialization has
been performed the previous values cannot be
recovered.
Top level Maintenance entry menu
Prev
Next
Cont
Back
A
B
C
D
<Prev> Returns to Flow Balance selection
Maint: Gravimetric
<Next> Proceeds to LSC Offset selection
Prev
Next
Cont
Back
A
B
C
D
<Cont> Enters LSC Check
Back>
<Prev> Proceeds to Security selection
<Next> Proceeds to Res Offset selection
<Cont> Enters Gravimetric test
<Back> Prompts for Maintenance exit
Selecting <Back> from the Maintenance entry
menu will result in a Yes or No prompt, as shown
next to confirm the intent to leave Maintenance.
23
Prompts for Maintenance exit
TDA-110P
Air Techniques International
1800103 Rev C
Maint: LSC Offset
Prev
Next
Cont
Back
A
B
C
D
Maint: ID
<Prev> Returns to LSC Check selection
Prev
Next
Cont
Back
A
B
C
D
<Next> Proceeds to View Sensor selection
<Prev> Returns to Cal Sensor selection
<Cont> Enters LSC Offset
<Next> Proceeds to Console selection
<Back> Prompts for Maintenance exit
<Cont> Enters ID
<Back> Prompts for Maintenance exit
Maint: View Sensor
Prev
Next
Cont
Back
A
B
C
D
Maint: Console
<Prev> Returns to LSC Offset selection
Prev
Next
Cont
Back
A
B
C
D
<Next> Proceeds to Cal Sensor selection
<Prev> Returns to ID selection
<Cont> Enters View Sensor
<Next> Proceeds to Security selection
<Back> Prompts for Maintenance exit
<Cont> Initiates Console
<Back> Prompts for Maintenance exit
Maint: Cal Sensor
Prev
Next
Cont
Back
A
B
C
D
Maint: Security
<Prev> Returns to View Sensor selection
Prev
Next
Cont
Back
A
B
C
D
<Next> Proceeds to ID selection
<Prev> Returns to Console selection
<Cont> Enters Cal Sensor
<Next> Proceeds to Gravimetric test
<Back> Prompts for Maintenance exit
<Cont> Enters Security
<Back> Prompts for Maintenance exit
24
1800103 Rev C
Air Techniques Incorporated
Loading --
14.2. Gravimetric
TDA-110P
### sec left
Flow = ##.#
Abort
Maint: Gravimetric
A
Prev
Next
Cont
Back
A
B
C
D
B
C
D
“Sec left” is the time remaining in the Gravimetric
test until completion.
“Flow=##.#” is the sample flow passing through the
weighed sample in liters per minute. If the flow
decreases during the test due to loading, this may
be adjusted using the Flow Adjustment control.
A gravimetric reading is used to determine the
aerosol concentration of the unit. Measuring the
amount of aerosol collected and the total airflow
during the sample time performs this as explained
below.
<Abort> Stops the test in progress and returns to
Maintenance entry menu.
<Prev> Proceeds to Security selection
<Next> Proceeds to Res Offset selection
Loading Complete
<Cont> Enters Gravimetric test
<Back> Prompts for Maintenance exit
AvgFlow = ##.#
A
B
Enter
Exit
C
D
Loading Complete
Returns to Maintenance entry menu.
TotFlow = ###.#
A
Wait
30 second countdown begins before entry into
Gravimetric test
D
Determine the amount of weight increase for the
media used during the gravimetric test by
subtracting the starting weight from the ending
weight measured after completion of “Loading”
Exit
C
C
<Cont> proceeds to the next step
Close Chuck to Start
B
B
Cont
“TotFlow = ###.#” is the total volume of sample flow
that passed through the pre-weighed media during
the Gravimetric test period. Record this number.
30 seconds remaining
A
D
<Cont> proceeds to the next step
<Enter> Enters Gravimetric test routine
<Exit>
C
“AvgFlow = ##.#” is the average sample flow
present during the Gravimetric test period.
Gravimetric Test
A
B
Cont
Ending weight – starting weight = weight increase
D
Divide the weight increase by the recorded TotFlow
value.
The result obtained is the aerosol
concentration expressed as weight of aerosol (ug)
per volume of air (l).
Weigh a sample of high efficiency media large
enough to completely fill the test fixture opening
and record the weight. No sample air flow should
be able to pass through the test fixture without
passing through the media.
Weight increase ÷ TotFlow = concentration in ug/l
Place the weighed media in the test fixture.
With the media in place depress both actuator
switches at the same time to initiate the Gravimetric
test.
25
TDA-110P
Air Techniques International
1800103 Rev C
<UP> Increases the concentration limit
Store New Values?
<DWN> Decreases the concentration limit
Yes
No
A
B
C
<ENTER>Accepts the value displayed as the lower
concentration limit and continues to the next step.
D
The calculated “Lo Con” value displayed will be the
valued entered for “Con” - 20%. This value may be
increased or decreased at the discretion of the enduser.
<Yes> Proceeds to the next step, memory
storage routine, to set and capture the gain levels
necessary to achieve a 100% response from the
available aerosol.
<No> Exits the Gravimetric test menu and returns
to the Maintenance entry menu.
Clear Chuck & Close
Con. =
Exit
### ug/l
UP
DWN
A
B
A
ENTER
C
D
<DWN> Decreases the displayed Conc. Value
<ENTER> Accepts the entered Conc. Value as the
calculated aerosol concentration and continues to
the next step.
The operating program then immediately returns to
the top level Maintenance entry menu.
### ug/l
DWN
A
B
14.3. Res Offset
ENTER
C
Maint: Res Offset
D
“Hi Con. = ### ug/l” is the upper concentration limit
for the “Con Check” variable.
<UP> Increases the concentration limit
<DWN> Decreases the concentration limit
<ENTER>Accepts the value displayed as the upper
concentration limit and continues to the next step
The calculated “Hi Con” value displayed will be the
valued entered for “Con” + 20%. This value may be
increased or decreased at the discretion of the enduser.
Lo Con. =
DWN
A
B
ENTER
C
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Maintenance entry menu
<Next>
Proceeds to Flow Balance selection
<Cont>
Enters Res Offset menu
<Back> Prompts for Maintenance exit
### ug/l
UP
D
After a loading period, the microprocessor will
adjust the sensitivity of the LSC to obtain the
required voltage level. At the end of this adjustment
the display will briefly show the 100% voltage level.
The Sample/Clear valve then changes to the clear
position to provide the LSC with particle free air.
After a short delay the Clear 0 baseline voltage
(stray light) will be stored to memory and briefly
displayed.
<UP> Increases the displayed Conc. value
UP
C
Make sure that nothing is in the test fixture.
Depress both actuator switches at the same time to
initiate the completion of this routine.
“Con. = ### ug/l” is calculated using the values
obtained during the gravimetric test.
Hi Con. =
B
D
“Lo Con. = ### ug/l” is the lower concentration limit
of the “Con Check” variable.
26
1800103 Rev C
Res Offset -
Air Techniques Incorporated
TDA-110P
#.## mm/H2O
Set
Dsbl
A
B
Flow is now
Exit
C
D
“Res Offset #.## mm/H2O” is the resistance
offset value currently stored in memory.
<Enbl/Dsbl> Toggles
between
enabled
and
disabled states. The Enabled state displays the
stored offset value (±) to be applied to each test
result. The Disabled state displays -----.
A
B
B
C
D
The flow balance is adjusted by alternating between
the Sample and Clear states, using the ON and
OFF selections, while observing the sample flow
display and adjusting the gate valve installed
immediately below the 0 Reference filter. The flow
balance setting is complete when the displayed
values during each state match with ± 0.1 LPM.
Enter
C
A
EXIT
<Exit> Returns to the Maintenance entry menu
#.## mm/H2O
Dwn
OFF
<OFF> Disables the Sample/Clear solenoid
valve. CLEAR displayed. Sample flow path is
through the 0 Reference filter.
<Exit> Returns to the Maintenance entry menu
Up
ON
<ON> Enables the Sample/Clear solenoid valve.
Sample displayed. Sample flow path is through the
test fixture.
<Set> Proceeds to the Res Offset adjustment menu
below
Res Offset -
SAMPLE (CLEAR)
D
14.5. LSC Check
“Res Offset #.## mm/H2O” is the resistance
offset value currently stored in memory.
Maint: LSC Check
<UP> Increases the Res Offset value
<DWN> Decreases the Res Offset value
<ENTER>Accepts the value displayed as the Res
Offset and returns to the previous screen.
14.4. Flow Balance
Maint: Flow Balance
Prev
Next
Cont
Back
A
B
C
D
Returns to Gravimetric selection
<Next>
Proceeds to Flow Balance selection
<Cont>
Enters Res Offset menu
Next
Cont
Exit
A
B
C
D
<Prev>
Returns to Gravimetric selection
<Next>
Proceeds to Flow Balance selection
<Cont>
Enters Res Offset menu
<Exit> Returns to the Maintenance entry menu
The menu is used to observe the LSC operating
voltages for basic troubleshooting purposes.
This routine is used to correct for air flow resistance
differentials between the sample flow path and the
clear airflow path.
This is done to minimize
response lag in the mass flow meter.
<Prev>
Prev
LSC Check
- Close
EXIT
A
B
C
D
<Exit> Returns to the Maintenance entry menu
<Back> Prompts for Maintenance exit
The test fixture should be closed at this time.
Closure may be accomplished using either manual
actuators or an EXT Start signal sent to the User
Port.
Close Chuck
The test fixture should be closed at this time.
Closure may be accomplished using either manual
actuators or an EXT Start signal sent to the User
Port.
27
TDA-110P
Air Techniques International
LSC Voltage
=
<Enbl> indicates the feature is disabled and
pressing the key will enable the LSC Offset
#.####
ON
EXIT
A
B
C
<Exit> Returns to the Maintenance entry menu.
D
“LSC Voltage = #.####” displays the voltage output
signal for the LSC.
<ON>/<OFF> Enable/Disables the Sample/Clear
solenoid valve.
<ON> Sample valve is enabled. The displayed
voltage represents the signal generated by the
100%
sample
of
the
available
aerosol
concentration.
<OFF> Sample valve is disabled. The displayed
voltage represents the signal generated by
sampling particle free air through the 0 Reference
filter. This is the “straylight” voltage.
<Exit> Returns to the Maintenance entry menu
14.6. LSC Offset
Maint: LSC Offset
Prev
Next
Cont
Exit
A
B
C
D
Used to access and set the LSC Offset variable.
The value is used to correct for signal shifts in the
LSC output caused by DC power variations during
operation.
<Prev>
Returns to LSC Check selection
<Next>
Proceeds to View Sensor selection
<Cont>
Enters LSC Offset menu
<Exit> Returns to the Maintenance entry menu
Note: Under most applications this variable is more
easily and accurately set using Console mode
access.
LSC Offset
-
-###
Set
Dsbl
A
B
“LSC Offset Offset value.
disabled state.
Exit
C
1800103 Rev C
D
-###” displays the stored LSC
A display of “------“indicates a
<Set> Enters LSC Offset adjustment routine.
<Dsbl>/<Enbl> Enables or disables the LSC Offset
feature.
<Dsbl> indicates the feature is enabled and
pressing the key will disable the LSC Offset.
28
1800103 Rev C
Air Techniques Incorporated
14.7. View Sensor
Insert Clear Filter
Maint: View Sensor
CONT
A
<CONT>
B
C
D
Used as an acknowledgement that a
known filter has been placed in series
with the LSC inet port, replacing the
Sample/Clear valve sample line. The
Sample/Clear valve will energize at
this time.
C
Cont
Exit
A
B
C
D
<Prev>
Returns to LSC Offset selection
<Next>
Proceeds to Cal Sensor selection
<Cont>
Enters View Sensor menu
D
The TDA-100P will sample for a period of time to
stabilize and record the LSC output signal in this
sampling state.
Selecting Pressure
Remove Filter
CONT
<CONT>
Next
The menu is intended as a troubleshooting aid and
allows viewing of the raw input voltages as
interpreted by the 100P microprocessor.
B
A
Prev
<Exit> Returns to the Maintenance entry menu
Finding Zero
A
TDA-110P
B
C
D
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to LSC sensor output
<Next>
Proceeds to Flow senor output
<Cont>
Enters Pressure sensor output
<Back> Returns to the Maintenance entry menu
Used as an acknowledgement that the
previously installed filter has been
removed and the LSC inet port tubing
connections restored. The
Sample/Clear valve will de-energize at
the appearance of this screen.
Voltage = #.#### V
Exit
A
B
C
D
“Voltage = #.### V” displays the output signal being
received from the differential pressure transducer.
Finding Zero
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
A
B
C
D
The TDA-100P will sample for a period of time to
stabilize and record the LSC output signal in this
sampling state before returning to the Maintenance
entry menu.
Selecting Flow
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Pressure sensor output
<Next>
Proceeds to Aux Flow senor output
<Cont>
Enters Flow sensor output
<Back> Returns to the Maintenance entry menu
29
TDA-110P
Air Techniques International
1800103 Rev C
Voltage = #.#### V
Temp = ##.## C
Exit
A
B
C
(Raw = #.#### V )
D
“Voltage = #.### V” displays the output signal being
received from the mass flow meter..
A
B
Exit
C
D
“Temp = ##.## C” displays the output signal being
received from the PCBA mounted temperature
sensor converted to degrees centigrade using the
conversion constant stored in memory.
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
“(Raw = #.#### V)” displays the PCBA mounted
temperature sensor output as a voltage.
Selecting Aux Flow
Prev
Next
Cont
Back
A
B
C
D
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
This feature is currently not used and is
intended for future applications.
<Prev>
Proceeds to Flow sensor output
<Next>
Proceeds to Temp senor output
<Cont>
Enters Aux Flow sensor output
Selecting Baro Pres
<Back> Returns to the Maintenance entry menu
Voltage = #.#### V
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Temp sensor output
<Next>
Proceeds to LSC senor output
<Cont>
Enters Baro Pres output
<Back> Returns to the Maintenance entry menu.
Exit
A
B
C
B.P. = ##.## inHg
D
“Voltage = #.### V” displays the output signal being
received from the auxiliary mass flow meter, if
installed and enabled.
(Raw = #.#### V )
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
“Temp = ##.## inHg” displays the output signal
being received from the PCBA mounted barometric
pressure sensor converted to degrees inches of
mercury using a fixed conversion constant.
A
.
B
Exit
C
D
“(Raw = #.#### V)” displays the PCBA mounted
barometric pressure sensor output as a voltage.
Selecting Temp
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Aux Flow sensor output
<Next>
Proceeds to Baro Pres senor output
<Cont>
Enters Temp output
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
<Back> Returns to the Maintenance entry menu.
30
1800103 Rev C
Air Techniques Incorporated
TDA-110P
Selecting LSC
Prev
Next
Cont
14.8. Cal Sensor
Back
Maint: Cal Sensor
A
B
C
D
<Prev>
Proceeds to Baro Pres sensor output
<Next>
Proceeds to Pressure senor output
<Cont>
Enters LSC sensor output
LSC=#.####
Up
Dwn
Samp
Exit
A
B
C
D
Next
Cont
Back
A
B
C
D
WARNING: The screens associated steps within
this section are used to verify and/or calibrate the
sensors used during the operation of the TDA100P.
Known calibration standards, specific
supplies and fixtures may be required and should
be readily available
<Back> Returns to the Maintenance entry menu
DAC = ##,
Prev
Only Trained and Authorized personnel should
access these features.
Improper or partial
completion of the steps outlined will render the
TDA-100P inoperable until successful completion.
The View LSC screen is intended as the primary
working maintenance interface for light scattering
chamber. Using this screen, all necessary output is
available to enable optical alignment, PMT
alignment & troubleshooting as well as leak testing.
<Prev>
Returns to View Sensor selection
<Next>
Proceeds to ID selection
<Cont>
Enters Cal Sensor menu
<Back> Returns to the Maintenance entry menu.
The test fixture should be closed at this time.
Closure may be accomplished using either manual
actuators or an EXT Start signal sent to the User
Port.
Selecting Pressure
“DAC = ##” displays the stored gain sensitivity
setting for the LSC photomultiplier tube
“LSC = #.####” displays the output voltage from the
light scattering chamber under its current
conditions.
Prev
Next
Cont
Back
A
B
C
D
<Prev> Proceeds to Gravimetric test.
See Section 14.2
<Up> Increases the DAC value by one at each
press. Changes made using this feature are not
retained upon exit.
<Next>
Proceeds to Flow calibration menu
<Cont>
Enters Pressure sensor calibration
<Dwn> Decreases the DAC value by one at each
press. Changes made using this feature are not
retained upon exit.
<Back> Returns to the Maintenance entry menu
<Samp>/<Cler> Enable/Disables the Sample/Clear
solenoid valve.
Pressure Check?
<Samp> Sample valve is disabled. The displayed
voltage represents the signal generated by
sampling particle free air through the 0 Reference
filter. This is the “straylight” voltage.
Yes
No
A
B
C
D
<Yes> enters a pressure verification menu.
<No> Proceeds to Pressure sensor calibration.
<Cler> Sample valve is enabled. The displayed
voltage represents the signal generated by the
100%
sample
of
the
available
aerosol
concentration.
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
31
TDA-110P
Air Techniques International
1800103 Rev C
Set for – 0 - WC
** Are You Sure? **
CONT
A
B
C
D
Disconnect both lines from the pressure transducer
so that the transducer inlets are bleeding to
atmosphere (0” water column).
Attach a pressure calibrator and pressure source in
series with the transducer inlet port labeled P2
(uppermost). The pressure calibrator/source should
not be sourcing pressure to the transducer at this
stage.
<CONT> enters a pressure verification routine.
Pressure
#.####
YES
NO
A
B
DEFAULT
C
D
<YES>
Proceeds to pressure calibration
<NO>
Returns to Cal Sensor entry menu.
<DEFAULT> Loads a default system pressure value
to memory and returns to Cal Sensor
entry menu.
NOTE:
Loading “Default” values is not
recommended except in extreme circumstances.
Always be prepared to select <Yes> and set up the
pressure transducer with a NIST traceable pressure
calibrator.
Exit
Set for – 0 - WC
A
B
C
D
CONT
“Pressure
#.###” displays the output signal
being received from the pressure sensor in mm/H2O
calculated using the conversion constant stored in
memory.
A
B
C
D
Disconnect both lines from the pressure transducer
so that the transducer inlets are bleeding to
atmosphere (0” water column).
Close any open bleed valves on the pressure
calibrator/source and slowly apply pressure to the
pressure transducer inlet port labeled P2.
Attach a pressure calibrator and pressure source in
series with the transducer inlet port labeled P2
(uppermost). The pressure calibrator/source should
not be sourcing pressure to the transducer at this
stage.
Allow the pressure to stabilize at 10, 40 & 80
mm/H2O and compare the TDA-100P pressure to
the pressure calibrator displayed pressure. The
TDA-100P pressure must correspond to the
pressure calibrator display within ± 0.5 mm/H2O at
all three (3) pressure set points.
<CONT> enters a pressure verification routine.
Record - 0 -
<Exit> Proceeds to the Pressure Setup menu
#.####
ENTER
A
Pressure Setup
A
B
Exit
Enter
C
D
B
C
D
“Pressure
#.###” displays the output signal
being received from the pressure transducer as
voltage.
<ENTER> Stores the transducer output signal at
atmospheric pressure memory.
<Exit> Returns to Cal Sensor entry menu.
<Enter> Proceeds to pressure sensor calibration.
If all pressure set point readings in the pressure
check menu were acceptable, restore the pressure
transducer tubing connections and press ‘C’
<EXIT>.
If any pressures set point readings in the pressure
check menu were unacceptable, press ‘D’
<ENTER>.
32
1800103 Rev C
Air Techniques Incorporated
14.9. Primary (HIGH) Flow Sensor
Setup
Set Pressure
“Flow1” Input on PCBA
Feature used to enter Flow Sensor calibration
data.
CONT
A
B
C
D
Note – if the flow sensor has been calibrated by
its manufacturer then this procedure is not required.
<CONT> This screen allows a user controlled
pause for adjustment of the external pressure
applied to the transducer. The applied pressure
value, as measured by the pressure calibrator,
should be in the upper 25% of the transducer range
(between 75 to 100 mm/H2O). Press CONT when
the reading displayed on the calibrator has
stabilized within the acceptable range.
Record Pres.
Selecting Flow
#.####
B
C
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Pressure sensor calibration
<Next>
Proceeds to Aux Flow senor calibration
<Cont>
Enters Flow sensor calibration
<Back> Returns to the Maintenance entry menu
ENTER
A
TDA-110P
Adj Flow to Zer0
D
CONT
“Pressure
#.###” displays the output signal
being received from the pressure transducer as
voltage.
A
EXIT
B
C
D
<ENTER> Records the transducer output signal at
the applied pressure to memory.
<CONT> Continues to the flow meter zero output
signal display screen.
NOTE: It is important that the exact value displayed
on the pressure calibrator in mm/H2O be noted at
the moment <ENTER> is pressed. This value is
required for the next step.
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
At this step the blower should be OFF to prevent
any air currents from being detected..
Flow Zero =
Input Pres
=
UP
DWN
A
B
###.#
SET
ENTER
C
A
displayed
B
C
D
<SET> Stores the displayed voltage to memory.
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
<UP> Increases the displayed pressure calibration
point.
the
EXIT
“Flow Zero = #.####” V displays the flow meter
signal being received with no air flow through the
sensor.
D
“Input Pres = ## #.#” displays the pressure that will
be use as the calibration point determined in the
preceding step.
<DWN> Decreases
calibration point.
#.####
Select Cal Type
pressure
VEL
<ENTER> Stores the entered value to memory and
returns to the initial View Sensor menu, “Selecting
Pressure”.
A
VOL
B
C
D
<VEL> Proceeds to the Velocity based calibration
routine using feet per minute (fpm) and the crosssectional area where the sensor is installed.
.
33
TDA-110P
Air Techniques International
<VOL> Proceeds to the Volumetric based
calibration routine using cubic feet per minute (cfm).
VOLUMETRIC CALIBRATION
When Flow = ### cfm
UP
DN
A
B
ENTER
C
D
<UP> Increases the displayed flow calibration value
in cfm.
<DWN> Decreases the displayed flow calibration
value in cfm.
<ENTER> Stores the entered value to memory.
Flow Volts =
#.####
UP
DWN
A
B
ENTER
C
D
“Flow Volts = #.####” displays a user entered Ref
Flow voltage (5.00VDC typical) from a NIST
Traceable calibration sheet..
<UP> Increases the displayed flow calibration
voltage.
<DWN> Decreases the displayed flow calibration
voltage.
<ENTER> Stores the entered value to memory and
returns to CAL SENSOR menu.
VELOCITY CALIBRATION
When Flow = ####.# fpm
UP
DN
A
B
ENTER
C
D
<UP> Increases the displayed flow calibration point
in feet per minute.
<DWN> Decreases the displayed flow calibration
point in feet per minute.
<ENTER> Stores the entered value to memory.
34
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
Flow Volts =
Enable Aux Flow Sens
#.####
UP
DWN
A
B
ENTER
C
D
“Flow Volts = #.####” displays a user entered Ref
Flow voltage (5.00VDC typical) from a NIST
Traceable calibration sheet.
YES
NO
A
B
<Yes>
A
B
B
C
D
<CONT> Continues to the flow meter zero output
signal display screen.
ENTER
C
EXIT
A
#.##”
DWN
Enters Aux Flow sensor calibration
CONT
<ENTER> Stores the entered value to memory.
UP
D
Adj Aux Flow to Zero
<DWN> Decreases the displayed flow calibration
voltage.
=
C
<No> Returns to the Maintenance entry menu
<UP> Increases the displayed flow calibration
voltage.
Duct Dia.
TDA-110P
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
D
“Duct Dia. = #.####” displays the diameter to be
used for calculating the cross-sectional area
(inches).
At this step the blower should be OFF and cap the
probe to eliminate any air currents from being
detected.
<UP> Increases the displayed duct diameter.
Flow Zero =
<DWN> Decreases the displayed duct diameter.
#.####
SET
<ENTER> Stores the entered value to memory and
returns to CAL SENSOR menu.
EXIT
A
B
C
“Flow Zero = #.####” V” displays the flow meter
signal being received with no air flow through the
sensor.
14.10. Auxiliary (LOW) Flow Sensor
Setup
“Flow2” Input on PCBA
<SET> Stores the displayed voltage to memory.
Feature used to enter Flow Sensor calibration
data for the auxiliary flow sensor.
<Exit> Returns to the initial View Sensor menu,
“Selecting Pressure”.
Note – if the flow sensor has been calibrated by its
manufacturer then this procedure is not required.
Select Cal Type
Selecting Aux Flow
VEL
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Flow sensor calibration
<Next>
Proceeds to Temp senor calibration
<Cont>
Enters Aux Flow sensor calibration
D
A
VOL
B
C
D
<VEL> Proceeds to the Velocity based calibration
routine using feet per minute and the crosssectional area where the sensor is installed.
<VOL> Proceeds to the Volume based calibration
routine using cubic feet per minute.
<Back> Returns to the Maintenance entry menu
35
TDA-110P
Air Techniques International
1800103 Rev C
VOLUMETRIC CALIBRATION
<ENTER> Stores the entered value to memory.
When Flow = ### cfm
Duct Dia.
UP
DN
A
B
C
= #.##”
ENTER
UP
DWN
D
A
B
ENTER
C
D
“Duct Dia. = #.####” displays the diameter to be
used for calculating the cross-sectional area
(inches).
<UP> Increases the displayed flow calibration value
in cfm.
<DWN> Decreases the displayed flow calibration
value in cfm.
<UP> Increases the displayed duct diameter.
<ENTER> Stores the entered value to memory.
<DWN> Decreases the displayed duct diameter.
Flow Volts = #.####
<ENTER> Stores the entered value to memory and
returns to CAL SENSOR menu.
UP
DWN
A
B
ENTER
C
D
.Selecting Temp
“Flow Volts = #.####” displays a user entered Ref
Flow voltage (5.00VDC typical) from a NIST
Traceable calibration sheet..
Prev
Next
Cont
Back
A
B
C
D
<UP> Increases the displayed flow calibration
voltage.
<Prev>
Proceeds to Aux Flow sensor calibration
<Next>
Proceeds to Baro Pres sensor calibration.
<Cont>
Enters Temp output calibration
<DWN> Decreases the displayed flow calibration
voltage.
<ENTER> Stores the entered value to memory and
returns to CAL SENSOR menu.
<Back> Returns to the Maintenance entry menu.
VELOCITY CALIBRATION
Current Temp = 20.0 C
When Flow = ####.# fpm
UP
DN
A
B
ENTER
C
D
ENTER
C
C
D
<ENTER> Stores the temperature value to memory
and returns to the initial View Sensor menu,
“Selecting Pressure”.
Flow Volts = #.####
B
B
<DWN> Decreases the displayed temperature
value.
<ENTER> Stores the entered value to memory.
A
A
ENTER
<UP> Increases the displayed temperature value.
<DWN> Decreases the displayed flow calibration
point in feet per minute.
DWN
DWN
“Current Temp = 20.0 C” displays the default
temperature value.
<UP> Increases the displayed flow calibration point
in feet per minute.
UP
UP
D
“Flow Volts = #.####” displays a user entered Ref
Flow voltage (5.00VDC typical) from a NIST
Traceable calibration sheet.
<UP> Increases the displayed flow calibration
voltage.
<DWN> Decreases the displayed flow calibration
voltage.
36
1800103 Rev C
Air Techniques Incorporated
Selecting Baro Pres
TDA-110P
Set#####
Prev
Next
Cont
Back
UP
DWN
A
B
C
D
A
B
<Prev>
Proceeds to Temp sensor calibration
<Next>
Proceeds to LSC senor calibration
<Cont>
Enters into Baro Pres calibration.
ENTER
C
D
“Set##### displays the current value stored as the
variable UNIT_ID.
<UP> Increases the ID numeric value
<DWN> Decreases the ID numeric value
<Back> Returns to the Maintenance entry menu
<ENTER> Stores the entered value to memory.
The ability to alter the ID value is also available
using the “SET UNIT_ID=#####” command through
the Console feature
Not Available
CONT
A
<Cont>
B
C
14.12. Console
D
Console
Returns to the Maintenance entry menu.
Prev
Calibration of the barometric pressure sensor is
performed at the manufacturing level. No postmanufacturing calibration is possible
Next
Cont
Back
<Prev>
Proceeds to ID selection
<Next>
Proceeds to Security selection
<Cont> Enters Console mode
See related Appendix
Selecting LSC
Prev
Next
Cont
Back
A
B
C
D
<Prev>
Proceeds to Baro Pres sensor calibration
<Next>
Proceeds to Pressure senor calibration
<Back> Prompts for Maintenance exit.
Note: Once Console mode has been engaged an
”END” command must be received through the RS232 communication port to restore unit control
panel functionality.
<Cont> Proceeds to Gravimetric test.
See Section 14.2
14.13. Security
Maint: Security
<Back> Returns to the Maintenance entry menu
Prev
14.11. ID
Maint: ID
Prev
Next
Cont
Back
A
B
C
D
Proceeds to Cal Sensor menu
<Next>
Proceeds to Console menu
<Cont>
Enters ID menu
Cont
Back
<Prev>
Proceeds to Console selection
<Next>
Proceeds to Gravimetric selection
<Cont>
Enters Security menu
<Back> Prompts for Maintenance exit
The machine ID is a, user selectable, five digit
number between 00000 & 30000 which is included
in the result strings output to the serial port.
<Prev>
Next
The security feature is available to prevent
unauthorized or inadvertent access to the Setup
Menu during normal operation.
This prevents
changes to the alarm values or critical load and
sample times.
<Back> Prompts for Maintenance exit
37
TDA-110P
Air Techniques International
Enable Security
Yes
No
<Yes> Enables security and restricts operator
access to “Setup” parameter menu.
<No> Disables security and allows full operator
access to “Setup” parameter menu.
The unit will immediately return to the Maintenance
entry menu upon selection of either Yes or No.
14.14. Setting the Machine ID
This screen is used to set the machine ID. This
ID is a four digit number which is included in the
result strings output to the serial port.
ID 9999
UP
9999
<UP>
<DWN>
<EXIT>
DWN
EXIT
Current machine ID number
Increase the number
Decrease the number
Set the number and return to the
Maintenance Menu
38
1800103 Rev C
1800103 Rev C
Air Techniques Incorporated
TDA-110P
15. Console Commands
The 100P provides a mechanism to issue
commands and set and recall operating parameters
through the RS-232 serial communications port.
This sections documents the available functions.
port. The prompt string is “Ready?” Characters
sent to the serial port prior to the receipt of the
prompt will be ignored. Each command sent to the
serial port must be terminated by a carriage return
character (0D in hexidecimal).
The console mode is initiated in one of two
ways:
1)
Through the Maintenance Mode
Security/Console Menu (Section 14.2).
2)
By request through the serial port. If an
ASCII ESC (Escape, hexidecimal 1B)
character is received while the TDA100P is idle, the machine will issue a
“READY?” prompt and begin console
mode. If the machine is busy when the
ESC character is received, the character
will be held until the machine is idle.
The available commands are: VER, END,
DUMP, LOAD, SET, PROMPT, PENCAL and END.
The function of each command is described below.
In the descriptions references to “PPPPP” refers to
one of the operating parameters listed in the 100P
Parameter Table below. Also, “VVVVV” represents
the value of the parameter or the value to be
assigned to the parameter. The values can take
several forms depending on the parameter. The
forms are:
When the 100P is ready to accept a new
command, it will issue a prompt string to the serial
Integer
a sequence of decimal digits. For
example, “150”.
Real
a sequence of decimal digits with an
optional decimal point. For example,
“2.3”.
True/False A true value is entered as a “Y” (for
“yes”) or an “N” (for “no”).
TDA-100P Serial Port Commands
VER
Displays the version of the software.
END
Terminates console entry mode and returned control to the 100P front panel buttons.
DUMP
Displays the value of the 100P’s operating parameters in the form PPPPP=VVVVV.
Each parameter is displayed on a line by itself followed by a carriage return and line feed
(hexidecimal 0D and 0A, respectively).
LOAD
Accepts a series of operating parameters in the same form as that produced by the DUMP
command.
SET
Used to set the value of an individual operating parameter.
The form of the command is “SET PPPPP=VVVVV”. No additional spaces are allowed.
PROMPT
This command will cause the TDA-100P to ask for changes to all of the parameters. For each
parameter, the 100P will display “PPPPP[VVVVV]=”, where the PPPPP is the parameter name and
VVVVV is the current value of the parameter. If the user wishes to leave the parameter unchanged,
the he can enter a carriage return. If he wishes to change the value, enter the new value of the
parameter followed by a carriage return.
PENCAL
Initiates an LSC calibration cycle.
39
TDA-110P
Air Techniques International
1800103 Rev C
16. LSC Maintenance
16.1. Photomultiplier Tube
Replacement
A. Loosen the wing nut on the top of the
Photomultiplier Tube Elbow (see Figure 4) and
lift out the Phototube housing.
B. Remove and replace the phototube.
C. Insert the housing back into the elbow.
Figure 10
NOTE: Make sure that when separating these two
pieces, the Light Barrel is on the bottom and the
Cone is lifted upward (See Figure 6).
WARNING – Failure to do this will cause the
lens to fall out damaging it.
Figure 9 - Phototube
D. Turn on the unit power and enter the Main
Maintenance menu LSC Check (Section 5.3).
Turn off the aerosol to obtain the low reading.
E. Rotate the phototube housing until the highest
reading is achieved.
F. Tighten the wing nut.
NOTE: Follow the aligning procedure if any other
changes were made to the LSC assembly.
16.2. Cleaning
A. Remove the power from the unit.
B. Remove the LSC from the unit.
Figure 11
NOTE: Do not remove the Light Assembly screw or
loosen the Light End Plate screws during cleaning.
C. Remove the three thumbnuts and the
Photomultiplier Tube Elbow (See Figure 5).
D. Remove the three screws securing the Lens
End Plate to the Sample Cone.
E. Remove the three screws securing the Light
Barrel to the Sample Cone.
F. Wipe the outer lens on the Light Barrel with a
clean, lint free cloth to eliminate all foreign
matter.
NOTE: Do not disturb the rubber aperture.
G. Wipe the lens and Lens End Plate with a clean,
lint free cloth to eliminate all foreign matter.
H. Inspect the inside of the Sample Cone and
remove any buildup or foreign matter.
NOTE: Any lint or foreign matter left in the Cone or
on the lenses will decrease the sensitivity of the
LSC.
I. Reverse this procedure to reassemble the LSC
and to reinstall it into the unit.
40
1800103 Rev C
Air Techniques Incorporated
16.3. Aligning
A. Turn on the unit power and enter the Main
B.
Maintenance menu LSC Check. Turn off the
aerosol to obtain the low reading.
If the Light Assembly was removed, loosen and
move in and out until the lowest reading is
displayed. Tighten the screw.
NOTE: If the Light Assembly is started from the full
in position and pulled out, the reading will start high,
drop, then increase and drop again when the
assembly is in the out position. The low point of the
first drop is the correct setting.
C. If the Light End Plate was moved, loosen the
D.
thumbscrews and realign the plate even with
the barrel. Tighten the screws.
Loosen the thumbscrews on the Photomultiplier
Tube Elbow and adjust for the lowest reading.
Tighten the screws.
Figure 12
E. Repeat the Light Assembly adjustment
procedure for the lowest reading.
41
TDA-110P
Appendix A.
Dip Switch Settings
Switch Number
Definition
Function
1
Not used
2
External Air Chuck Control
3
Not used
4
Remote Mode
5
Chuck Close Audible
6
Chuck Latched
Not Used
Off–Chuck close air controlled by unit (Default)
On – Chuck close air controlled externally. Unit
ignores position sensor inputs
Not Used
Off – Normal operation (Default)
On – Remote operation. Unit bypasses normal
operator prompts.
Off – No Audible (Default)
On – Audible beep when chuck starts to close
Off – No audible (Default)
On – Audible beep when chuck is closed
7
8
8)
Not Used
Not Used
Alarm Prompt
Off – Disables user input requirement (Default)
On – Requires user input to clear alarm state
1800103 Rev C
Appendix B.
Air Techniques Incorporated
TDA-110P
TDA-110P Operating Parameters
UNIT_ID
A numeric identifier that can be used to distinguish one TDA-110P from another. This number
is included in the result output string reported at the serial port.
PENMAX
Maximum allowed penetration during a test. An alarm will be raised if this limit is exceeded.
PENMAX_ON
“Y” if maximum penetration limit is enabled; if “N” the upper limit will not be tested.
PENMIN
Minimum allowed penetration during a test. An alarm will be raised if the sample penetration
falls below this limit.
PEN_LO_ON
“Y” if minimum penetration limit is enabled; if “N” the upper limit will not be tested.
RESMAX
Maximum allowed resistance during a test. An alarm will be raised if this limit is exceeded.
RESMAX_ON
“Y” if maximum resistance limit is enabled; if “N” the upper limit will not be tested.
RESMIN
Minimum allowed resistance during a test. An alarm will be raised if the sample resistance falls
below this limit.
RESMIN_ON
“Y” if minimum resistance limit is enabled; if “N” the upper limit will not be tested.
FLOW_TOL
Allowable variation tolerance of the sample flow during testing in LPM.
FLOW_TOL_ON
“Y” if flow tolerance limit is enabled; if “N” the limits will not be tested.
CYC_TO_CAL
Number of test cycles performed between mandatory recalibrations.
CYC_TO_CAL_ON “Y” if the mandatory calibration based on test count is enforced.
LOAD_TIME
Delay time (in seconds) allowed for aerosol loading and stabilization.
LOAD_TIME_ON
“Y” if load time will be used. If “N”, load time will be ignored.
SAMP_TIME
Sample testing time (in seconds).
SAMP_TIME_ON
“Y” if sample testing is limited to the value of “SAMP_TIME”. If “N”, testing will run continuously
until an alarm limit is exceeded or the chuck is opened by the operator.
EXT_LOAD_TIME
Extended load time (in seconds). Extended load time is used on the first test after calibration
and after a limit error.
Parameters below this line will only appear if the trace level (TLEVEL) parameter is > 1 (one). These
parameters are normally set during the calibration process and should not be changed manually without a
thorough understanding of the impact on test results.
RES_OFF
Provides an end-user feature to shift the displayed pressure differential by a fixed amount due
to sample path variations occurring in non-encapsulating test fixtures. Default setting is 0.
RES_OFF_ON
“N” disables the resistance offset. Default setting is N (Disabled).
ZERO_OFF
An offset in Analog to Digital converter counts that is be applied to the LSC reading, at each
test cycle, during calculation of the % Penetration. Default setting of -120.
ZERO_OFF_ON
“N” disables the zero offset. Default setting is N (Disabled)
CAL_100
100% LSC Reading from gravimetric calibration
FLOWCON
Flow calibration slope constant
FLOWZERO
Zero flow output level for flow sensor (in counts)
GRAVCON
LSC slope constant created during gravimetric calibration
GRAVHI
Upper response limit of the 100% penetration reading (set during gravimetric routine).
GRAVLO
Lower response limit of the 100% penetration reading (set during gravimetric routine).
LSC_DAC
LSC DAC (Gain) Setting
RESCON
Pressure (resistance) calibration slope constant.
TLEVEL
Trace Level determines the amount of diagnostic information displayed through the serial port.
Available levels are 0, 1, and 3, representing minimal, normal and maximum amounts of output.
Default value of 0 to match legacy output content.
QZT
Quick Zero Tolerance is a factor applied to the LSC reading returned at the start of each test
cycle to determine if the LSC state has recovered from the previous test. 0=Off. Default value
of 6.
For operating revisions ≥1.17
BPSENSTYPE
Specifies the type of barometric pressure sensor installed.
0=None (Default), 1 = MPX4105A, 2= MPX5100A
FLOWSENSTYPE
Specifies the mass flow meter installed.
A=Alicat (Default), T=Teledyne Hastings
For operating revisions ≥2.0
ALT100PCT
“N” disables the non-encapsulating feature (Default). Default setting is N (Disabled). “Y”
enables non-encapsulating feature utilizing an optional dedicated valve for obtaining 100%
during daily ‘PenCal’ routine.
43
TDA-110P
Air Techniques International
Appendix C.
1800103 Rev C
Default Settings
The table below lists the default values for the TDA-110P operating parameters. In the
event of a failure in the internal parameter memory, the parameters will be reset to these
defaults.
Default Parameter
Pressure unit
Flow Calculation
Pen Max
Pen Min
Max Resistance
Min Resistance
Flow Tolerance
Load Time
Extended Load Time
Sample Time
Open Time
Cycles to Cal
Con Check
Default Value
mm/H2O
STD
0.0299
0.0005
25.0
2.0
2.0
8.0
12.0
1.0
0.0
1000
NA
44
Unit
mm/H2O
NA
Percentage
Percentage
mm of H2O
mm of H2O
CFM
Seconds
Seconds
Seconds
Seconds
Counts
mg/m3 or g/l
Enabled Y/N
NA
NA
Y
N
Y
N
Y
Y
Y
Y
Y
Y
N
1800103 Rev C
Appendix D.
Air Techniques Incorporated
TDA-110P
Fault and Error Messages
Error Message
Problem Fault
Corrective Action
**ALL VALUES
RESET**
This message occurs if the memory used to retain system
operating parameters has been corrupted and lost. This message
will only occur during the first machine initialization immediately
following memory loss.
Press <CONT> to continue.
You will automatically enter the
maintenance mode and must
complete all calibrations before
the machine is operational.
Cal Penetration?
This message occurs when the selected number of tests
between calibrations has been reached. (See “Cycles to Cal” for
information about this setting.) This prompt allows the user to enter
the penetration calibration procedure or to skip the calibration.
Whether the calibration is performed or not, this prompt will not
reappear until the number of tests between calibrations count is
reached again.
Calibration
Required
Indicates that one or more sensors (flow, pressure, or LSC)
must be calibrated.
Enter <YES> to enter the
penetration calibration
procedure (See section 11.7.)
Enter <NO> to return to the
Start Menu (Section 8.3) without
calibration.
Press <CONT> to continue.
You will automatically enter the
maintenance mode and must
complete all calibrations before
the machine is operational.
Pressure sensor needs to be
calibrated.
If problem persists, contact
manufacturer
Run system gravimetric test
If problem persists, contact
manufacturer
Flow sensor needs to be
calibrated.
If problem persists, contact
manufacturer
PRES SENSOR
- RUN MAINTEN
– RESET
Memory Error. Memory storage corrupted or missing
GRAVIMETRIC
- RUN MAINTEN
– RESET
Memory Error. Memory storage corrupted or missing
FLOW SENSOR
- RUN MAINTEN
– RESET
Memory Error. Memory storage corrupted or missing
CANCEL
DURING TEST
Occurs when the chuck is opened without completion of the test
or setup
Press <CONT> to return to
the main menu of the canceled
function
Chuck
Obstructed Error
The Chuck Obstructed Error occurs when the chuck close
switches are not pressed simultaneously and held until the chuck is
completely closed.
Press <CONT> to return to
the previous menu
LSC PROBLEM
- RUN MAINTEN
– RESET
No LSC output or LSC output less than 30% of initial reading
Check LSC LED
CON SYSTEM
ERROR
- RUN MAINTEN
– RESET
Penetration reading exceeded maximum available span
CON TO HI
- RUN MAINTEN
– RESET
Concentration check higher than Hi setting from gravimetric test
45
Check generator level
Check drip jar
Check for proper dilution air
Check nozzle pressure
Clean LSC and run gravimetric
test
Check generator level
Check drip jar
Check for proper dilution air
Check nozzle pressure
Clean LSC and run gravimetric
test
TDA-110P
Air Techniques International
1800103 Rev C
CON TO LOW
- RUN MAINTEN
– RESET
Concentration check lower than Low setting from gravimetric test
Check generator level
Check drip jar
Check for proper dilution air
Check nozzle pressure
Clean LSC and run gravimetric
test
CHECK
ORIFICE
TURN MAIN
POWER OFF
Generator tank pressure too high
Remove generator and check
impactor orifice for obstruction
CHECK FLOW
CTRL
TURN POWER
OFF
System flow is outside of set flow tolerance
Check vacuum pressure
Readjust flow rate
Check test flow filters
The system has lost vacuum or signal from the mass flow meter
Check vacuum system
Check mass flow meter cabling
CHECK
VACUUM
TURN MAIN
POWER OFF
SYSTEM
ERROR
TURN POWER
OFF
CHECK
SYSTEM AIR
TURN MAIN
POWER OFF
General system fault
System pressure switch has not actuated.
46
Turn power off, wait two minutes
and turn power on
If problem persists contact
manufacturer
Verify adequate pressure supply
of 10 cfm @ 80 psig to unit.
Verify incoming compressed air
supply line has a min 0.375” ID
(9.5 mm).
Check system pressure switch
operation.
Appendix E.
Remote Mode Operation
In remote mode, the TDA110P can be controlled from an external computer through a combination of the
user interface port and commands sent and received through the serial port. The remote mode is selected by
DIP switch 2 (two) and 4 (four) (See Appendix A). The program trace level (TLEVEL parameter) should also be
set to level 1 for remote operation (See Appendix B)
When the machine is restarted in this mode:
1) The machine will wait for air.
2) The check level prompt is bypassed.
3) The system warm-up will proceed.
4) Personnel will be required to either enter or bypass barometric pressure and ambient temperature
selection.
5) The message “Insert Filter; and Close” will print on the VF display. Personnel will be required to
close the test fixture locally or through an HMI, and an external start signal sent through the user
interface port.
6) The sample flow rate is then adjusted using the “Flow Adjustment “ valve located on the front panel
and accepted using the ‘D’ function key on the control panel.
7) The machine will enter an LSC calibration cycle. In the calibration cycle:
a. The machine will remain busy.
b. The message “!Close Chuck” will be sent to the serial port, while the message “Clear
Chuck & Close” will print on the VF display. In response to this, the chuck should be
closed, and an external start signal sent to the machine through the user interface port.
c.
Upon receipt of the external start, a 100% calibration will be performed.
d. At the end of the 100% calibration, two messages will be sent to the serial port. The first:
“ PENCAL: LSC_100% = nnnnnn” will display the penetration calibration constants. The
second: “PENCAL: Res Tare = nnnnnn” will display the resistance tare value.
The numbers in the above strings will not occupy a fixed number of digits; they will be as
large as needed to represent the values.
e. A 0 (zero) baseline routine will be performed with no external intervention required.
f.
At the end of the calibration process, the “Ready” contact at the User Interface port will
become active.
Note: The calibration beginning from step 7 will also be entered if a “PENCAL” command is sent to the
RS-232 port using the console command protocol (See) or the number of cycles specified by the
“Cycles to Cal” variable is exceeded.
Other operational features of the machine can be controlled through parameter setting commands available
as described in the Console Commands section of this manual (See).
The RS-232 printed statements listed above are the minimum that will be printed during these stages of unit
operation. As greater levels of program trace output are enabled there may be significantly more data
available at the RS-232 port.
Appendix F.
USER INTERFACE Circuit Example
Appendix G.
4B Lite Aerosol Generator
Generator Description
The generator type used on the ATI 110P Penetrometer is based on a modified Laskin nozzle
used in conjunction with a virtual impactor. The particle size distribution produced by this
combination is characterized as 0.185 CMD ± 0.020 micron with a standard deviation of <
1.6.
Generator Maintenance
1. If clean, dry, compressed air is used with this unit, little maintenance should be
required.
2. Inspect and drain the compressed air filter/regulator daily, or more often, as required.
3. Yearly, under daily operation, drain all liquid and flush with a solvent to remove any
residue from the unit.
BEFORE SHIPPING GENERATOR UNIT
1. Drain all liquid from unit.
2. Verify that the LIQUID FILL cap is tight.
3. Block aerosol outlet flange with liquid-absorbing cloth or paper to prevent oil from
damaging shipping container.
4. Tape or plug the compressed air inlet on air filter/regulator to prevent internal damage
by foreign material.
Package the unit in a triple wall carton with a minimum of 3 inches of loose packing fill on all
sides.
49
Appendix H.
PAO-4 Safety Data Sheet
MATERIAL SAFETY DATA SHEET
1. Identification of the Substance/ Mixture and of the Company/ Undertaking
Product Name: ATI PAO-4
Preparation Date: 11/21/2000
Revision Date: 06/26/2012
Recommended Use: Particle filter testing
Supplier: Air Techniques International
11403 Cronridge Drive
Owings Mills, MD 21117-2247
Telephone: 1 (410) 363-9696
Emergency Telephone Number: Chemtrec (USA) 1-800-424-9300
2. Hazard Identification
GHS Classification: Aspiration Hazard: Category 1
H304 – May be fatal if swallowed and enters airways.
GHS Labeling:
Symbol:
Signal Word: Danger
Precautionary Statements: P301 + P310: IF SWALLOWED: Immediately call a POISON
CENTER OR doctor/physician
P331: Do NOT induce vomiting
P405: Store locked up
P501: Dispose of contents and container in accordance with local regulations
Other Hazards Not Classified: No significant hazards
50
3. Composition/Information on Ingredients
Chemical Name
1-Decene, homopolymer,
hydrogenated or 1Decene, tetramer mixed
with 1-decene
CAS Number
Concentration, Wt. %
68037-01-4
or
68649-12-7
100%
4. First Aid
Inhalation: If inhaled, move to fresh air. If victim has stopped breathing give artificial
respiration, preferably, mouth to mouth. Contact a physician immediately.
Eyes:
Flush with large amounts of cold water for at least 15 minutes. Do not let victim
rub eyes. If irritation develops, contact a physician immediately.
Ingestion:
Do not induce vomiting. If victim is conscious and able to swallow, promptly
have victim drink water to dilute. Do not give sodium bicarbonate, fruit juices or
vinegar. Never give anything by mouth if victim is unconscious or having
convulsions. Contact a physician immediately.
Skin:
Wash affected area with soap and water. Remove contaminated clothing.
Launder contaminated clothing before re-use.
Most important symptoms and hazardous effects:
Local necrosis as evidenced by delayed onset of pain and tissue damage a few
hours after exposure.
Indication of immediate medical attention and special treatment needed:
If ingested, material may be aspirated into the lungs and cause chemical
pneumonitis. Treat appropriately.
5. Fire-fighting Measures
Suitable extinguishing media: Carbon dioxide, Dry chemical, Foam, Water spray
Specific hazards: Smoke, fumes and incomplete combustion products.
Specific protective equipment and precautions for fire fighters:
Use water spray, dry chemical, foam or carbon dioxide. Water may be
ineffective but should be used to keep fire exposed containers cool. If a spill or
a leak has not ignited, use water spray to disperse the vapors. Water spray may
be used to flush spills away from fire.
Perform only those firefighting procedures for which you have been trained.
Firefighters should wear self-contained breathing apparatus in the positive
pressure mode with a full face piece where there is a possibility of exposure to
smoke, fumes or hazardous decomposition products.
51
Flammability Properties:
Flash point: 222oC (432oF)
Auto ignition Temperature: 343oC (649oF)
6. Accidental Release Measures
Personal precautions, protective equipment and emergency procedures:
Use personal protective equipment. Ensure adequate ventilation.
Environmental Precautions:
Do not allow spilled material to enter sewers or streams.
If spills are likely to enter any drain, waterway or groundwater, contact the
appropriate governmental agency.
Methods and materials for containment:
Add dry material to absorb (if large spill, dike to contain). Using recommended
protective equipment, pick up bulk of spill and containerize for recovery or
disposal. Flush area with water to remove residues.
7. Handling and Storage
Precautions for safe handling:
Read label for instructions in use of product. Prevent small spills and leakage to
avoid slip hazard. Material can accumulate static charges which may cause an
electrical spark (ignition source).
Conditions for safe storage:
Store in closed containers in a cool, dry well ventilated area. Maintain closure of
bungs. Store at temperatures between 5oC and 50oC. Do not reuse container.
Avoid container damage while storing.
Empty containers retain residue (liquid and/or vapor) and can be dangerous. Do
not pressurize, cut, weld, bronze, solder, drill, grind or expose such containers
to heat, flame, sparks, static electricity or other sources of ignition; they may
explode and cause injury or death. Do not attempt to refill containers since
residue is difficult to remove. Empty drums should be completely drained,
properly bunged and returned to a drum re-conditioner. All containers should be
disposed of in an environmentally safe manner in accordance with
governmental regulations.
8. Exposure Controls/Personal Protection
Control parameters: 1-decene, homopolymer, hydrogenated CAS # 68037-01-4;
TWA 5 mg/m3 (ExxonMobil)
For mist and aerosols: 5 mg/m3ACGIH TLV; 10 mg/m3 ACGIH STEL
52
Appropriate engineering controls:
Proper protection and controls is dependent upon the potential exposure
conditions. No special requirements are needed under ordinary conditions
where adequate ventilation is available.
Individual protective measures:
Respiratory protection: Needed when airborne contaminant concentrations are
at a level which cannot protect worker health. Then an approved respirator must
be used. Selection of the respirator is dependent upon regulatory conditions.
For high airborne concentrations, use an approved supplied-air respirator,
operated in positive pressure mode.
Eye protection: No eye protection is needed under conditions of normal use. If
there is a possibility that the product can be splashed into the eyes, then safety
glasses with side shields or chemical goggles are required. Contact lenses
should also not be worn if the product could be splashed into the eyes.
Hand protection: No gloves are required for single, short duration exposures.
For prolonged or repeated exposures, wear rubber gloves.
Body protection: If product use involves single, short duration exposures, then
no additional protective wear for covering the skin is required. For prolonged or
repeated exposures to the skin, wear impervious, protective clothing including
rubber safety shoes to avoid skin contact.
9. Physical and Chemical Properties
Appearance: Colorless liquid
Odor: Odorless
Odor Threshold: N/A
pH : N/A
Melting point/freezing point: < -20oC
Initial boiling point and boiling range: > 316oC
Flash Point (Method): 222oC (Cleveland Open Cup)
Evaporation Rate: N/A
Flammability (Solid, Gas): N/A
Upper/lower flammability or explosive limits: UEL: No data available; LEL: No data available
Vapor pressure: <0.013 kPa (0.1 mm Hg) at 20oC
Vapor density: N/A
Relative density: 0.82 @ 15.5oC
Partition coefficient n-octanol/water: N/A
Auto ignition Temperature: No data available
Decomposition Temperature: N/A
Viscosity: 18 cSt at 40oC/ 4 cSt at 100oC
53
10. Stability and Reactivity
Chemical stability: Stable at normal conditions
Possibility of hazardous reactions: Not expected and hazardous polymerization will not
occur
Conditions to avoid: Excessive Heat. High energy sources of ignition.
Incompatible Materials: Strong acids, bases and oxidizing agents.
Hazardous decomposition products: Carbon dioxide and carbon monoxide
11. Toxicological Information
Inhalation Toxicity: Practically non-irritating. LC50 > 5 mg/l Based on testing of similar
products
Oral Toxicity (Rabbits): LD 50 > 15 g/kg Practically non-irritating
Skin Irritation (Rabbits): Practically non-irritating. Primary irritation score 1. (Scale 0-8)
Dermal Toxicity (Rabbits): LD50 > 5 g/kg Practically non-irritating
Eye Irritation (Rabbits): Practically non-irritating. Eye irritation score: 2.0 at 1 hour, 2.0 at
24 hours (Scale 0-110)
12. Ecological Information
Aquatic/terrestrial ecological toxicity:
Toxicity to fish:
LC 50 > 1,000 mg/liter (96 hours)
Species
Rainbow Trout
Toxicity to daphnia: EC 50: 190 mg/liter (48 hours)
Species
Daphnia
Mobility:
Not established
Persistence and degradability:
Biodegradability of this material is greater than or to 20%. This material is
inherently biodegradable.
13. Disposal Considerations
Disposal methods:
Product can be disposed of by burning in an enclosed, controlled burner for fuel
value or disposal by supervised incineration. Such burning may be limited by
the controlling authority. In addition, the product is suitable for processing by an
approved recycling facility or can be disposed of at any licensed waste disposal
site.
Precaution for disposal:
All recovered material should be packaged, labeled, transported and disposed
or reclaimed in conformance with Good Engineering Practices. Comply with all
applicable governmental regulations. Avoid land filling of liquids. Reclaim
where possible.
54
14. Transport Information
USA DOT: Not designated as a hazardous material by the USA DOT
RID/ADR: Not regulated by RID/ADR
IMO: Not regulated by IMO
IATA: Not regulated by IATA
15. Regulatory Information
Europe REACH: In compliance with the inventory
EU Labeling:
Not required
Regulations in other countries:
Product is listed on the database of the following countries: US, Canada, Australia, New
Zealand, Japan, China, Korea and the Philippines.
16. Other Information
References and Sources:
Information contained in this safety data sheet is based on Air Techniques International
owned data and public sources deemed valid or acceptable. The absence of data
elements required by ANSI or 2001/58/EC indicates, that no data meeting these
requirements is available.
Disclaimer:
This document has been prepared in good faith and from information provided to us by our
suppliers and other sources considered to be reliable. No warranty, express or implied is
given. The buyer is responsible to evaluate all available information when using this
product for any particular use. The buyer is also responsible for complying with all Federal,
State, Provincial, and Local Laws and regulations when using this product.
55
Appendix I.
Blower Motor
The blower motor manufacturer’s installation, maintenance and operating instructions
are included in this addendum. Please follow the manufacturer’s recommended
instructions for service and operation.
56
Appendix J.
Flow Sensor Operation and Service Manual
The flow sensor manufacturer’s installation, operating and service instructions are
included in this addendum. Please follow the manufacturer’s recommended
instructions for service and operation.
Appendix K.
Manual Revision
Base Manual version
Limited release firmware manual
Updated to include operating code &
design revisions incurred since 2007
Rev B release
Date
Revision
September 2006
September 2007
A
B
December 2012
C
58