MANUAL OPERATIONS
3.3.8 PRINT NUMBER
Print NO = 0
Range : 0 - 9999
The Print NO will start from 0 when the tester is turned On. The number will
increase by 1 when “TEST” button is pressed until 9999 and loop back to 0. The
user can set the starting Print NO by entering the number in this mode. Enter the
starting Print NO and press “ENTER” key. When the tester is turned OFF, the
Print NO will be reset to 0.
When the “SETUP” key is pressed, the display will loop back to 3.3.1 PLC
REMOTE SELECTION or press “EXIT” to exit parameter setting mode.
3.4 DISPLAYED MESSAGES
3.4.1 AC WITHSTAND VOLTAGE TEST:
3.4.1.1 Abort
3.4.1.1.1 If the test in process is Aborted with the “RESET” button or remote control, the
display will show:
ACW Abort
XXX.X s
MXX-X X.XX KV XX.XX mA
3.4.1.1.2 If the test in process is Aborted with the “RESET” button or remote control before
the meter readings are taken, the display will show:
ACW Abort
MXX-X - . - - KV
XXX.X s
- - . - - mA
or
ACW Abort
XXX.X s
MXX-X X.XX KV - - . - - mA
3.4.1.2 Ramp
3.4.1.2.1 At the beginning of AC Withstand Voltage test when the voltage begins to ramp but
before the meter readings are taken, the display will show:
ACW Ramp
MXX-X - . - - KV
XXX.X s
- - . - - mA
3.4.1.2.2 During the AC Withstand Voltage test when the values are being updated in real time
during the ramp cycle, the display will show:
ACW Ramp
XXX.X s
MXX-X X.XX KV XX.XX mA