Download Logiq 9 - internetMED

Transcript
GE MEDICAL SYSTEMS
DIRECTION 2294853-100, REVISION 5
Section 8-10
8-10-7
LOGIQ™ 9 SERVICE MANUAL
System Setup (cont’d)
Standby Operation Checkout
To place the system in the Standby mode:
-
Press the On/Off Switch, select “Standby” and select “OK”.
-
Wait approximately 30 seconds. The system should be in Standby mode. Both display
screens are black and the On/Off switch indicator is green.
To wake the system from Standby mode:
-
Press On/Off switch QUICKLY! Less than .25 seconds. If you press the On/Off switch too long
the system will shutdown. It takes about 45 seconds for the system to become fully functional.
NOTE:
Waking the system from Standby without AC connected will turn off the system.
NOTE:
Standby mode is good for approximately 12-15 minutes.
The system will “beep” while in Standby Mode if AC power is removed (unplugged).
NOTE:
Charge time for batteries is approximately 4 hours.
8-10-8
Functional Checkout
Refer to the customers “Quick Guide” to preform a functional checkout.
Table 8-1
Quick Guide Direction Numbers
Application Software Version
Direction Number
R2.X
2309820-100
R3.X
2384458-100
R4.X
5113779-100
Chapter 8 Replacement Procedures
8-35