Download Operating And Service Manual Ramsey MINI CK100 Weight Indicator
Transcript
3.10.3.4 Load Cell Resistance The resistance value of the bridge of each load cell has to be entered here. The number of scroll depends on the number of load cells of the system specified on the MENU 3. --SC DATA SCROLL 9A Resistance Load cell N # 350.000 Ohms Password:Service X = A, B, C, D, E, o F N = 1, 2, 3, 4, 5, o 6 Default: 350 Ohms Min: 10 Ohms Max: 2000 Ohms Note:. For each load cell must be inserted the value on the corresponding screen. 3.10.4 Stable Weight These Scrolls define the parameters for the stable weight indication. Motion band defines the range on which the weight should stay in order to be considered stable. -SC DATA SCROLL 11A -Motion Band division 1 Password:Service ENTER Default: 1 Min: 0 Max: 3 3.10.5 Stable Weight Delay The motion delay defines how many times this condition should be true before stable weight indication turns on. -- SC DATA SCROLL 11B Motion delay 1.0 sec ENTER Default: 1 Min: 0 Max: 60 3-24 Password:Service