Download Leica ASP300 S
Transcript
7. Cleaning and Maintenance Draining the condensate bottle • Remove and empty the blue condensate bottle and reinsert (Fig. 77). Dispose of waste solvents with care according to local regulations and the waste management policy of the company or institution. 14 Fig. 77 Exchanging the activated carbon filter • The life of the carbon filter cartridge (13) will depend on the reagent types used and the frequency of vacuum cycles (which expel air from the instrument through the filter). • In general, carbon filter cartridges should be replaced after approx. 3 months. (Fig. 78) After replacing the filter press open the menu SMART MONITOR to reset the filter age to "0". See also chapter 5.5. • Prior to changing the filter we recommend a smart clean. 13 Filter label Fig. 78 When inserting the new carbon filter, make sure the arrow on the filter label points up. 70 Operating Manual V 1.2 – 01/2008