Download DMF - Dwyer Instruments, Inc.
Transcript
9. CALIBRATION CONVERSIONS FROM REFERENCE GASES The calibration conversion incorporates the K factor. The K factor is derived from gas density and coefficient of specific heat. For diatomic gases: 1 d X Cp where d = gas density (gram/liter) Cp = coefficient of specific heat (cal/gram) K gas = Note in the above relationship that d and Cp are usually chosen at standard conditions of one atmosphere and 25F C. If the flow range of a Mass Flow Controller or Controller remains unchanged, a relative K factor is used to relate the calibration of the actual gas to the reference gas. K = where Qa Qr Ka Kr = = = = Qa Qr = Ka Kr mass flow rate of an actual gas (sccm) mass flow rate of a reference gas (sccm) K factor of an actual gas K factor of a reference gas For example, if we want to know the flow rate of oxygen and wish to calibrate with nitrogen at 1000 SCCM, the flow rate of oxygen is: QO = Qa = Qr x K = 1000 X 0.9926 = 992.6 sccm 2 where K = relative K factor to reference gas (oxygen to nitrogen) 20