Download 1 - Cerner Learning Manager
Transcript
sec09.fm Page 5 Thursday, July 14, 2005 3:49 PM Plum A+® Infusion Systems Time To Detect Downstream Occlusions RATE OCCLUSION ALARM PRESSURE SETTING TYPICAL TIME TO ALARM (MACRO) MAXIMUM TIME TO ALARM (MACRO) 25 mL/hr 6 psig 38 seconds 47 seconds 15 psig 98 seconds 106 seconds 6 psig 17 minutes 22 minutes 15 psig 45 minutes 49 minutes 1 mL/hr * Baseline backpressure is 0 psig * Bolus Volume Released After Downstream Occlusions Are Corrected RATE OCCLUSION ALARM PRESSURE SETTING TYPICAL BOLUS VOLUME (MACRO) MAXIMUM BOLUS VOLUME (MACRO) 25 mL/hr 6 psig 0.23 mL 0.28 mL 15 psig 0.55 mL 0.63 mL 6 psig 0.24 mL 0.30 mL 15 psig 0.59 mL 0.70 mL 1 mL/hr * Baseline backpressure is 0 psig * 430-95597-001 (Rev. 06/05) 9- 5