Transcript
UC Line User's Guide Power Requirements The UC requires 100-240 VAC, 50/60 Hz, 2.2A power. Wiring must include a proper ground. The integrity of the power ground for equipment is important for both safety and dependable operation of the electronic device. A poor ground can result in an unsafe condition should an electrical short develop in the equipment. A good ground connection minimizes extraneous electrical noise pulses. The electronic device should not share power lines with noise-generating equipment. Examples include electrical motors, compressors, heating elements, lights, and relays that may generate noise spikes and power surges. To confirm ground integrity, use a commercial branch circuit analyzer or grounded outlet tester. If adverse power conditions exist, a dedicated power circuit or power line conditioner might be required. Network Connections The UC line can connect to an Ethernet network using a cable or using a wireless radio. If you use a wired network connection and experience connection problems, make sure your Ethernet cable is connected and seated securely in the jack, located under the touchscreen at the front of the scale or behind the rear door on the UC-LP. If you are connected using a wireless radio, you will see a cover protruding from the left side of the counter scale, the right side of the tower scales (UC-GT/UC-hTT), or the UC-LP will have external antennas (see below). UC-LP Wireless Antennas UC-CWx/UC-GT/UC-hTT Cover for Wireless Radio Ethernet network connection (shown on a counter scale) UC-LP (behind door) RJ45 Ethernet Jack for wired network connections. 1-4