Warning messages and malfunctions
6.3 Power failure
In case of a power failure, the incubator operates as follows:
In normal and week time switch operating modes
After the power supply has been restored, operation is continued with the parameters set.
The time and the duration of the power failure is documented in the log memory.
For remote operation
If there is a power failure in remote operation, the appliance starts in the normal operating
mode for safety reasons and all setpoint values are changed to safe default values (see table).
The programme can only be continued from the computer. The time and duration of the
power failure are documented in the log memory.
Parameters
Default value
Temperature
20 °C
CO2
0%
rh
off
O2 (only for models with O2 module)
OFF