Download Technical Documentation Revision 05

Transcript
5 3.2.22
[_______OK]
Tubing systems as per Instructions for Use
The tubing systems as per the Instructions for Use are undamaged.
[OK]
5 3.2.23
[_______OK]
Dräger test lung (adults)
The adult test lung comprises the mask tube, ISO size 7 catheter connection socket and 2 L
breathing bag. The mask tube, the ISO size 7 catheter connection socket and the 2 L breathing
bag are complete and undamaged. (Note: The breathing bag must not be over-inflated or torn.)
[OK]
5 3.2.24
[_______OK]
SIEMENS test lung (adults)
The SIEMENS test lung is not cracked or porous and is undamaged.
[OK]
5 3.2.25
[_______OK]
Bellows K (neonatal)
The bellows K are undamaged.
[OK]
5 3.2.26
[_______OK]
Accessories for neonatal flow (option)
- Flow sensor cable
The flow sensor cable is undamaged. The socket on the flow sensor cable locks safely into the
flow sensor insert.
- Flow sensor ISO 15
The ISO 15 flow sensor and the flow sensor insert are undamaged.
- Water trap for expiratory valve (option)
The water trap for the expiratory valve is undamaged. The water trap for the expiratory valve must
be used with the optional ”neonate flow” feature.
[OK]
3.2.27
Drug nebuliser (white housing)
WARNING:
If other pneumatic drug nebulisers are used, significant variances in the breathing volume and the
inspiratory oxygen concentration may result!
INFO:
Do not connect the drug nebulizer with the "black" housing to the unit!
5 3.2.27.1
[_______OK]
Components of the drug nebuliser
Check the following components for damage and dirt contamination:
1 = Patient connection
2 = Drug nebulizer housing, white
3 = Flat seal
4 = O-ring
5 = Nozzle
6 = O-ring
7 = Atomizer
8 = Container
9 = Drug nebulizer tube
Evita XL
2005-07 Released
- page 13 -