SECTION 7—COMPRESSOR
3. Remove protective boot to reveal capacitor wires.
4. Disconnect the spade connectors from the capacitor terminals (not shown).
5. Remove the screw that secures the capacitor clamp to the soundbox brace.
6. Remove the capacitor from the concentrator base assembly.
7. Install new capacitor by reversing STEPS 3-6.
8. Re-install the cabinet. Refer to Removing Cabinet on page 26.
Protective
Boot
Capacitor Clamp
Capacitor
Screw
Soundbox
Brace
FIGURE 7.4 Replacing Capacitor
Rebuilding the Thomas Model 2650 and 2660
Compressor
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
This kit is designed for rebuilding the Thomas Model 2660 Series compressor. The serial
number is located on the front of the compressor. If the serial number does not begin in
this way, this kit will not work.
The 5-Liter Thomas 2660 Series Compressor Rebuild Kit contains the following:
DESCRIPTION QUANTITY
DESCRIPTION
QUANTITY
Teflon®Piston Cups
2
Aluminum Piston Sleeves
2
Flapper Valves - Intake and Exhaust
4
Valve Keepers
4
O-Ring - Sleeve
2
Part No 1118396
41
Platinum™Oxygen