CHAPTER 3 SEMIAUTOMATIC MODE Start with CPR Option Your AED Pro unit may be configured to start a rescue with a CPR period of 30 to 180 seconds that begins after you power on the AED Pro unit. You can end this initial CPR period at any time by pressing the Analyze softkey. When you press the Analyze softkey, the AED Pro unit immediately begins analyzing the patient’s ECG rhythm if the electrode pads are attached correctly. The Analyze softkey appears only during the initial CPR period and does not appear during any subsequent CPR periods unless it is configured to do so and/or the subsequent CPR intervals are set to Extended. Messages in Semiautomatic Mode During semiautomatic mode, the unit can issue the following voice prompts and text messages. The unit issues each voice prompt only once, but the equivalent message remains on the screen until you take action, time expires, or the device status changes. Message Description UNIT OK The unit successfully passed the power-on self-test. UNIT FAILED The unit failed the power-on self-test and is not usable for patient care. CHANGE BATTERY The self-test detected a low energy condition that is insufficient for patient care. Replace the battery pack immediately. ADULT PADS The unit detected the specified type of electrode pads and adjusted defibrillation energy settings accordingly. PEDIATRIC PADS PLUG IN CABLE The unit started up without an electrode cable plugged in. Plug the cable into the unit. ATTACH DEFIB PADS TO PATIENT’S BARE CHEST The unit does not detect attachment of the defibrillation electrode pads to the patient. Check the cable for damage. Make sure that the pads are properly applied to the patient. If this message continues, check the electrode sites to ensure that they are clean, dry, and free of excess hair. Check the expiration date on the electrode package. 3–12 CHECK PATIENT Check the patient for responsiveness or consciousness by gently shaking the patient and shouting “Are you all right?” CHECK PULSE Check the patient for a pulse or other signs of circulation, such as normal breathing, movement, or coughing. START CPR Indicates the beginning of a CPR period. Begin CPR. STOP CPR Indicates the end of the CPR period. Discontinue CPR. DON’T TOUCH PATIENT, ANALYZING Indicates the beginning of an ECG analysis period. Make sure everyone stands clear of the patient. Keep the patient motionless during ECG analysis. NO SHOCK ADVISED ECG analysis did not detect a shockable rhythm. SHOCK ADVISED ECG analysis detected a shockable rhythm that requires therapy. www.zoll.com 9650-0350-01 Rev. G