4.12 Leakage current test
4.12 Leakage current test
Leakage current test
Auxiliary devices required
● Line-frequency, high-resistance measurement voltage source
(isolation transformer)
● A measuring circuit (MD) that meets the requirements of IEC 60 6011.
● Voltmeter
Alternative
Complete test units, e.g. the "Bender tester", fulfill these requirements as
well.
Equivalent device leakage current measurement
båÖäáëÜ
Test assembly
1. Connect the measurement voltage source to the measuring circuit
(MD).
2. Connect the voltmeter to the measuring circuit (MD).
3. Connect the measuring circuit (MD) to the metallic part (B) of the wall
module housing.
4. Short-circuit the power connection (N and L) with a suitable cable.
62 15 102 D3507
D3507.076.01.07.02
08.2011