Download Finesse E+ and ME+ Operator Guide
Transcript
Note 1 2 The ‘ROCK’ and ‘RETRACT’ facilities cannot be used at the same time. When ‘WAX’ is selected as the chosen medium, the status of ‘ROCK’ and ‘RETRACT’ can be different for ‘TRIM’ and ‘SECTION’. To enable the rock facility, the rock position must be saved. Turn the Handwheel to move the specimen to the start, or top, of the cutting sector (a). Press [ROCK] on the Control Unit to save the start of the rocking sector (b). The green LED lights when the facility is selected (e). a ROCK Move the Handwheel so that the specimen passes across the knife and then back up again (c). b ROCK When the specimen is above the set rock position, the amber LED on the button (d) lights. The Specimen Head advances by the set distance. ROCK c Note 1 d If a Cutting Sector has previously been set, the Finesse will use the existing top position when [ROCK] is pressed. e To cancel the rock facility, press [ROCK] again. 4.7 TO USE THE MEMORY To help reposition similar sized specimens, press [MEMORY] on the Control Unit when the Specimen Head has advanced to the required position to start to trim. The green LED is lit to indicate that a position is saved. MEMORY AUTO If [AUTO] is then pressed, the Specimen Head rewinds a set distance back from the knife so that the specimen cassette can be removed. (The distance can be set in the Auto Load Options Menu section of User Setup, described in Appendix A, section A.3.6). A77510300GB Issue 3 51