Download PRISMA System
Transcript
Chapter 3: Continuous Renal Replacement Therapies (CRRT) 2. After the blood pump restarts (following an alarm or after pressing RESUME from the Stop screen). 3. After the operator presses the CONTINUE softkey from a pressure trending Advisory alarm screen. Pressure Trending Limits If the access or return pressure changes 50 mmHg negative or positive from its pressure operating point, the control unit notifies the operator by issuing an Advisory alarm, as shown in Figure 11. These alarms can be cleared by pressing the CONTINUE key on the alarm screen. This resets the pressure operating points to the current pressures in each pod. mmHg +50 Advisory alarm Advisory: Access Pressure is Rising Advisory: Return Too Positive Pressure Operating Point -50 Advisory: Access Too Negative Advisory: Return Pressure is Dropping Figure 11. Pressure Trending Limits, CRRT Therapies “Cannot Detect Disconnection” Limits If the access pod operating point is set more positive than -10 mmHg, or if the return pod operating point is set below +10 mmHg, a “Cannot Detect Disconnection” Advisory alarm occurs, as shown in Figure 12. The operator PRISMA System Operator’s Manual 69