Download service manual for BSC
Transcript
arms to prevent tiredness and shake during the experiment. 4.1.9 Wipe the cabinet and lower bracket with cloth and water or neutral household detergent after installation. 4.1.10 Note: the worktable and brackets for worktable must be taken out firstly if the biosafety cabinet has to be carried to the room in side dumping way in order to prevent destroying the ultraviolet lamp; 4.2 Installation step of B2 biosafety cabinet 4.2.1 Former installation steps are the same to A2 biosafety cabinet; 4.2.2 Installation of outward exhaust fan a. Cover the pipe clamp on one end of the outward exhaust pipe which is then covered at the exhaust port on top of biosafety cabinet, then fasten the pipe clamp; b. Keep the outward exhaust pipe level and determine the position of drilling on the wall; c. Drill circular holes with diameter being 20cm at selected positions; d. Open the attached packing box of outward exhaust fan and check the materials as per packing list; e. Connect the fixation frames for outward exhaust fan as per following figures; 22 / 98