Download CASE/ CardioSoft

Transcript
CASE/ CARDIOSOFT/ CS
DICOM CONFORMANCE STATEMENT
GE MEDICAL SYSTEMS
DIR 2040396-087 REV A
7.2.3 General Series Module
TABLE 7.2-3
GENERAL SERIES MODULE ATTRIBUTES
Attribute Name
Tag
Type
Attribute Description
Modality
(0008,0060)
1
Defined term “ECG” used.
Series Instance UID
(0020,000E)
1
Uniquely generated by the equipment.
Series Number
(0020,0011)
2
Fix: set to 1
Laterality
(0020,0060)
2C
Not used
Series Date
(0008,0021)
3
Not used
Series Time
(0008,0031)
3
Not used
Performing Physicians’ Name
(0008,1050)
3
Not used
Protocol Name
(0018,1030)
3
Filled with Exercise- or Ergospirometry test
protocol name.
Series Description
(0008,103E)
3
Not used
Operators' Name
(0008,1070)
3
Not used
Referenced Performed Procedure Step
Sequence
(0008,1111)
3
SOP Class UID: 1.2.840.10008.3.1.2.3.3
SOP Instance IOD: Uniquely created PPS SOP
Instance UID
> ‘Referenced SOP Class / Instance UIDs’
Body Part Examined
(0018,0015)
3
Not used
Patient Position
(0018,5100)
2C
Not used
Smallest Pixel Value in Series
(0028,0108)
3
Not used
Largest Pixel Value in Series
(0028,0109)
3
Not used
Request Attributes Sequence
(0040,0275)
3
Used if Requested Procedure ID, Scheduled
Procedure Step ID and Scheduled Protocol Code
are available from the worklist.
>Requested Procedure ID
(0040,1001)
1C
Used
>Accession Number
(0008,0050)
3
Not used
>Study Instance UID
(0020,000D)
3
Not used
>Referenced Study Sequence
(0008,1110)
3
Not used
>Requested Procedure Description
(0032,1060)
3
Taken from the worklist if available
>Requested Procedure Code Sequence
(0032,1064)
3
Not used
(0040,1002)
3
Not used
(0040,1001)
3
Not used
>>‘SOP Instance Reference Macro'
>>‘Code Sequence Macro’
>Reason for the Requested Procedure
> Reason for Requested Procedure
Code Sequence
7–4