Download Service Manual - Flight Medical
Transcript
Functional Description Remote Alarm Connection 3.5 Remote Alarm Connection The Flight-60 ventilator can be connected to third party remote alarm system. The remote alarm station displays all visible and audible alarms and alerts if the ventilator is shutdown. Remote alarm cable should be connected to the COM2 RS232/RA port (the lower port) in the back of the ventilator. The COM2 port provides a dry contact output, both Normally Open (NO) and Normally Closed (NC). Normally Open – Pin 7 – closing the contacts to transmit an alarm. Normally Closed – Pin 8 – opening the contacts to transmit an alarm. The dry contacts current should not exceed 1mA. The voltage on the contacts must be lower than 12 Volts. Remote Alarm Electrical schematics The legal and regulatory responsibility for this integration of the Flight 60 device into any external system lays on the integrator and not the Flight 60 manufacturer. This include but is not limited to responsibilities for the Design, Implementation, Installation validation verification and safety of the connection of the Flight-60 to the remote alarm. Falling to comply with the specification given here may result in damage to the device and will void its warranty. Flight 60 Service Manual | 3-9