Download Operation/Service Manual
Transcript
Appendix 5.2.5 Patient Parameter Specifications Pulse Rate from NIBP Accuracy: ± 1 bpm Resolution: 1 bpm Range: ADULT MODE PEDIATRIC MODE 40 to 240 bpm 40 to 240 bpm 5.2.6 NIBP Sub-System Functional Requirements 5.2.6.1 Maximum Cuff Pressure The software-controlled over-pressure monitor vents to atmosphere at the following pressures: ADULT MODE PEDIATRIC MODE 297 ±3 mmHg 243 ±3 mmHg Under single-fault conditions, the hardware controlled over pressure mechanism vents the cuff to atmosphere so that the pressure in the cuff does not exceed the following: 5.2.6.2 ADULT MODE PEDIATRIC MODE 300 (+10%) mmHg 300 (+10%) mmHg Cuff Inflation The inflation source is capable of supplying sufficient air to bring a volume of 500 cc to a pressure of 300 mmHg in no more than 20 seconds. If the cuff is not inflated 5 mmHg within 18 seconds, the cuff is vented and the measurement is stopped. 5.2.6.3 Maximum Leakage The maximum allowed pressure drop with the bleed valves closed is 6 mmHg in 60 seconds as measured with a 500 cc volume at differential pressures of 250 mmHg, 150 mmHg and 50 mmHg. 5.2.6.4 Vent Rate A volume of 500 cc, when vented, is reduced from a pressure of 260 mmHg to a pressure of 15 mmHg in a maximum of 10 seconds. 5.2.6.5 Initial Conditions An NIBP Zero is performed automatically before the NIBP can be initiated. An NIBP measurement will not initiate until the unit has been powered on for 5 seconds, in order to allow time for the Zero. Duo™ Operating Instructions 0070-10-0603-01 5-5