Download SERVICE MANUAL - Hill
Transcript
4.18 Line Voltage Fuses Chapter 4: Removal, Replacement, and Adjustment Procedures Figure 4-21. Fuse Location 4 Replacement CAUTION: Ensure that the fuse housing is properly oriented for the voltage rating (110-120V or 220-240V) on the bed that you are servicing. Failure to do so could result in equipment damage. 1. Orient the fuse housing properly for the voltage rating (110-220V or 220-240V) on the bed you are servicing. 2. Insert a serviceable line voltage fuse (C) into the fuse housing (A). 3. To install the replacement line voltage fuse (C), reverse the removal procedures. 4. To ensure proper operation of the Affinity® Three Birthing Bed, perform the “Function Checks” on page 2-6. Affinity® Three Birthing Bed Service Manual (man272ra) Page 4 - 47