Download EC9810 Service Manual
Transcript
EC9810 O3 ANALYZER SERVICE MANUAL 7. Set the PUMP SPEED COARSE to 99 and press Enter> (↵). 8. Allow approximately 1 minute for the pressure to drop to a stable reading. This reading should be typically 100 to 200 torr lower than the ambient pressure (depending upon the capacity of the pump). 9. Set this reading (in TORR) as PRESSURE 1 LOW in the PRESSURE CALIBRATION MENU and press Enter> (↵). 10. Disconnect the pressure transducer from the inlet and turn the pump OFF. 11. To verify that the pressure transducers on the standard flow control board are calibrated, view the GAS PRESSURE reading in the PRESSURE CALIBRATION MENU. it should be at ambient pressure +/- 2 TORR. This completes the pressure transducer calibration procedure. 3.5.1.3 Flow Calibration 1. From the HIDDEN MENU, select the FLOW CALIBRATION MENU. The menu of Figure 18 should be displayed. FLOW CALIBRATION MENU DESIRED FLOW FLOW CONTROL ZERO FLOW TRANSDUCER DF CONTROL LOOP VALVE SEQUENCING FLOW ZERO POINT FLOW SPAN POINT PUMP SPEED COARSE PUMP SPEED FINE GAS PRESSURE GAS FLOW 0.50 SLPM 81 CANCEL ENABLED ON CANCEL 0.500 SLPM 28 7 710.0 TORR 0.500 SLPM Figure 18. Flow Calibration Menu for Standard flow block 2. Verify the pump is turned OFF .and the CONTROL LOOP is DISABLED in the FLOW CALIBRATION MENU. 3. Load the linearization table into the flow control board EEPROM by setting FLOW TRANSDUCER DF to SET and pressing Enter> (↵). The linearization table contains the following parameters: Flow points: 0, 0.5, 1.75, 2.5, 3.25, 3.75. Voltage points: 0, 0.13, 0.91, 1.783, 2.5, 3.478. 38 98107601 Rev D