EC9852 SULFUR COMPOUNDS ANALYZER SERVICE MANUAL 4.2.5 Instrument Status INSTRUMENT STATUS GAS FLOW GAS PRESSURE REF. VOLTAGE CONC. VOLTAGE ANALOG SUPPLY DIGITAL SUPPLY : : : : : : 0.375 - 0.625 690 – 760 2.3 – 2.7 0.0 - 4.2 11.6 - 12.2 4.8 - 5.2 SLPM TORR VOLTS VOLTS VOLTS VOLTS HIGH VOLTAGE LAMP CURRENT : : 700 - 720 34 - 36 VOLTS MA VERSION 1.18.0012 TRS EXIT Figure 26. Instrument Status Ranges If any of the parameters displayed on the INSTRUMENT STATUS screen vary significantly from the values shown in Figure 26, the fault or operational problem is probably related. This is also true if one of the parameters is demonstrating a rapid change or is oscillating strongly around the desired setpoint. For more information on the operational limits and fault messages, refer to the table in section 4.2.7 System Faults.. Several of the parameters displayed on the INSTRUMENT STATUS screen are affected by the potentiometer settings on the PREPROCESSOR POTS screen. If a parameter is out of the normal operating range, make note of the parameter value, proceed to the respective menu, and examine the pertinent potentiometer settings. 4.2.6 System Temperatures SYSTEM TEMPERATURES CELL TEMP. : 47 - 53 DEG C CHASSIS TEMP. : FLOW TEMP. : COOLER TEMP. : 25 - 35 45 – 55 10 - 14 DEG C DEG C DEG C Figure 27. System Temperatures and Tolerances The SYSTEM TEMPERATURES screen provides the temperatures of the fluorescence cell, the chassis, and the thermoelectric cooler used to cool the PMT. Figure 27 contains the nominal values that should be displayed on this screen. If any of the parameters are outside the acceptable ranges, a 54 98527601 Rev. C