Download Dual Controller - Agilent Technologies
Transcript
Technical Information List of Commands 5 Remote I/O Commands Coding Tab. 32 Bitfield Remote I/O Input Remote I/O Output 01h id. I/O board (always 1) High Voltage enable 02h I/O board OK (always 1) Set Point2 active 04h Step mode Set Point1 active 08h Remote mode Interlock active 10h Protect mode High Voltage fault 20h Output HV enable Remote mode or Local mode (0) - Serial mode (1) 40h Confirm HV on Protect mode 80h Remote I/O interlock Not used (always 0) The following example shows the reading on the High Voltage 1, the status of the Start/Protect. Binary Protocol: ü 04 C01 ? X 81h 30h 34h 43h 30h 31h 3Fh 78h • 04 C0 1 0 W 1h 30h 34h 43h 30h 31h 30h 77h Ascii Protocol: @ 04 C0 1 ? 0391 40h 30h 34h 43h 30h 31h 3Fh 30h 33h 39h 31h $ 04 C0 1 0 0348 24h 30h 34h 43h 30h 31h 30h 30h 33h 34h 38h Dual Controller User Manual / 87-900-094-01 (I) 171/194