Download Airflo Mk3
Transcript
English Airflo Mk3 MI0259 issue 9 Trouble shooting guide Compressor does not run when ‘up’ is selected on the push button hand control. Battery State Indicator light flashing ‘Red’. Ensure that the air hose from the product being inflated is connected to the Airflo. This happens when the compressor’s internal temperature cut-out has tripped. Leave the Airflo for 60-90 minutes to cool. Compressor keeps stopping when trying to inflate the lifting aid. Compressor causes electrical interference with another piece of electrical equipment. Check that there are no kinks in the air hose. Move either the compressor or other item further away from each other until the interference stops. Battery State Indicator light flashing ‘Red/Green’. This happens when the current overload cut-out has tripped. Leave for 3 seconds to self reset. Specifications Airflo Weight: Input to Compressor: Output pressure: Battery: Environmental protection: Type B equipment 4.6Kg 12Vdc, 7Amp 0.35 bar (5 psi) 1 x 12Vdc - 2.3Ah (Valve Regulated Sealed Lead Acid) recycle IP51 (Airflo housing) IP67 (Push-button hand control) do not dispose of in household waste This equipment has been tested and meets the requirements of current EMC legislation. Charger Input: 230Vac, 50 Hz 10.5 VA Output: 15Vdc, 350mA Internal protection: 200mA 250Vac Fuse; Environmental protection: IP30 Double insulated Class II equipment Thermal Fuse 125ºC 250Vac / 2A freedom through lightweight solutions 7