EM7000_08_英_再校 13.6.10 1:26 PM ページ21 6-6-4 ACA (AC 6 A) 1) Objects of measurement AC current below 6 A in compact power supply circuits, etc. 2) Measurement procedure Insert the red plug of test leads into the power supply DC 6 A+/AC 6 A input AC6A SW Load terminal, and the black plug into the –COM terminal. Turn the function/range selector to the AC 6 A AC6A range at the lower center. Turn on the power switch of this unit (power ON: Lamp blinks). Turn the electrical zero adjuster (ZERO ADJ) to adjust the pointer to the zero position of red AC 6 A scale at the bottom. Connect the black test pin to a measuring point on one side of the circuit to be measured, and the red test pin to a measuring point on another side through the load (in series). (To ensure safety, turn off the power switch of the circuit to be measured before the test pins are connected, and after that, turn on the power switch of the circuit to be measured.) Read the meter indication on red AC 6 A scale and 0 - 6 figures in a unit of A (ampere). Detach the test pins from the circuit to be measured. Turn off the power switch (power OFF: Lamp blinking goes off). The continuous measurable time of the current 3 - 6 A is within 30 seconds as the shunt resistor in this unit overheats. The internal resistance of the AC 6 A range is very small and therefore if a voltage is applied by mistake to this range, very large current will flow, causing a danger. Though this current is shut off by the built-in fuse (6.3 A), take care not to make a wrong measurement. Further, observe the same precautions as those for the measurement of DC current (DCA ). 0 — 47 — DRAWING No.EM7000 09-1306 2040 2040