Download pH測定と電極の選び方 第一版 - 株式会社ティ・アンド・シー・テクニカル
Transcript
pH By Erich L.Springer © 2013 7 7 by T&C Technical 1. .......................................................................................................................... 5 1.1 ............................................................................................................. 7 2. PH ................................................................................................................ 8 3. PH ......................................................................................................................... 12 3.1 ........................................................................................................ 12 3.2 PH ...................................................................................................... 16 3.2.1 ........................................................................................................... 16 3.2.1.1 ................................................................................................. 18 3.2.1.2 ..................................................................................................... 19 3.2.2 ........................................................................................................... 21 3.2.3 ........................................................................................................... 22 3.2.4 KCl 3.2.5 ( ( ) ............................................ 23 ) .......................... 24 3.2.5.1 3.2.6 ...................................................................................... 25 pH ............................................................................. 26 3.3 PH ...................................................................................................... 27 3.3.1 ...................................................................................... 28 3.3.2 ...................................................................................... 29 3.3.3 ........................................................................................................ 31 3.3.4 pH ....................................................... 32 2 by T&C Technical 3.3.5 3.4 .................................................................................................... 34 .......................................................................................................................... 36 3.4.1 ........................................................................................................... 40 3.4.2 ............................................................................... 42 3.4.4.1 (EVEREF).................................................... 42 3.4.4.2 -B 3.5 ...................................................... 43 ......................................................................................... 44 3.5.1 .................................................................................................... 44 3.5.2 ............................................................................................................... 45 3.6 ............................................................................................. 46 3.7 ........................................................................................................ 48 3.8 .......................................................................................................................... 51 3.8.1 ................................................................................................. 51 3.8.2 ................................................................................................. 52 3.9 .......................................................................................................................... 54 3.9.1 pH 3.9.2 ............................................................................. 54 ........................................................................................................ 57 3.9.2.1 ............................................................................ 58 3.9.2.2 pH .......................... 61 3.9.3 ............................................................................................................... 62 3.9.4 ........................................................................................................... 62 3.9.5 / ..................................................................................... 64 3 by T&C Technical 3.10 ...................................................................................................... 65 4. ............................................................................................................. 66 4.1 5. .......................................................................................................................... 66 4.1.1 ................................................................................................. 67 4.1.2 ................................................................................................. 68 4.1.3 ................................................................................................. 68 4.2 ................................................................................................. 69 4.3 ........................................................................... 71 PH PH .................................................................................. 72 5.1 ......................................................................................... 72 5.2 ......................................................................................... 72 5.3 ............................................................................................. 73 5.4 ............................................................................................. 73 5.5 ...................................................................................... 75 4 Page 3 1. pH Measurement utilising pH Measurement Glass Electrodes utilising Theory and Practice Glass by Electrodes Erich K. Springer Page 3 by T&C Technical Theory and Practice 1 1 by Erich K. Springer INTRODUCTION pH sour and bitter when eating liquid or Since the creation of man, he could pH differentiate between INTRODUCTION solid foods. We say that vinegar and lemon juice taste sour and that soap water tastes bitter. pH This characteristic and thebetween marked degree of bitter sour or bittereating is indicated by the pH Since the creation of man,ofhethese couldliquids differentiate sour and when liquid or value. the pH value is today to athat great portion of tastes our population solid foods. WeThe saysignificance that vinegarof and lemon juice taste known sour and soap water bitter. alone through the fact that it is accepted that the pH value of swimming-pool water has to bepH kept within This characteristic of these liquids and the marked degree of sour or bitter is indicated by the certain limits. The food eatiscontains a lot toofa water and ourofdrinks are nothing else than value. The significance of the pHwe value today known great portion our population alone andit flavoured water, andpH thisvalue waterofcan be either sour or bitter; inkept otherwithin words, it will through coloured the fact that is accepted that the swimming-pool water has toorbe have aThe certain value. certain limits. foodpHwe eat contains a lot of water and our drinks are nothing else than pH coloured and flavoured water, and this water can be either sour or bitter; or in other words, it will In general we can say that water is the most have a certain pH value. important substance on earth. Without water there will be life on planet. People living in areas of In general wenocan sayourthat water is the most 1onhave severe drought experienced the importance of important substance earth. Without water there water for their existence. Continuous droughts do will be no life on our planet. People living in areas of not onlyhave affect the farmers, are a disaster for severe drought experienced thethey importance of thetheir entire economyContinuous of a nation. It is therefore of water for existence. droughts do importance treat this precious not onlyutmost affect the farmers,that theyweare a disaster for liquid with care, use it responsibly and try to save the entire economy of a nation. It is therefore of every drop of it. that we treat this precious liquid utmost importance with care, use it responsibly and try to save every drop of it. Water is everywhere! Three quarters of the surface of the earth is covered with water in the form of oceans,Three rivers, lakes of andtheswamps. Water is everywhere! quarters whiskey, beer,with wine, cool-drinks, fruit surface Brandy, of the earth is covered water in the coffee and tea consist water containing form ofjuices, oceans, rivers, lakes andof swamps. certain additives which change the and Brandy, whiskey, beer, wine, cool-drinks, property fruit the taste theconsist water. of Blood, vital liquid of our juices, coffee andoftea waterthe containing body is 1mainly water, the in fact our entire 1 change certain additives which property and body of approximately 65%liquid water. the tasteconsists of the water. Blood, the vital of our body is Water mainlyis water, fact our entire body used ininindustry as cooling water, wash consists water, of approximately 65% water. boiler water, steam, condensate and solvent. Dirty water is usually disposed of as effluent. 2/3 Notably chemical industrywater, especially Water is used inthe industry as cooling wash makes use of this universal solvent, because most substances are somewhat soluble water. Dirty water is usually disposed of as effluent. water, boiler water, steam, condensate andinsolvent. Notably the chemical industry especially makes use of this universal solvent, because most substances are somewhat soluble in water. 5 by T&C Technical 65% pH pH pH pH pH pH pH pH 6 by T&C Technical 1.1 pH pH pH pH pH pH pH pH pH pH Dr.Hannes Bühler, Andreas Brügger, Dirk Tillich, and thank to my loving wife 7 by T&C Technical 2. pH 1(100) 10-7 Page 5 10-14mol/L 2 THE pH SCALE If we express the hydrogen ion concentration of an aqueous solution in relation to its molecular value we derive a scale of 1 (100) via 10-7 to 10-14 mole/litre. 0 14 pH This scale is impractical but if written as a function of its negative logarithm a real and simple scale of 0 – 14 has been created: the pH scale. pH H+ concentration (mole/litre) mol/L OH- concentration (mole/litre) pH mol/L 1 0.1 0.01 0.001 0.0001 0.00001 0.000001 0.00000000000001 0.0000000000001 0.000000000001 0.00000000001 0.0000000001 0.000000001 0.00000001 0 1 2 3 4 5 6 0.0000001 0.0000001 7 0.00000001 0.000000001 0.0000000001 0.00000000001 0.000000000001 0.0000000000001 0.00000000000001 0.000001 0.00001 0.0001 0.001 0.01 0.1 1 8 9 10 11 12 13 14 hydrogenii 1909term pH is pH the abbreviationpondus of pondus hydrogenii and means the weight of hydrogen. This was introduced in 1909 by the Danish biochemist S. P. L. Sørensen (1868 – 1939). S.P.L. SØrensen (1868 1939 ) The pH scale covers the active concentration of the H+ ions and OH¯ ions and therefore the pH value is defined as the negative + common logarithm of the active hydrogen ion pH H OHconcentration in an aqueous solution. pH 1 pH = log hydrogen ion concentration ( mole/litre) If the H+ ion concentration changes by a factor of ten, the pH value changes by one unit. This illustrates how important it is to be able to measure the pH value to a tenth of a unit or even a hundredth of a unit in special applications. The pH definition refers to the active hydrogen ion concentration and not just to the hydrogen ion concentration. It is important to understand this difference. Only in dilute solutions are all anions and all cations so far apart that they are able to produce the maximum of the chemical energy, 8 i.e. the H+ ion concentration and the H+ ion activity are identical. For instance 0,01 mole hydrochloric acid is still classified as a dilute solution which dissociates completely and therefore concentration equals activity. by T&C Technical 1 pH = Log (mol/L) H+ 10 pH 1 0.1 0.01 pH H+ H+ 0.01 0.01 mol HCl: (H+) 1 mol HCl: pH 9 (OH-) by T&C Technical pH pH pH pH pH0 HCl + NaOH pH0 + pH14 pH 1mol/L pH14 1mol/L → HOH + NaCl → H2O + NaCl → pH7 NaCl pH pH =7 pH 7 pH 10 8 by T&C Technical Page 7 pH The pH Scale -1 mol/L HCl -1 mole 1010 HCl pH 0 1 2 3 -1 10 mol/L NaOH 10-1 mole NaOH HOH HOH 4 5 6 7 8 9 10 11 12 13 14 1mol/L HCl 1mol/L NaOH 1 mole/l HCl 1 mole/l NaOH increasing acidity pH 0 1 2 3 4 5 neutral 6 7 8 increasing alkalinity 9 10 11 12 13 14 Gastric Milk Well Sour Juices Milk Coffee Water Household Yoghurt Soap Ammonia Suds Beer Blood Milk Lemon of Juice Magnesia (30%wt) 3 THE pH MEASUREMENT 3.1 The NERNST Equation To determine the active hydrogen ion concentration, a pH measurement is necessary. Three methods are generally used for the direct determination of the pH value in aqueous solutions: 11 1) The visual method, a colour comparison with pH sensitive indicator paper (litmus) to a standard colour scale. by T&C Technical 3. pH 3.1 pH pH 3 1 pH 2 pH 3 pH pH pH ( ) ( ) 12 If two hydrogen electrodes (each a thin plate of polished platinum) are immersed in two solutions, having different hydrogen ion concentrations, each electrode will generate a potential which depends on the active hydrogen concentration of the solution in which the electrode is immersed. To be able to measure this potential, the two solutions are connected by a salt bridge (electrolyte bridge) and the two electrodes are connected to a high impedance voltmeter. Both solutions are by T&C Technical saturated with pure hydrogen gas. Since the current passing during such a measurement is negligible, the chemical composition of the sample solution is not altered. The bridge acts as a phase boundary between solution C1 and solution C2 and closes the electric circuit. High impedance mV meter 0 -1500 +1500 mV A potential difference will be generated between C1,C2 the two platinum electrodes by the different active hydrogen-ion concentrations in the solutions. The relationship is expressed by the NERNST equation: Electrolyte Bridge Pt Pt E C2 C1 RxT C x log 1 nxF C2 where: E = potential difference (mV) R = gas constant (8,31439 J x mol-1 x K-1) R T constant (96495,7 C1 C x mol-1) F = Faraday E= Log n F C2 E= mV R= (8.31439J×mol-1×K-1) (96495.7C×mol-1) F= T= (K) n= ( C1 = C1 C2 = C2 nH = 1) 13 by T&C Technical R T E= C1 C1 C2 10:1 Log n F C2 E R T E= 10 UN Log n F UN 10 R 1 n R T T E= ℃ 2.303585 n F T = 273.15 + 20 = 293.15 K UN = 58.16mV 14 F The values of R and F are constant. The charge number n is known for each kind of ion and the temperature T can be calculated from the measured value in °C. The values of R and F are constant. The charge number n is known for each kind of ion and the If we assume the temperature of the solutions to be 20 °C, then: temperature T can be calculated from the measured value in °C. T of = 273,15 + 20 =to293,15 If we assume the temperature the solutions be 20 Kelvin °C, then: by T&C Technical This will give us a NERNST potential T =of273,15 + 20 = 293,15 Kelvin Page 7 pH This will give us a NERNST potential UNof= 58,16 mV The pH Scale UN = 58,16 mV HOH HOH -1 mol/L HCl 1010-1 mole HCl -1 -1 mol/L NaOH 1010 mole NaOH The pH Scale and the Related NERNST Potential -1 -1 10 mole HCl HOH 10 mole NaOH The pH Scale and the Related NERNST Potential pH 0 1 2 3 4 5 10-1 mole HCl 1mol/L HCl 1pH mole/l HCl2 0 1 3 4 5 6 7 8 9 10 11 12 13 14 HOH 10-1 mole NaOH 6 7 1 mole/l 8 9 10 11 12 13 14NaOH increasing pH 0 1 2acidity 3 4 5neutral 6 7 1 mole/l HCl 1mol/L NaOH alkalinity 8increasing 9 10 11 12 13NaOH 14 1 mole/l 1 mole/l HCl 1 mole/l NaOH 0 0 1 2 3 4 5 6 mV 7 8 9 10 11 12 13 407,12 14 +pH 407,12 0 mV mV mV 407,12 + 407,12 Gastric mV Milk Well Sour Juices pH MilkpH CoffeepH Water Household +58,16 2x Yoghurt Soap-58,16 mV mVAmmonia mV at 20 oC +58,16mV= pH pH Suds 116,32 mVpH Beer Blood +58,16 -58,16 Milk Lemon 2x o mV mV at 20 C +58,16mV= of mV Juice 116,32 mV Magnesia mV As the ion activity is temperature dependent, so is the NERNST potential (refer to the NERNST equation). The following table illustrates the temperature dependency: As the ion activity is temperature dependent, so is the NERNST potential (refer to the NERNST equation). The table UN UN the temperature UN dependency: T following T illustrates T o o o C C C mV mV mV UN UN U T T T 0o 54.20 35o 61.14 70o 68.08 N 5 C 55.19mV 40 C 62.13mV 75 C 69.08mV 056.18 54.20 45 3563.12 61.14 80 7070.07 68.08 557.17 55.19 50 4064.12 62.13 85 7571.06 69.08 1058.16 56.18 55 4565.11 63.12 90 8072.05 70.07 1559.16 57.17 60 5066.10 64.12 95 8573.04 71.06 2060.15 58.16 65 5567.09 65.11 100 9074.04 72.05 25 59.16 60 66.10 95 73.04 Temperature Dependency of the NERNST Potential 30 60.15 65 67.09 100 74.04 10 15 20 25 30 Temperature Dependency of the NERNST Potential 3 THE pH MEASUREMENT 3.1 The NERNST Equation To determine the active hydrogen ion concentration, a pH measurement is necessary. Three methods are generally used for the direct determination of the pH value in aqueous solutions: 15 1) The visual method, a colour comparison with pH sensitive indicator paper (litmus) to a standard colour scale. by T&C Technical Page 11 3.2 The pH Measuring System 3.2 pH A basic pHpH measuring system consists of 0 -1500 mV +1500 (1) the measuring electrode, a pH sensitive electrode, (3) High impedance voltmeter Electrode assembly (1) Measuring electrode 1 (2) and Reference electrode (2) the reference pH electrode (3) a high impedance voltmeter. 2 3 3.2.1 The Measuring Electrode 3.2.1 The purpose of the measuringpH electrode is to determine the pH value of an /aqueous solution. The platinum/hydrogen electrode was originally used to measure the hydrogen ion 1897 concentration in aqueous solutions (since 1897) and today still serves as a reference standard for the electrometric pH determination. The hydrogen electrode consists of a platinised pH with platinum black), subjected to a flow of gaseous hydrogen. A platinum plate or rod (coated silver wire coated with silver chloride serves as a reference electrode. The basic theory, when employing a hydrogen electrode, is as follows: If a metal rod (electrode) is immersed into an aqueous solution containing its own salt (silver electrode in silver nitrate), the atoms on the surface of that metal rod will ionise. The water molecules will attract the positively charged metal ions from the surface of the rod, which leaves the metal rod negatively charged. This charge exchange develops a potential difference at the phase boundary metal/solution. The potential depends on the ion concentration in the (solution and is known as the galvanic potential. ) Today the hydrogen electrode still serves as a reference standard especially as its measuring results are extremely accurate. However for practical reasons the hydrogen electrode has lost its / importance because of its difficult and complicated handling. Only the antimony electrode has survived out of various metal electrodes. Antimony is chemically resistant to hydrofluoric acid and can therefore be used for the pH measurement in solutions containing hydrofluoric acid. However, the accuracy and reproducibility of the measuring result incorporates large tolerances. 16 by T&C Technical pH 17 3.2.1.1 The Glass Electrode It was not until the development of the glass electrode that pH measurement became a simple and reliable tool for all kinds of applications. In recent years the glass electrode has outgrown all other types of indicator electrodes for pH measurements. The pH determination of an aqueous by T&C Technical solution is today as common as temperature and pressure measurements, thanks to the reliability and accuracy of the glass electrode in combination with extreme stable electronic amplification. However, the successful application of the glass electrode requires some 3.2.1.1 knowledge about its functionality and its maintenance, which this booklet will provide. The Construction of a Glass Electrode pH cable pH Electrode head Electrode glass shaft A glass electrode consists of a shaft pH made from glass which should be highly resistant to hot alkaline solutions and its electrical resistance must be several times greater than that of the membrane glass. The pH sensitive part of the glass electrode is the normally cylindrically shaped electrode tip, the glass membrane. The membrane is made from special hydrogen ion sensitive glass and is fused to the electrode shaft. The pH glass electrode is partly filled with a buffer solution, normally having a pH value of 7. A defined amount of potassium chloride (KCl) is added to this internal buffer. Screen Internal conductor electrode (Ag/AgCl) Inner buffer+KCl(pH7) + KCl (pH 7) A silver wire, coated with silver chloride (Ag/ AgCl) is inserted into the glass electrode right down into the internal buffer and serves as a conducting electrode. Via the core of the coaxial pH cable, the Ag/AgCl wire is connected to one terminal of a pH meter. pH sensitive pH membrane glass 3.2.1.2 The Glass Membrane All types of glasses possess the property of producing a potential difference relative to the hydrogen ion concentration in aqueous solutions. However only special types, such as the pH conventional Mc-Innes glass (Corning 015) produce galvanic potentials which satisfy the NERNST equation over a wide range of the pH scale. Every manufacturer of pH electrodes is constantly researching for better pH sensitive glasses. Through constant development HAMILTON has achieved results which have not previously been available without unsatisfactory compromises. pH7 KCl When the membrane glass of a measuring electrode comes into contact with an aqueous solution, it forms a thin gel layer of approximately 10-4 mm thickness between the glass surface and the solution. The thickness of the gel layer depends on the quality and composition of the membrane glass, the temperature and the pH value of the measured solution. As the pH internal side of the glass membrane is in contact with the inner buffer (an aqueous solution of pH 7) a gel layer is also pH formed on the inside of the glass membrane. 18 by T&C Technical 3.2.1.2 Mc-Innes (Corning 015) pH pH pH 0.1µm pH H+ H+ H+ H+ 0mV pH Li+ 19 by T&C Technical A continuous exchange of H+ ions in the gel layers and H+ ions of the solutions takes both sides of the membrane. This ion exchange is1 controlled by the H+ ion concentrati pH 2 solutions. 24 ion concentration of each solution is identical on both sides of If the hydrogen membrane, the ion exchange stops after an equilibrium has been reached between the the solutions and the H+ ions in the gel layers. Therefore, both sides of the membrane g the same potential and the potential difference is 0 mV. The Glass Membrane Ag/AgCl electrode / Glass membrane 1mm appr. 0.5mm 0,5 to 1mm thick Inner gel layer approx. 10 mm thick + + Li+ pH + + + + ++ Li + H+ < pH 7 7 H+= const. Li Li+ Li + Outer gel0.1 layerm approx. 10 mm thick H+ > pH 7 Measured medium If a difference of a hydrogen ion conexists between the inner buffer and solution, a potential difference develops the inner and outer sides of the membr 0.1 m which is proportional to the differen between the inner buffer and the outer s be able to measure the membrane pot membrane itself has to be conductiv achieved by the mobility of the alkaline membrane glass (Li+ ions in most glasse Na+ ions in older membrane glasses). The thickness and composition of the determine the response time characteristic slope of the glass Therefore the gel layer is of critical im to the electrode performance. Without the gel layer there can be no pH measurement. Unfortunately it takes approxim to two days until a gel layer is fully developed. Therefore a measuring electrode ne hydrated (immersed into normal clean tap water) for at least 24 hours prior to manufacturers deliver their electrodes already hydrated (the membrane is kept wet w solution in a plastic cap) which renders the electrode ready for immediate use. 3.2.2 The Reference Electrode The reference electrode represents a defined electrical connection between the medium and the pH meter. The accuracy of the pH measurement is often determin reference electrode and therefore the choice of the reference electrode is of s importance. An ideal reference electrode should produce a predictable potential, wh respond only in accordance with the NERNST equation. A good and stable reference should also have a low temperature coefficient and possess no temperature hysteresis. A reference electrode consists of an internal electrode (similar to the measuring electro is immersed into a defined electrolyte. This electrolyte must be in contact with the medium. Over the years various reference systems have been employed, but only two the mercury-mercurous 20 chloride (calomel) and the silver-silver chloride electrodes were found reliable with respect to an accurate and stable potential. HAMILTON applies exclusively the silver-silver chloride reference system. (refer als EVEREF reference system” on page 25). Page 14 by T&C Technical At low and stable temperatures (max. 80 °C) the calomel electrode has a high potential stability 3.2.2and a high accuracy down to a hundredth of a millivolt. The Construction of a Reference Electrode pH pHcable Electrode head Electrode glass shaft Refill opening Internal reference system 33mol/L mole KCl KCl But today the silver/silver chloride electrode has gained pH practical acceptance and is by far the most frequently employed reference system. It is easy to manufacture, its pH potential rapidly attains equilibrium between –30 °C and 135 °C, and is very reproducible. The Ag-AgCl reference electrode remains stable and accurate especially with wide temperature fluctuations and at high temperatures up to 135 °C. The internal electrode of an Ag-AgCl reference electrode consists of a silver chloride coated silver wire which is immersed into potassium chloride of 3 mole concentration situated in a large chamber formed bypH the glass body of the reference electrode. A diaphragm (normally a small porous ceramic rod) is fitted at the bottom of this chamber to permit the potassium chloride to diffuse or leak into the measured medium. To complete the electric circuit the silver-silver chloride wire is connected via a coaxial cable to the pH meter. The Construction of a Combination Electrode Diaphragm pH cable 3.2.3 The Combination (Single Rod) Electrode Electrode head Since 1947 electrode manufacturers have combined the measuring electrode and the reference electrode into one unit, hence the name combination electrode. Today, the combination electrode is almost exclusively employed in laboratories/ and industrial plants. Only when the life expectancy is significantly different for the measuring and the reference EVEREF ( electrode, is the use of a pH) measuring system consisting of two separate electrodes recommended. Electrode glass shaft Refill opening 3 mole KCl Reference system / In a combination electrode the concentric space surrounding the measuring electrode is filled with the reference electrolyte and 1) (80℃) 1/100mV contains the internal reference system. A diaphragm near the bottom of the electrolyte chamber serves as the junction / between the KCl solution and the measured medium. As the reference electrolyte is a conductive medium, it acts as a screen 2) -30℃ 135℃ to the measuring electrode. Internal conductor electrode (Ag/AgCl) Inner Buffer + KCl (pH 7) pH sensitive glass membrane Diaphragm 21 by T&C Technical calomel electrode has a high potential stability 3) / volt. 135℃ lver/silver chloride electrode has gained nce and is by far the most frequently ce system. It is easy to manufacture, its ains equilibrium between –30 °C and 135 °C, ducible. The Ag-AgCl reference electrode d accurate especially with wide temperature high temperatures up to 135 °C. KCl de of an Ag-AgCl reference electrode consists coated silver wire which is immersed/ into of 3 mole concentration situated in a large the glass body of the reference electrode. pH ally a small porous ceramic rod) is fitted at the ber to permit the potassium chloride to diffuse sured medium. To complete the electric circuit 3.2.3 oride wire is connected via a coaxial cable to 1947 The Construction of a Combination Electrode 1 pHcable pH ode Electrode head ombined the into one unit, Today, the employed in e expectancy he reference consisting of Electrode glass shaft Refill opening 3 mole KCl 3mol/L KCl Reference system KCl rrounding the lectrolyte and agm near the the junction dium. As the s as a screen Internal conductor electrode (Ag/AgCl)(Ag/AgCl) Inner Buffer +KCl(pH7) + KCl (pH 7) pH pHsensitive glass membrane Diaphragm 22 by T&C Technical 3.2.4 KCl ( ) Page 15 3.2.4 Combination Electrode filled with KCl-Gel as a Reference Electrolyte ( Reference electrodes incorporating a liquid reference electrolyte are maintenance intensive as their electrolyte level has to be controlled and regularly topped up. The search for a 3mol/L assembly KCl maintenance-free electrode led to the development of the gel reference electrode. pH 3mol/L KCl (Ag/AgCl) +KCl (pH7) Diaphragm pH ) The gel electrode is a low-maintenance electrode. The reference electrolyte chamber is filled with 3 mol/l KCl electrolyte in gel form. The diaphragm is normally made from ceramic. The glass shaft is often reinforced with an outer plastic sleeve or the electrode shaft is made completely out of plastic (Epoxy). The gel electrode is mostly used in combination with portable pH meters or laboratory pH meters for simple measuring applications, e.g. public swimming pools. This electrode does not need to be topped-up with reference electrolyte – which pH reduces maintenance time. However it has to be taken into account that the gel electrode has pHa reduced accuracy and a shorter life span than its counterpart with a liquid reference electrolyte. The response time of a gel electrode is somewhat slower than an electrode filled with a liquid electrolyte. 3.2.5 Reference Electrode with POLISOLVE Electrolyte (Polymer) In the early 1980s electrode manufacturers created the polymer electrode. The polymer electrode is either a standard combination glass electrode or a separate reference electrode. In both cases the reference electrolyte chamber is completely filled with a semi-solid polymerised plastic material into which the KCl is embedded. No diaphragm junction is required. Therefore the KCl saturated polymer has direct contact with the measured solution. The contact is established through an aperture, which could either be one or more holes near the bottom of the reference electrolyte chamber or a fissure separating the bottom electrolyte chamber from the measuring electrode. - What does not exist can not clog up! – 23 maintenance-free electrode assembly led to the developm maintenance-free electrode assembly led to the development of the gel reference electrode. of the gel reference electrode. 3.2.5 The gel electrode. electrode is The gel electrode is a low-maintenance Thea low-maintenance electrode. reference electrolyte chamber is filled with 3 mol/l reference electrolyte chamber is filled with 3 mol/l KCl electrolyte in gel form. diaphragm is normally m electrolyte in gel form. The diaphragm is normally made by TThe &C Technical from ceramic.with Theanglass from ceramic. The glass shaft is often reinforced outershaft is often reinforced with an o or the out electrode shaft is made completely o plastic sleeve or the electrode shaft plastic is madesleeve completely of ( ) plastic (Epoxy). plastic (Epoxy). gel electrode is mostly used in combination with port The gel electrode is mostly used in The combination with portable pH for meters or measuring laboratory pH meters for simple measu pH meters or laboratory pH meters simple applications, e.g. public swimming applications, e.g. public swimming pools. This electrode does 1980 pools. This electrode d not need to be topped-up not need to be topped-up with reference electrolyte – which with reference electrolyte – w reduces maintenance time. However it has to be taken reduces maintenance time. However it has to be taken ( ) pHinto thataccuracy the gel electrode has a reduced accuracy account that the gel electrode has account a reduced and a shorter span than its counterpart with a liquid refere a shorter life span than its counterpart with a life liquid reference electrolyte. electrolyte. response slower time of than a gel electrode is somewhat slower The response time of a gel electrodeThe is somewhat Diaphragm an electrode filled with a liquid electrolyte. an electrode filled with a liquid electrolyte. hragm 1 KCl 2 with3.2.5 Reference Electrode with POLISOLVE Electrolyte (Polymer) Reference Electrode POLISOLVE Electrolyte (Polymer) In the earlycreated 1980s the electrode manufacturers created the polymer electrode. The poly early 1980s electrode manufacturers polymer electrode. The polymer electrode is either a standard combination glass electrode ode is either a standard combination glass electrode or a separate reference electrode. Inor a separate reference electrod both cases the isreference electrolyte chamber is completely filled with a semi-s cases the reference3 electrolyte chamber completely filled with a semi-solid 2 plastic material into the KCl is embedded. No diaphragm junctio merised plastic material intopolymerised which the KCl is embedded. No which diaphragm junction is required. Therefore KClcontact saturated has direct contact with the measu ed. Therefore the KCl saturated polymer has the direct withpolymer the measured solution. The contact is established through an aperture, on. The contact is established through an aperture, which could either be one or more which could either be one or m holeselectrolyte near the chamber bottom oforthe electrolyte near the bottom of the reference a reference fissure separating thechamber bottom or a fissure separating the bo electrolyte chamber from the measuring electrode. olyte chamber from the measuring electrode. What - What does not exist can not clog-up! – does not exist can not clog up! – Hole in electrolyte chamber Hole in electrolyte chamber 1 2 24 by T&C Technical 3 4 pH2 90℃ 5 3.2.5.1 (POLISOLVE) 1 pH0 2 3 4 130℃ 600kPa 5 6 2µS/cm pH pH SIP pH 25 pH As the KCl saturated polymer is free of AgCl, there is no possibility of silver sulphide contamination when measuring the pH value of solutions containing sulphides. The high pressure rating of 600 kPa, its extended temperature rating of up to 130°C and its maintenance-free operation should make the POLISOLVE reference system always a first choice. Electrodes, utilizing the POLISOLVE polymer electrolyte, are even suitable for steam sterilisation by T&C Technical in biotechnology applications. 3.2.6 pH 3.2.6 The Measurement of the Potential Difference The pH measuring pH electrode and the pH reference electrode form a so-called pH measuring chain within the measured medium. This chain can pH be compared to a battery of which the voltage produced depends on the measured medium. High impedance voltmeter 0 -1500 mV +1500 pH Electrode assembly Reference The difference in potential between the measuring electrode Measuring 2 electrode and the reference electrode is a function electrode of the pH value of the measured medium. In theory the voltage changes by 58,16 mV per The pH pH unit at 20 °C according to the NERNST measuring pH chain equation. The voltage produced by the pH measuring chain is large enough not to present any problem for a measurement. But the Measured medium measuring chain is a voltage source from which no pH current can be drawn, not even the low current which a moving coil of a DC voltmeter draws. The potential difference of the measuring chain has 20℃ 58.16mV pH the voltage to be measured without drawing any current from1pH the voltage source, otherwise would be reduced and the pH measurement would be drastically falsified. The reason is the high electrical resistance of the glass electrode which is mainly determined by the resistance of the glass membrane. pH The resistance values of a glass membrane vary between 10 M and 1 000 M at 25 °C and increase 10 times at a temperature decrease of 25 °C. The lowest possible operating pH temperature of a pH electrode is often determined by the resistance increase of the glass membrane at low temperatures, the internal resistance of the measuring instrument, the required accuracy of the pH measurement and the freezing point of the electrolyte. 25℃ 25℃ 10 pH 26 1000MΩ pH pH ±0.1pH 10MΩ by T&C Technical 1/100 5000MΩ 1012Ω 3.3 pH pH 27 3.3 The Characteristics of a pH Measuring Chain The characteristics of a pH measuring chain are the result of the individual properties of the measuring and the reference electrode. Most of the electrode assemblies in use today are combination electrodes. For this reason we will refer to the combination electrode when examining the different properties of a measuring chain. However everything that is said about by T&C Technical the properties of a combination electrode may be applied to the individual measuring and reference electrode as well. 3.3.1 3.3.1 The Different Potentials of a Combination Electrode Etotal E total EE11 : EE22 : Asymmetry potential of the glass membrane. 0 -1500 mV Potential developing on the outer gel layer. +1500 developing on EE33 : Potential the inner gel layer. developing on EE44 : Potential the internal conductor electrode. developing on EE55 : Potential the reference electrode. E5 E6 potential E6 : Diffusion the diaphragm. E4 E3 E6 of he pH to ta l otal E total :EtTmeasured p ot e nt i a l by a pH E E6 electrode1 assembly is the sum of potentials E1 to E6. E2 E1 pH pH mV 6 E2 E1 E6 pH E1 28 pH by T&C Technical E2 pH 0mV E3 E3 E4 E5 / E4 E6 E1 E2 E2 E6 E6 3.3.2 0mV 29 E5 by T&C Technical pH pH7 pH 0mV Etotal E1 7 E6 pH -30mV +30mV pH ( pH ( pH6.8) ) ±0.02pH(1.16mV) pH pH pH a) b) pH c) 30 / by T&C Technical 3.3.3 pH ( pH7) 0mV ±47mV(±pH0.8) pH 31 by T&C Technical 3.3.4 pH pH 1pH pH = U pH 58.16mV 0 pH7 pH7 -58.16mV 99.8% pH pH 32 pH14 slope potentiometer of the pH meter/transmitter. As with the zero point adjustment, the slope adjustment has to be performed at regular intervals. The slope of a pH electrode assembly is temperature dependent in accordance with the NERNST equation. The slope increases with the rise in temperature of the measured solution, as by T&lines C Technical can be seen from the diagram below. In theory, all temperature dependent slope intersect the theoretical zero point (pH 7). pH 450 400 350 300 The Slope or Sensitivity of a pH Electrode Assembly + Slope = pH 200 150 100 50 U 1 2 3 pH 4 5 6 mV 50 100 150 200 300 350 400 450 7 8 9 pH Theoretical Values o Slope at 0 C = 54,20 mV o Slope at 20 C = 58,16 mV o Slope at 50 C = 64,12 mV pH a) b) 10 11 12 13 14 pH c) d) e) 33 0 oC o 20 C 50 oC d) The electrode assembly must be symmetrical, i.e. measuring and reference electrode must have identical conducting systems in order to neutralise their galvanic potentials. e) The diffusion potential of the diaphragm should be as small and as constant as possible. by T&C Technical Slope and Asymmetry Potential pH of a pH Electrode Assembly 450 400 350 300 + 200 150 100 50 U pH Slope = pHas U 1 2 3 pH 4 5 6 7 8 9 mV pH 50 100 150 200 Asymmetry Potential Theoretical Values Slope at 0 oC = 54,20 mV 300 350 400 450 Slope at 20 oC = 58,16 mV Slope at 50 oC = 64,12 mV 3.3.5 (0mV pH7) pH Uis(Uis= 10 11 12 13 14 ) 34 o 0 C 20 oC 50 oC Page 22 3.3.5 The Isotherm Intersection Point by T&C Technical Again we have a disagreement between theory and practice. In theory all temperature dependent slope lines intersect the theoretical zero point (0 mV/pH 7). When an asymmetry potential is present – and pH that is always the case – this intersection shifts either to the right or to the left of the zero point, as can be seen from the diagram below. 2 of a pH electrode assembly vary with temperature. The temperature dependency of All potentials each individual potential cannot be accurately defined, but it will shift the resultant intersection pH away from the theoretical pH zero point and away pH from the point of all temperature slope lines asymmetry potential. This intersection point is then known as the isotherm intersection point mV pH (Uis = isotherm potential). In order to perform an accurate pH measurement, the position of the isotherm intersection point Uis Uis has to be established. Two buffer solutions are required. The position of the isotherm intersection point can then be determined by measuring the potential difference of the pH electrode assembly pH against various temperatures (heated buffer solutions). The mV outputs of the electrode assembly are then plotted against their pH value on graph paper and thereby the position of the pH Uis isotherm intersection point is established. The voltage and polarity of the established isotherm potential Uis can only be compensated for if the pH meter/transmitter used is equipped with an Uis potentiometer. Nowadays, only the microprocessor based pH meters/transmitters have the capacity to compensate for the Uis potential. pH 25℃ 60℃ New electrodes from HAMILTON show a maximum compensation error of 0,1 pH when 0.1pH in a solution having 60 °C. calibrated at 25 °C and thereafter measuring pH Slope and Isotherm Potential of a pH electrode Assembly 450 400 350 300 + 200 U 150 100 50 mV 50 100 150 200 300 350 400 450 Isotherm Intersection Point pH 1 2 3 Uis 4 5 6 7 8 9 Slope = U pH 10 11 12 13 14 pH Theoretical Values Slope at 0 oC = 54,20 mV o Slope at 20 C = 58,16 mV o Slope at 50 C = 64,12 mV 35 o 0 C o 20 C o 50 C by T&C Technical 3.4 Page 23 3.4 The Diaphragm ( The diaphragm is a very important and critical part of the reference electrode. It provides an electrolytic interface between the silver/silver chloride conducting system and the measured ) solution. In most cases the diaphragm consists of a porous ceramic plug fused into the glass wall at the lower end of a reference electrode (porous ceramic diaphragm). Various diaphragm types (differing in construction and shape) are available, each type has its advantages and limitations. It is normally the measurement application which determines the use of a specific diaphragm. a) Porous Ceramic Diaphragm a) The porous ceramic diaphragm is probably the most frequently used today. It possess a high chemical resistance and it is easy to manufacture. This junction provides a reproducible electrolyte flow but because of its large surface it is very vulnerable to contamination. b) Platinum Fibre Diaphragm b) Platinum fibre diaphragms consist of very fine platinum wires which are spun loosely together and fused into the glass. This type of junction resist contamination to a certain extent but their electrolyte flow is less reproducible than ceramic diaphragms. c) SINGLE PORE (Trademark of HAMILTON) Diaphragm The SINGLE PORE diaphragm is strictly speaking not a diaphragm at all. It is a very small glass capillary which allows a larger leakage rate than a ceramic or platinum diaphragm. A constant and very reproducible electrolyte flow is assured. Clogging or contamination is barely possible. It gives the most accurate and repeatable results. In combination with a polymeric electrolyte the SINGLE PORE principle is adapted for industrial electrodes. Due to the lack of36contamination and maintenance it has a lot of advantages. The German Federal Physical Technical Institute (PTB) decided during a traceability test in 1997 Platinum fibre diaphragms consist of very fine platinum wires which are spun loosely together and fused into the glass. This type of junction resist contamination to a certain extent but their electrolyte flow is less reproducible than ceramic by T&C Technical diaphragms. c) ( ) c) SINGLE PORE (Trademark of HAMILTON) Diaphragm The SINGLE PORE diaphragm is strictly speaking not a diaphragm at all. It is a very small glass capillary which allows a larger leakage rate than a ceramic or platinum diaphragm. A constant and very reproducible electrolyte flow is assured. Clogging or contamination is barely possible. It gives the most accurate and repeatable results. In combination with a polymeric electrolyte the SINGLE PORE principle is adapted for industrial electrodes. Due to the lack of contamination and maintenance it has a lot of advantages. The German Federal Physical Technical Institute (PTB) decided during a traceability test in 1997 that the SINGLE PORE pH electrode is the most accurate laboratory electrode. (“Traceability of pH measurement” by Petra Spitzer; ISBN 3-89429-877-4 or ISSN 0947-7063) (PTB) pH 1997 (“Tracebility of pH measurement” by Petra Spitzer; ISBN 3-89429-877-4 37 ISSN 0947-7063) Page 24 Page 24 d) Annular Ceramic Diaphragm by T&C Technical d) Annular Ceramic Diaphragm d) annular ceramic diaphragm is formed by a porous The ceramic layer ceramic between two glass tubes. The direction of the The annular diaphragm is formed by a porous measured medium is not critical due to the annular shape ceramic layer between two glass tubes. The direction of the of this junction. The electrolyte flow tois the notannular reproducible. measured medium is not critical due shape Hence is mainlyThe applied in gel-type of this itjunction. electrolyte flowelectrodes. is not reproducible. Hence it is mainly applied in gel-type electrodes. e) Ground Sleeve Diaphragm Electrolyte Electrolyte e) Ground Sleeve Diaphragm Ground sleeve diaphragms are ideally suited for applications in suspensions as these Ground sleeve diaphragmsandareemulsions, ideally suited for e) diaphragms can be cleaned easily by only pulling up the applications in suspensions and emulsions, as these glass sleeve. successful diaphragms can Another be cleaned easily byapplication only pullingofup this the diaphragm is theAnother pH measurement in low ionic solutions or glass sleeve. successful application of this in non-aqueous The electrolyte flowrate diaphragm is themedia. pH measurement in low ionic depends solutions on or the roughness of the ground glass surface of the sleeve in non-aqueous media. The electrolyte flowrate depends on and roughness the tightness theground sleeveglass fit. However the of of the surfacethis of diaphragm the sleeve is not the this pH diaphragm electrode and thesuitable tightnessfor of applications the sleeve fit.where However assembly is subjected to vibrationwhere as thisthe might the is not suitable for applications pHloosen electrode diaphragmissleeve. assembly subjected to vibration as this might loosen the diaphragm sleeve. The selection of the right diaphragm for a measuring application is of utmost importance but not always easy. Very the experimental “trial and error” method will lead to a successful The selection of theoften right only diaphragm for a measuring application is of utmost importance but not application of Very a certain type. For detailed one haswill to consult technical always easy. oftendiaphragm only the experimental “trial information and error” method lead to the a successful pH data sheets of the electrode manufacturers. application of a certain diaphragm type. For detailed information one has to consult the technical data sheets ofprovides the electrode manufacturers. A diaphragm a deliberate leak of the electrolyte solution into the measured medium whilst preventing unrestricted mixing ofthe both solutionssolution within into the the reference electrode. A diaphragm provides a deliberate leak of electrolyte measured medium Penetration of the unrestricted measured solution intoofthe reference electrolyte, thus poisoning of the whilst preventing mixing both solutions within and the reference electrode. reference conducting system solution occurs frequently during pHelectrolyte, measurements, especially when Penetration of the measured into the reference and thus poisoning of the measured conducting solution is pressurised. reference system occurs frequently during pH measurements, especially when the measured is pressurised. There are solution pH electrodes on the market where the electrolyte storage vessel can be pressurised in order to pH counteract theon penetration of where the measured solutionstorage throughvessel the diaphragm. As a rule There are electrodes the market the electrolyte can be pressurised of thumb a pressure of 100 kPa above the pressure of the measured solution will normally in order to counteract the penetration of the measured solution through the diaphragm. As a rule suffice. result aofsmall of electrolyte solution penetrate into the of thumbAsa apressure 100 amount kPa above the pressure of thewill measured solution willmeasured normally solution As which is generally no significance to the process. However this the suffice. a result a small of amount of electrolyte solution will penetrate into decreases the measured resistance of theisreference to between 0,1 k and 2 k , improves of solution which generallyelectrode of no significance to the process. However the thisreproducibility decreases the the measurement and prevents the diaphragm from clogging up. resistance of the reference electrode to between 0,1 k and 2 k , improves the reproducibility of the measurement and prevents theisdiaphragm up. Before the electrode assembly immersedfrom intoclogging the measured medium the stopper which closes the electrode refill opening must be removed. Periodic the electrolyte level should Before assembly is immersed into the inspection measuredofmedium the stopper which be part of the electrode maintenance programme. closes the refill opening must be removed. Periodic inspection of the electrolyte level should be part of the electrode maintenance programme. 38 by T&C Technical pH pH 100kPa 0.1 2kΩ 1 39 by T&C Technical 3.4.1 Page 25 3.4.1 The1Diffusion Potential (E6) Another disturbing factor of the diaphragm is its diffusion potential (E6). This potential always develops at the phase boundary between two electrolytes of different concentration or composition. The diffusion potential can be contributed to the different migration velocities of ions, which again depends on the polarity and size of the ion type. The illustration below explains the diffusion potential between two HCl solutions of different HCl concentration: H+ H+ ions ClThe diffuse nearly5 five times faster to the right than the Cl¯ ions. This creates a potential across the boundary of the two solutions. In order to keep the diffusion potential at the diaphragm of a reference electrode as small as possible, the different ions in the reference electrolyte should have identical ionic mobility. With a 3 mole KCl solution this ideal condition is nearly reached. Diffusion Potential Potential Cl H+ H+ Cl Cl H+ Cl H Cl H+ H+ + H + + Cl Cl Cl H+ Cl Cl H+ H Cl H+ H+ Cl H+ In general it can be said: Cl Distance 1. The higher the KCl concentration of the reference electrolyte, the lower the diffusion potential. 3mol/L KCl 2. The larger the flowrate of the reference electrolyte through the diaphragm, the smaller the diffusion potential. 1. The more the pH value of the measured solution differs from pH 7, the larger the diffusion potential. KCl 2. 3. pH7 Diffusion Potentials which develop between various solutions and a saturated KCl electrolyte pH 1,0 mole HCl 0,1 mole HCl 0,01 mole HCl 0,1 mole KCl Buffer pH 1,68 Buffer pH 4,01 Buffer pH 4,65 Buffer pH 7,00 Buffer pH 10,01 0,01 mole NaOH 0,1 mole NaOH 1,0 mole NaOH = = = = = = = = = = = = 14,1 mV 4,6 mV 3,0 mV 1,8 mV 3,3 mV 2,6 mV 3,1 mV 1,9 mV 1,8 mV 2,3 mV -0,4 mV -8,6 mV 40 From the above it can be seen that different measured solutions will create different diffusion potentials at the diaphragm of a reference electrode. Distance electrolyte, the lower the diffusion potential. 2. The larger the flowrate of the reference electrolyte through the diaphragm, the smaller the diffusion potential. by T&C Technical The more the pH value of the measured solution differs from pH 7, the larger the diffusion potential. Diffusion Potentials which develop between various solutions and a saturated KCl electrolyte 1,0 mole HCl 0,1 mole HCl 0,01 mole HCl 0,1 mole KCl Buffer pH 1,68 Buffer pH 4,01 Buffer pH 4,65 Buffer pH 7,00 Buffer pH 10,01 0,01 mole NaOH 0,1 mole NaOH 1,0 mole NaOH = = = = = = = = = = = = 14,1 mV 4,6 mV 3,0 mV 1,8 mV 3,3 mV 2,6 mV 3,1 mV 1,9 mV 1,8 mV 2,3 mV -0,4 mV -8,6 mV From the above pH7 it can be seen pH that different measured solutions will create different diffusion potentials at the diaphragm of a reference electrode. 41 by T&C Technical 3.4.2 Page 26 3.4.2 Diaphragm Contamination through Chemical Reaction pH Chemical reaction at the diaphragm between the reference electrolyte and the measured solution must be avoided at all costs. This reaction will lead to diaphragm contamination, increase of resistance across the diaphragm and falsified measuring results. The reference electrolyte contains silver chloride which is prone to chemical reactions, especially with sulphides. For this reason great care must be taken when measuring pH in solutions containing sulphides, as the diaphragm may be contaminated with silver sulphide deposits. Silver sulphide contamination can easily be identified by a blackened diaphragm. As a result, the response time of an electrode assembly increases substantially, the diaphragm resistance increases radically and it may be impossible to calibrate such a contaminated electrode assembly. 3.4.4.1 (EVEREF) 3.4.2.1 The EVEREF Reference System Reference conductor Silver wire Reference electrolyte with constant chloride concentration, but free of AgCl Silver chloride reservoir Diffusion pipe filled with cotton wool Diaphragm In order to counteract silver sulphide contamination at the diaphragm HAMILTON have invented the EVEREF reference system. This system consists of a silver (EVEREF) chloride reservoir from which the silver reference wire leads to the electrode plug. The reservoir is separated from the reference electrolyte by a diffusion barrier consisting of densely packed cotton wool in a glass tube. The barrier prevents the loss of silver chloride into the reference electrolyte induced by temperature variations. The EVEREF reference system enhances the stability of the reference potential and extends the life of the combination electrode considerably. 3.4.2.2 The EVEREF-B Double Liquid Junction If a chemical reaction at the diaphragm is unavoidable, the application of a reference electrode having an inter-mediate electrolyte provides a solution. 42 The EVEREF-B double liquid junction system from HAMILTON with its intermediate electrolyte reservoir provides such a solution. The reference electrolyte is completely separated from the EVEREF reference system by an internal diaphragm situated in a second reservoir filled with an intermediate electrolyte. The life of the pH electrode is greatly prolonged as the reference by T&C Technical Page 27 3.5 Alkaline and Acid Error 3.5.1 Alkaline Error 3.4.4.2 -B pH Electrode pH “Chemotrode Bridge” with Intermediate Electrolyte Electrode Head (Variopin or SMEK) VP SMEK Refill opening Intermediate electrolyte Reference system EVERREF-B -B Internal diaphragm 3 mole KCl 3mol/L KCl Internal conductor electrode (Ag/AgCl) Internal buffer KCl plus KCl At a value above pH 10 the gel layer structure at the membrane of a measuring electrode is subject to certain changes which lead to a measuring inaccuracy, the alkaline error. This alkaline error is caused by the presence of a high concentration of alkaline ions, especially sodium ions (Na+). These ions replace, partly or completely, the hydrogen ions in the outer gel layer of the glass membrane, and by doing so, contribute to the potential at the outer phase boundary. -B(EVEREF-B) As a result a lower pH value will be measured than the actual pH value of the measured solution. In earlier days the alkaline error of glass electrodes already developed between pH 9 and pH 10. Today, where the glass membranes contain lithium instead of sodium, the alkaline error is only noticeable from between pH 12 and pH 13. The alkaline error increases with increasing pH value, with higher alkaline concentration and with rising pH temperature. In order to counteract the alkaline error, electrode manufacturers use special membrane glasses with low alkaline errors for electrodes which are used to measure high -B pH values. pH 50 1 External diaphragm 2 3 4 5 6 7 8 9 10 11 12 13 14 0 50 100 150 mV 200 250 300 350 400 43 Alkaline Error 3.5 Alkaline and Acid Error 3.5.1 Alkaline Error pH Electrode “Chemotrode Bridge” with Intermediate Electrolyte 3.5 Electrode Head (Variopin or SMEK) 3.5.1 pH10 Refill opening At a value above pH 10 the gel layer structure at the membrane of a measuring electrode is subject certain by T&CtoTechnical changes which lead to a measuring inaccuracy, the alkaline error. This alkaline error is caused by the presence of a high concentration of alkaline ions, especially sodium ions (Na+). These ions replace, partly or completely, the hydrogen ions in the outer gel layer of the glass membrane, and by doing so, contribute to the potential at the outer phase boundary. As a result a lower pH value will be measured than the actual pH value of the measured solution. In earlier days the alkaline error of glass electrodes already developed between pH 9 and pH 10. Today, where the glass membranes contain lithium instead of sodium, the alkaline error is only noticeable from between pH 12 and pH 13. Intermediate electrolyte Reference system EVERREF-B Internal diaphragm The alkaline error increases with increasing pH value, pH with rising with higher alkaline concentration and temperature. 3 mole KCl Internal conductor electrode (Ag/AgCl) 13 pH9 10 In order to counteract the alkaline error, electrode manufacturers use special membrane glasses pH12 with low alkaline errors for electrodes which are used to measure high pH values. Internal buffer plus KCl pH 50 1 External diaphragm 2 3 4 5 6 7 8 9 10 11 12 13 14 0 50 100 150 mV 200 250 300 350 400 44 Alkaline Error by T&C Technical Page 28 3.5.2 Acid Error 3.5.2 400 At low pH pH values (< pH 2) the potential difference between measuring and reference electrode will not conform exactly to the NERNST equation. Through experiments it has been proven that the gel layer of the membrane will absorb acid molecules at very low pH values. Acid Error 350 300 + mV 250 200 150 pH the activity of the H+ ions This absorption decreases and results in a lower potential at the outer membrane phase boundary. The pH measurement shows a higher pH value than the actual pH value of H+ the measured solution. This effect is known as the acid error. 100 50 1 2 3 4 5 6 7 8 9 10 11 12 13 14 0 50 pH pH As with the alkaline error, manufacturers supply measuring electrodes with membrane glasses having specially low acid errors. HAMILTON membrane glasses show no acid error above pH 1. 3.6 Temperature Influence and Temperature Compensation pH1 The pH measurement is temperature dependent. Three temperature factors have to be considered in order to perform a nearly perfect pH measurement: 1. the temperature dependency of the NERNST equation 2. the position of the isotherm intersection point 3. the pH/temperature dependency of the measured solution 450 400 350 300 + Measuring Error without Temperature Compensation 31.72 mV Calibration at 25 oC o Measurement at 65 C 200 150 100 50 mV 50 100 150 200 300 350 400 450 The temperature dependency of the Nernst equation and with it the temperature dependency of the theoretical slope of a pH electrode assembly has already been discussed in paragraph 3.1. The slope of a pH electrode assembly is temperature dependent in accordance with the NERNST equation. 0,47 pH 1 2 3 4 5 6 7 8 9 10 11 12 13 14 pH Theoretical Values Slope at 25 oC = 59,16 mV o 25 C 65 C o o Slope at 65 C = 67,09 mV 45 The temperature dependency of the NERNST equation is easily calculated, and as a rule, only this temperature influence is considered by instrument manufacturers when they incorporate conventional manual or automatic temperature compensation in their pH measuring products. The adjacent graph illustrates the theoretical error which is compensated for by conventional temperature compensation. 100 50 1 2 3 4 5 6 7 8 9 10 11 12 13 14 0 50 and results in a lower potential at the outer membrane phase boundary. The pH measurement shows a higher pH value than the actual pH value of the measured solution. This effect is known as the acid error. by T&C Technical As with the alkaline error, manufacturers supply measuring electrodes with membrane glasses having specially low acid errors. 3.6 HAMILTON membrane glasses show no acid error above pH 1. pH pH 3.6 Temperature Influence and Temperature Compensation The pH measurement is temperature dependent. Three temperature factors have to be considered 1 in order to perform a nearly perfect pH measurement: 1. the temperature dependency of the NERNST equation 2 2. the position of the isotherm intersection point 3 450 400 350 300 + pH/ 3. the pH/temperature dependency of the measured solution Measuring Error without Temperature Compensation 31.72 mV 25 Calibration at 25 oC o Measurement at 65 C 65 200 150 100 50 0,47 pH 1 2 3 mV 4 5 6 7 8 9 10 11 12 13 14 pH 50 100 150 200 300 350 400 450 The temperature dependencypHof the Nernst equation and with it the temperature dependency of the theoretical slope of a pH electrode assembly has already been 3.1discussed in paragraph 3.1. The slope of a pH electrode assembly is temperature pH dependent in accordance with the NERNST equation. Theoretical Values 25 65 o Slope at 25 C = 59,16 mV 59.16mV 67.09mV o Slope at 65 C = 67,09 mV o 25 C 65 oC The temperature dependency of the NERNST equation is easily calculated, and as a rule, only this temperature influence is considered by instrument manufacturers when they incorporate conventional manual or automatic temperature compensation in their pH measuring products. The adjacent graph illustrates the theoretical error which is compensated for by conventional pH temperature compensation. 3.3.5 pH7 pH pH Uis pH 46 Uis by T&C Technical 3 pH/ pH pH pH/ pH pH 47 by T&C Technical 3.7 pH Page 30 Page 30 3.7 Various Electrode Shapes Various Electrode Shapes It is 3.7 not possible to use one electrode shape for every application. More often in the laboratory different electrode shapes are required as there are numerous pH measurement applications. It is not possible to use one electrode shape for every application. More often in the laboratory Electrode manufacturers try to cover most of these applications by offering varying electrode different electrode shapes are required as there are numerous pH measurement applications. constructions. a) manufacturers try to cover most of these applications by offering varying electrode Electrode Form a) constructions. Form a) This is the most common electrode shape and a wide field of applications can be covered with this type of electrode, both in the laboratory and in process control. Measuring, reference and This is the most common electrode shape and a wide field of applications can be covered with combination electrodes are manufactured using this construction. this type of electrode, both in the laboratory and in process control. Measuring, reference and combination electrodes are manufactured using this construction. Form b) Form b) b) A standard combination electrode construction with a ground sleeve diaphragm. This electrode is mainly used in the laboratory where dirty or strongly contaminated solutions have to be A standard combination electrode construction with a ground sleeve diaphragm. This electrode is measured. Its application includes non-aqueous media as well. The diaphragm can easily be mainly used in the laboratory where dirty or strongly contaminated solutions have to be cleaned by pushing the sleeve upwards. There is a limited use for this construction in process measured. Its application includes non-aqueous media as well. The diaphragm can easily be control (be aware of vibration). cleaned by pushing the sleeve upwards. There is a limited use for this construction in process control (be aware of vibration). Form c) Form c) This construction example includes two electrode features: a ground sleeve diaphragm and a pipe connector. In order to minimise maintenance time and to pressurise the reference electrolyte This construction two electrode ground sleeve diaphragm and a an external electrolyteexample reservoirincludes is connected to the features: electrode avia the pipe connector. This pipe connector. In order to minimise maintenance time and to pressurise the reference electrolyte electrode construction can be used in the laboratory and in process control, especially for high external electrolyte reservoir is connected to the electrode via the pipe connector. This purityanwater control in power stations. electrode construction can be used in the laboratory and in process control, especially for high 48 purity water control in power stations. A standard combination electrode construction with a ground sleeve diaphragm. This electrode is mainly used in the laboratory where dirty or strongly contaminated solutions have to be measured. Its application includes non-aqueous media as well. The diaphragm can easily be cleaned by pushing the sleeve upwards. There is a limited use for this construction process by T&C in Technical control (be aware of vibration). c) Form c) This construction example includes two electrode features: a ground sleeve diaphragm and a pipe connector. In order to minimise maintenance time and to pressurise the reference electrolyte an external electrolyte reservoir is connected to the electrode via the pipe connector. This electrode construction can be used in the laboratory and in process control, especially for high purity water control in power stations. Page 31 Form d) d) Page 31 Form d) This electrode construction is mainly used in the chemical industry and in biotechnological processes. The electrode features a large electrolyte vessel which is sometimes combined with an intermediate electrolyte vessel. The electrolyte can be pressurised and also sterilised with SIP hot steam. A special electrode holder is required for this electrode. No laboratory application. This electrode construction is mainly used in the chemical industry and in biotechnological processes. The electrode features a large electrolyte vessel which is sometimes combined with Form e) an intermediate electrolyte vessel. The electrolyte can be pressurised and also sterilised with hot steam. A special electrode holder is required for this electrode. No laboratory application. Form e) e) The above electrode construction is used in the laboratory where small samples have to be measured. Form f) The above electrode construction is used in the laboratory where small samples have to be measured. Form f) The feature of this construction is the flat membrane which enables the operator to measure the pH of surfaces, e.g. skin, leather, paper etc. HAMILTON supplies this electrode with an unbreakable plastic shaft, as it will often be carried 49 around and used with a portable pH meter. The feature of this construction is the flat membrane which enables the operator to measure the Form g) pH of surfaces, e.g. skin, leather, paper etc. HAMILTON supplies this electrode with an processes. The electrode features a large electrolyte vessel which is sometimes combined with an intermediate electrolyte vessel. The electrolyte can be pressurised and also sterilised with hot steam. Form e)A special electrode holder is required for this electrode. No laboratory application. The above electrode construction is used in the laboratory where small samples have to be measured. Form e) by T&C Technical The above electrode construction is used in the laboratory where small samples have to be Form f) f) measured. The above electrode construction is used in the laboratory where small samples have to be measured. Form f) The feature of this construction pHis the flat membrane which enables the operator to measure the pH of surfaces, e.g. skin, leather, paper etc. HAMILTON supplies this electrode with an Form f) unbreakable plastic shaft, as it will often be carried around and used withpH a portable pH meter. The feature of this construction is the flat membrane which enables the operator to measure the pH ofg)surfaces, e.g. skin, leather, paper etc. HAMILTON supplies this electrode with an Form unbreakable plastic shaft, as it will often be carried around and used with a portable pH meter. The feature of this construction is the flat membrane which enables the operator to measure the pH of surfaces, e.g. skin, leather, paper etc. HAMILTON supplies this electrode with an unbreakable plastic shaft, as it will often be carried around and used with a portable pH meter. Form g) g) This construction is used exclusively for combination electrodes featuring a gel or polymer reference Form g) electrolyte. The shaft is made completely from plastic. This design makes the shaft unbreakable as it is often used with a portable pH meter. This construction is used exclusively for combination electrodes featuring a gel or polymer reference design makes the shaft Form H) electrolyte. The shaft is made completely from plastic. This pH unbreakable as it is often used with a portable pH meter. This construction is used exclusively for combination electrodes featuring a gel or polymer reference electrolyte. The shaft is made completely from plastic. This design makes the shaft unbreakable as it is often used with a portable pH meter. Form H) The above construction, pointed electrode, are normally applied in the food laboratory and in the dairy They facilitate the pH measurement in meat and cheese. Formindustry. H) h) The above construction, pointed electrode, are normally applied in the food laboratory and in the dairy industry. They facilitate the pH measurement in meat and cheese. The above construction, pointed electrode, are normally applied in the food laboratory and in the dairy industry. They facilitate the pH measurement in meat and cheese. pH 50 by T&C Technical 3.8 3.8.1 1 2 3 4 60℃ 5 6 a) b) c) d) pH4 pH8 18 90℃ 2 51 by T&C Technical pH13 90℃ 2 3 pH 3.8.2 3.4 ( ( 200kPa) ) KCl 60℃ 52 by T&C Technical pH2 25℃ 6 (POLISOLVE) 130℃ pH0 90℃ 130 12 53 12 by T&C Technical 3.9 pH 3.9.1 pH pH pH (NIST) pH 9 pH NIST pH NIST 0℃ ±0.005pH 60℃ 95℃ ±0.008pH ±0.02pH 1 CO2 54 NIST 60℃ by T&C Technical 5 pH9.21 pH10.01 pH NIST PTB pH Values ofNIST Standard Buffer Solutions pH ac cording to NIST NBS NBS A Code Temp. Potassium tetraC oxalate 1,668 1,670 1,672 1,675 1,679 1,683 1,688 1,691 1,694 1,700 1,707 1,715 1,723 1,743 1,766 1,792 1,806 0 5 10 15 20 25 30 35 38 40 45 50 55 60 70 80 90 95 B Potassium hydrogen tartrate H C Potassium Potassium dihydrogen hydrogen phthalate citrate 3,863 3,840 3,820 3,802 3,788 3,776 3,766 3,759 3,755 3,753 3,750 3,749 3,750 3,753 3,763 3,780 3,802 3,815 3,557 3,552 3,549 3,548 3,547 3,547 3,549 3,554 3,560 3,580 3,609 3,650 3,674 4,010 4,004 4,000 3,999 4,001 4,006 4,012 4,021 4,027 4,031 4,043 4,057 4,071 4,087 4,126 4,164 4,205 4,227 D F Phosphate Borax 6,984 6,951 6,923 6,900 6,881 6,865 6,853 6,844 6,840 6,838 6,834 6,833 6,834 6,836 6,845 6,859 6,877 6,886 7,534 7,500 7,472 7,448 7,429 7,413 7,400 7,389 7,384 7,380 7,373 7,367 - 9,464 9,395 9,332 9,276 9,225 9,180 9,139 9,102 9,081 9,068 9,038 9,011 8,985 8,962 8,921 8,885 8,850 8,833 B H C D E I G 55 I G Sodium carbonate/ Calcium sodium hydrogen hydroxide carbonate 10,317 10,245 10,179 10,118 10,062 10,012 9,966 9,925 9,903 9,889 9,856 9,828 - 13,423 13,207 13,003 12,810 12,627 12,454 12,289 12,133 12,043 11,984 11,841 11,705 11,574 11,449 - Page 35 A F E Phosphate / by T&C Technical Page 36 Stability Comparison of HAMILTON Alkaline Buffer to Conventional Alkaline Buffer Both buffer solutions have been exposed to blowing air 9,3 9,2 Page 36 pH Stability Comparison of HAMILTON Alkaline Buffer to Conventional Alkaline Buffer 9,1 Both buffer solutions have been exposed to blowing air 9,3 9,0 HAMILTON DURACAL Buffer 9,2 pH9.2 9,2 Conventional Buffer pH 8,9 -12 9,1 0 9,0 12 24 36 Hours 48 60 72 HAMILTON DURACAL Buffer 9,2 Conventional Buffer Used buffer solutions should always be discarded and never be returned to their original storage 8,9 60 storage 12 24 36 48 72 bottle. For this reason HAMILTON -12 have0 developed a unique bottle which includes a Hours calibration compartment with a non-return valve at the bottom, preventing the used buffer (CALPAK) solution to be returned into the storage bottle. This CALPACK bottle is practical: no additional Used buffer solutions should be discarded andonly neverthe be returned to their original calibration container is required, andalways it is economical: required amount ofstorage buffer solution bottle. For this reason HAMILTON have developed a unique storage bottle which includes a 1 is used. calibration compartment with a non-return valve at the bottom, preventing the used buffer solution to be returned into the storage bottle. This CALPACK bottle is practical: no additional The Practicality of the HAMILTON CALPACK Calibration Bottle calibration container is required, and it is economical: only the required amount of buffer solution 2 is used. The Practicality of the HAMILTON CALPACK Calibration Bottle 56 by T&C Technical pH pH ( (NIST/PTB) 1.09 ±0.02pH 60 1.68 ±0.02pH 60 2.00 ±0.02pH 60 3.06 ±0.02pH 60 4.01 ±0.01pH 60 5.00 ±0.02pH 60 6.00 ±0.02pH 60 7.00 ±0.01pH 60 8.00 ±0.02pH 60 9.21 ±0.02pH 60 10.01 ±0.02pH 60 11.00 ±0.05pH 24 12.00 ±0.05pH 24 3.9.2 ( ) pH pH pH pH 57 ) by T&C Technical mV pH pH pH 3.9.2.1 1) 2 pH 1 pH Page 38 pH7 3.9.2.1 Calibration of Analogue pH Meters 1 pH 1) Two buffer solutions of different pH values have to be selected. The pH value of one buffer solution should be as close as possible to the zero point of the electrode assembly, which is, under normal circumstances, pH 7. 2pH should have a pH value which lies as near as possible at the end The second buffer solution of the anticipated measuring range and the value should differ as much as possible from the zero3buffer solution, at least 2 pH units. 2 3 If the measuring range covers the acidity and the alkalinity range then a third buffer solution is 2pHsolution should lie in the acid region and the third required. The pH value of the second buffer buffer solution should lie in the alkaline region of the desired measuring range. Both values should differ as much as possible. pHpH Scale 0 1 2 3 4 5 6 1 I measuring range 7 8 9 10 11 12 13 14 measuring range II 2 Slope Calibration Zero Calibration Verifying HAMILTON Buffer Solutions 2) The pH value of any buffer solution is temperature dependent. Therefore the temperature of the applied buffer solution has to be measured in order to establish the correct pH value of the buffer solution. 58 The temperature adjustment potentiometer (manual temperature compensation) must be set to the buffer temperature. With automatic temperature compensation the temperature sensor, connected to the pH measuring instrument, must be immersed into the buffer solution. Slope Slope Calibration Calibration 2) Zero Zero Calibration Calibration Verifying Verifying by T&C Technical HAMILTON Buffer Solutions HAMILTON Buffer Solutions pH pH The value of any buffer solution is temperature dependent. Therefore temperature 2) 2) The pHpH value of any buffer solution is temperature dependent. Therefore thethe temperature of of ( ) the applied buffer solution has to be measured in order to establish the correct pH value of the applied buffer solution has to be measured in order to establish the correct pH value of thethe buffer solution. buffer solution. The temperature adjustment potentiometer (manual temperature compensation) must The temperature adjustment potentiometer (manual temperature compensation) must be be setset to the buffer temperature. With automatic temperature compensation temperature sensor, to the buffer temperature. With automatic temperature compensation thethe temperature sensor, connected to the measuring instrument, must immersed buffer solution. connected to the pHpH measuring instrument, must be be immersed intointo thethe buffer solution. Manual Temperature Compensation Manual Temperature Compensation Temperature Dependency Temperature Dependency of of Technical buffer Solutions Technical buffer Solutions 7,00 pH pH 7,00 Uis Uis zero pH 7,00 oC zero o C o C slope pH 4,01 pH 7,00 5 10 15 18 20 22 25 30 35 40 45 50 slope pH 4,01 pH pH 10,01 10,01 3) oC HAMILTON Buffer Solutions HAMILTON Buffer Solutions 54,014,01 104,004,00 154,004,00 184,004,00 204,004,00 224,004,00 254,014,01 304,014,01 354,024,02 404,034,03 454,044,04 504,054,05 7,097,09 7,067,06 7,047,04 7,037,03 7,027,02 7,017,01 7,007,00 6,996,99 6,986,98 6,976,97 6,976,97 6,976,97 10,19 10,19 10,15 10,15 10,11 10,11 10,08 10,08 10,06 10,06 10,04 10,04 10,01 10,01 9,979,97 9,929,92 9,869,86 9,839,83 9,799,79 pH7(0mV) 1 Page 39 3) After rinsing the electrode assembly with de-ionised water, it is immersed into the first buffer solution for the zero point compensation (normally pH 7 = 0 mV). The electrode assembly has @pH pHas to stay in the buffer solution for a short time, at least for one minute, till the indication of the pH meter settles near the pH value of the buffer solution (pH 7) and becomes stable. Thereafter pH the indicated value has to be adjusted with the zero potentiometer to the exact value of the buffer solution. The zero potentiometer is sometimes marked @pH or pHas. For analogue pH meters it is important that the zero point calibration is always performed 4) before the slope calibration. 4) After the zero point calibration, the electrode assembly has to be rinsed again with deionised water and dried with tissue paper. When drying the electrode, care must be taken not to rub the membrane, i.e. only dab the electrode with tissue paper. Under no circumstances must an electrode be rubbed. This might introduce static electricity into the glass shaft of the electrode and could make an accurate pH measurement impossible for hours. No rubbing! 59 second buffer solution (acid or base – 5) The electrode assembly is then immersed into the according to the measuring range). After the indication of the pH meter has stabilised, the indicated value has to be adjusted with the slope potentiometer to the exact pH value of the second buffer solution. The slope potentiometer is sometimes marked mV/ pH. Page 39 by T&C Technical 3) After rinsing the electrode assembly with de-ionised water, it is immersed into the first buffer solution for the zero point compensation (normally pH 7 = 0 mV). The electrode assembly has to stay in the buffer solution for a short time, at least for one minute, till the indication of the pH meter settles near the pH value of the buffer solution (pH 7) and becomes stable. Thereafter the indicated value has to be adjusted with the zero potentiometer to the exact value of the buffer solution. The zero potentiometer is sometimes marked @pH or pHas. 5) ( zero point calibration is always For analogue pH meters it is important that the performed before the slope calibration. ) pH 4) After the zero point calibration, the electrode assembly has to be rinsed again with deionised water and dried with tissue paper. When drying the electrode, care must be taken not to rub the membrane, i.e. only dab the electrode with tissue paper. Under no circumstances must an electrode be rubbed. This might introduce static electricity into the glass shaft of the electrode and could make an accurate pH measurement impossible for hours. ΔmV/ΔpH No rubbing! 6) 5) The electrode assembly is then immersed into the second buffer solution (acid or base – according to the measuring range). After the indication of the pH meter has stabilised, the indicated value has to be adjusted with the slope potentiometer to the exact pH value of the second buffer solution. The slope potentiometer is sometimes marked mV/ pH. It is advisable to re-check the zero point after the slope calibration has been performed. It is of utmost importance that the above calibration sequence must be adhered to, otherwise no valid calibration will be obtained. 7) 6) If the measuring range covers both the acidity and the alkalinity range the third buffer solution serves as a verifying solution. After the zero point (with the first buffer) and the slope calibration (with the second buffer) has been accomplished, the calibration over the entire measuring range has to be verified with the third buffer solution (be aware of acid and alkaline errors). 8) 7) In order to keep the temperature error as low as possible, especially errors due to the diffusion potential and the isotherm intersection point, it is recommended that the calibration be executed at the temperature at which the actual pH measurement will be performed. 8) After the successful calibration the used buffer solution should be discarded. Never re-use buffer solution and never return it to its original storage container. 6) 7) Technic al Buffer Solution pH 4,0 1 60 by T&C Technical 3.9.2.2 pH pH pH pH7 1 pH 2 pH pH pH 3.9.2.1 61 by T&C Technical 3.9.3 pH a) pH b) c) d) e) Page 41 The Frequency of re-calibration has to be established experimentally Ma y 2002 C alib ra pH ! te Calib ra pH ! te Calib ra pH ! te Calib ra pH ! te Calibr a pH ! te Calib ra pH ! te Calib ra pH ! te 3.9.4 3.9.4 Response Time pH pH If a pH electrode assembly is immersed into a buffer solution, it does not show the pH value of that buffer solution instantaneously. A response time of approximately 30 seconds is normal in pH the buffer value with a tolerance 30 of 0,01 0.01pH(0.6mV) order to reach pH units (0,6 mV). For this reason it is very important to wait long enough during the calibration cycle for the indication of the pH meter to become stable. Only then can the indicated value 62 be adjusted to the buffer value. A premature interruption of the electrode adaptation to the buffer value is often a major source for incorrect pH measurements. The response time of an electrode assembly is especially slow if the electrode temperature and by T&C Technical pH pH pH 10℃ 3mol/L pH 0.01pH 63 KCl by T&C Technical 3.9.5 / pH 3 1 2 3 (3 ) a) b) c) d) ( ) e) f) g) 10℃ h) pH pH pH7 0mV pH7 i) pH 64 by T&C Technical 3.10 pH pH pH 1 pH pH pH pH 100kPa 65 by T&C Technical 1012Ω pH ±0.03pH ±0.05pH ±0.02pH 4. pH pH pH 4.1 pH pH ) 66 ( by T&C Technical 4.1.1 48 67 Page 44 If a measuring electrode has to be stored for short periods between measurements it should be immersed in a container filled with tap water or be fitted with a watertight plastic cap filled with tap water. Therefore one should not throw the supplied plastic caps away. They should kept for by T&C be Technical re-use. 4.1.2Storage of Reference Electrodes 4.1.2 The Reference electrodes should always be stored wet, i.e. the diaphragm has to be covered with the same reference electrolyte with which the reference electrode has been filled. Wet storage must also be applied when the electrode is stored for a long time. It is not advisable to store reference electrodes dry as 3mol/L the reference electrolyte will slowly KCl penetrate through the diaphragm and crystallise and subsequently cover the complete bottom of the electrode. The crystallisation itself does not cause a problem but the reference electrode might dry out completely which will result in a substantial increase in the diaphragm resistance. Even when the reference electrode is refilled with its respective electrolyte, the high diaphragm resistance will not disappear immediately and will result in large measuring errors or even make a measurement totally impossible. Closed refill aperture KCl 3 mol/l or (KClRubber 3mol/L) plastic cap filled with reference electrolyte ( 3 mol/l KCl) Therefore for short or long periods of time it is essential to store reference electrodes in their respective reference electrolyte, with the refill aperture closed by a suitable stopper. Storage in tap water or in distilled water should be avoided. Any penetration of these liquids through the diaphragm will increase the diaphragm potential considerably and will significantly falsify the subsequent pH measurement. 4.1.3 The Storage of Combination Electrodes A combination electrode consists of a measuring electrode and a reference electrode combined into one single-rod electrode. The storage conditions must therefore be suitable for a measuring and for a reference electrode. As every reference electrolyte is an aqueous pH solution it has been found that the optimum storage liquid is the respective reference electrolyte of that combination electrode. The refill aperture has to be closed during storage time. All that has been stated about the storage of a reference electrode applies equally to the storage of a combination electrode. Gel-filled combination electrodes are an exception to the rule. These electrodes have no refill 4.1.3 aperture and the drying-out of their diaphragm has to be avoided at all costs. Therefore gel-filled combination electrodes must be stored wet in a 3 mol/l KCl solution. This statement applies to polymer electrodes as well. 68 by T&C Technical 3mol/L KCl 4.2 pH a) b) c) pH pH pH 1 0.1mol/L HCl 69 by T&C Technical 1% 0.1mol/L HCl ( ) / 0.1mol/L HCl 0.1mol/L NaOH 2 (3mol/L KCl) 1 70 12 by T&C Technical PN 238290 A (<3%) B 4% 500mL 500mL 3mol/L KCl 500mL 4.3 71 by T&C Technical 5. pH pH 5.1 5×109Ω pH pH pH (1012Ω) 1014Ω 1m 1017Ω pH 150pF/m 200pF/m pH 50 64pF/m 50m pH 102pF/m pH pH +80℃ 70℃ pH 130℃ 5.2 72 -30℃ by T&C Technical 5.3 5.4 pH 107Ω pH pH 73 5.3 Combination Electrode Connection Cable voltages which might be created when moving the coaxial cable. It is of importance that this black layer must be carefully removed when preparing the cable ends for connection to the electrode What has been said about the cable requirements for a separate measuring electrode must also plug and/or to the measuring instrument. If the black layer is not removed, it will cause a be observed for the connection of a combination electrode. It is advisable to use a double short circuit between the internal conductor and the copper screen. When stripping the screened coaxial cable (triax cable). To save costs it is normal practice to connect the inner coaxial cable, tools and hands of the technician should always be dry and clean. After stripping, screen of the coaxial cable to the reference electrode part of the combination electrode. the cable ends should be cleaned with alcohol or ether, which can be with a cloth or a by done T&C Technical brush, both of which should be absolutely clean. Touching the stripped cable ends with wet or 5.4 Cable Preparation and Cable fat-stained fingers will reduce the Routing insulation resistance down to, or below, 107 ohms, which will result in a short circuit of the high resistance measuring chain – a pH measurement becomes The insulation of the internal conductor of a pH connection cable is not only screened with impossible. copper wire mesh but also with an additional black semiconductor layer. This layer suppresses voltages which might be created moving the coaxial is taken of importance that this When routing the pHwhen connection cable carecable. must Itbe not to route theblack pH connection layer must be carefully removed when cables. preparingParallel the cable ends connection cable parallel to power situated power cables tointhe theelectrode vicinity of the pH pHfor plug and/orconnection to the measuring theelectro-magnetic black layer is not removed,(induction), it will cause a must be cables instrument. would leadIf to interference which short circuit between internal conductor screen. When stripping the always be avoided at allthe cost. The outer screen ofand the the triaxcopper cable electrode) should 50m pH (combination coaxial cable, tools on andone hands the In technician always be dry and After stripping, earthed sideof only. principle,should every pH connection cableclean. should be as short as possible, the cable ends shouldunder be cleaned with alcohol or ether, which can be done with a cloth or a but should no circumstance be longer than 50 metres. brush, both of which should be absolutely clean. Touching the stripped cable ends with wet or pH connection cables cannot be buried straight into the ground. If this has to be done these fat-stained fingers will reduce the insulation resistance down to, or below, 107 ohms, which will cables must be installed in a metal or plastic conduit. result in a short circuit of the high resistance measuring chain – a pH measurement becomes impossible. Triax cable Coaxial cable PVC PVCexternal PVC PVC external insulation insulation When routing the pH connection cable care must be taken not to route the pH connection insulation screenParallel situated power cables inscreen cable parallel to power cables. the 1vicinity of the pH insulation connection cables PE would lead to electro-magnetic interference (induction), which must be PE insulation avoided at all cost. The outer screen of the triax cable (combination electrode) should always be earthed on one side only. In principle, every pH connection cable should be as short as possible, screen 2 black internal but should under no circumstance be semiconductor longer than 50 metres. layer conductor internal black semiconductor conductor pH connection cables cannot be buried straight into the ground. If this has tolayer be done these cables must be installed in a metal or plastic conduit. Coaxial cable PVC external insulation Triax cable PVCPVC 1 screen PE insulation internal conductor insulation PVC PVCexternal insulation screen 1 PE insulation black semiconductor layer screen 2 2 internal conductor 74 black semiconductor layer by T&C Technical 5.5 pH 2 1 1m Page 48 5.5 Plug or Cable? pH pH in two configurations: either with Various electrode manufacturers supply their pH electrodes an integrally installed cable (normally 1 metre in length), or with an electrode plug connection. Both configurations have their advantages and disadvantages. Page 48 If the cable is installed to the electrode by the manufacturer, the customer can be assured that the cable connection is water tight and measuring faultspH attributed normally to the electrode/cable Various electrode manufacturers supply their pH electrodes two configurations: either hand with however, when connection (short circuit, moisture ingress) can beinruled out. On the other anthe integrally installed cable (normally 1 metre in length), or with an electrode plug pH electrode assembly has to be replaced (remember: a pH electrode is a consumable item connection. Both configurations have their advantages and disadvantages. with a certain life expectancy), the cable also has to be re-purchased. 5.5 Plug or Cable? If the cable is installed to the electrode by the manufacturer, the customer can be assured that An electrode connection eases faults the attributed electrode removal or replacement during the the cable connectionplug is water tight and measuring normally to the electrode/cable maintenance periodmoisture and is ingress) more economical in the runhand (saving of cable connection (short circuit, can be ruled out. On long the other however, when cost). However, the pH electrode assembly has toalways be replaced (remember: aelectrode pH electrodesocket is a consumable item care must be taken to connect the firmly to the cable plug, pH pH with a certain life expectancy), the cable also has to be re-purchased. otherwise moisture, the biggest enemy to the electrode/cable connection, might penetrate Anthe electrode plug connection electrode replacement during pH the measurement is socket/plug coupling.eases Once the moisture hasremoval enteredorthis joint, a reliable maintenance period and is more economical in the long run (saving of cable cost). However, no longer possible. care must be taken to always connect the electrode socket firmly to the cable plug, otherwise moisture, the biggest enemy to the electrode/cable connection, might penetrate the socket/plug coupling. Once moisture has entered this joint, a reliable pH measurement is Electrode socket – cable plug no longer possible. Electrode head with integrated cable pH and Electrode head different instrument plugs connection Electrode socket – cable plug connection with integrated cable and different instrument plugs 75 by T&C Technical pH pH pH pH pH Theory is necessary, Experience is essential. By Hamilton, pH Electrode system R&D team 76 by T&C Technical pH 77 pH 測定と電極の選び方 6.参考資料 pH 付属参考文書 150322_E-J by T&C Technical 水素イオン濃度 pH という用語を定義する前に、その化学的物理的構造について述べておく必要があります。その中心的な 存在は二つの水素原子(H2)と、1 つの酸素原子で構成される水についてです。 6.1 原子の構造 pH pH 古代ギリシャの哲学において、原子という言葉の意味は、物質のこれ以上小さくできない大きさとされて いました。この基本的な微粒子、現代においてこの概念に当たる用語として用いられるもの、はこれ以上 分解しないものとして考えられていました。事実、ギリシャ語の原子は「これ以上分けられないもの」と いう意味を持ちます。 その寸法と原子というものについての知識は、数世紀にわかりゆっくりと思索され続けましたが、その内 容は、まだ単にそれを推測するだけにとどまっていました。 pH それに変化が現れるのは、16 世紀から 17 世紀における実験的科学の発展により、原子理論の進歩が急速 に進むようになりました。その中で化学者は、すべての液体、気体そして固体は、究極の構成元素あるい は元素に分析されると認識しだしました。 pH 例えば塩は二つの異なった違いのある元素で構成されていることを見つけました。ナトリウムそして塩素 Theory is necessary, Experience is essential. は、現代では化学合成として知られるように、相互に深い絆で結ばれています。 By Hamilton, pH Electrode system R&D team 原子は、1 つの元素の建築ブロック あらゆる元素のすべての原子は化学的に同じ振る舞いをします。このように、化学視点から、原子とは考 慮される最も小さな実体ということになります。元素の化学的特徴は大きな違いがあります。それらの原 子は、多数の化学合成物を構成するための多数の異なる方法により結合しています。それは今日において、 112 の元素として知られています。 1911 年、イギリスの物理学者アーネスト・ラザフォード(1837-1937)は原子構造の理論を組織的にまと め上げました。それは原子を、電子の雲によって囲まれる高密度原子核として最初に視覚化したものでし た。 ラザフォードは、 「原子の質量はその原子核に集中している。原子核には電気の正電荷がある。一方 電子に は、おのおの負電荷がある。」としました。電子により生じる変化の量は、原子核にある正電荷と同じ電気 の量になります。このため通常の原子の電気的状態は中性になります。ラザフォードは原子核を構成する その微粒子を陽子と呼びました。このラザフォードの原子の捉え方は、1913 年、ニール・ボーア(1885-1962) 76 1 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical により見直されることになります。(次章 ボーアの原子) pH 1932 年、イギリスのジェームズ・チャドウイック(1891-1974)は、現在中性子として知られている原子 の原子核にある陽子以外の他の微粒子が存在することを発見しました。それは陽子と同じ質量を持ってい るのですが、電荷を持っていません。この発見により、原子核は陽子と中性子で構成されていることを知 ることになりました。 「すべての原子において、陽子の数は電子の数と同じである。ゆえに、それは原子の 原子番号となる。」(元素の周期律表における原子の位置) pH 6.2 pH ボーアの原子 原子の構造を説明する前に、デンマークの物理学者、ニールズ・ヘンドリック・デビッド・ボーアにより 1913 年に「ボーア 原子の理論 The Bohr Theory of the Atom」として知られる仮説が提唱されたことを 知っておく必要があります。彼は、原子核からかなりの距離まで、電子は定められた殻、量子準位に配置 されていると仮定していました。これら電子の位置は「電子配置」と呼ばれます。 pH ボーアの原子 pH 電子の数は原子の原子番号と等しい:水素は 1 つの殻に一つの電子、酸素は 8 つ、そしてウランは 92 を Theory is necessary, Experience is essential. 持ちます。電子の殻は第 1 の殻から第 7 の殻まで規則正しく並んでいます。それぞれの核は収納できる電 子の数に上限があります。殻は内側の殻から外側の殻まで、K 殻、L 殻から Q 殻と名付けられています。 By Hamilton, pH Electrode system R&D team K 殻は二つの電子まで持てます。L 殻は最大 8 つの電子 まで満たすことができます。M 核は 18 の電子、そして それを一般化すれば、n 殻は 2n の電子で満たすことが 2 できることになります。唯一外側のお殻における電子の 数は、原子の化学的な振る舞いにより決定されます。 原初の殻は、殻毎に 間なく電子で埋め尽くす必要はあ りません。元素の周期律表にある最初から 18 番目の元 素までは、電子は規則正しく増加していき、それぞれの 殻は新しい殻が始まる前まで制限いっぱいの電子で満 たされています。 76 2 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical 19 番目の元素からは、最も外殻の電子はそれより内側の殻の電子が完全に満たされている状態で始まりま pH す。規則性はまだ維持されています。しかし電子は満たされている内側の殻との間を行ったり来たりしだ します。この原子量を増加させるためのその化学的性質の定性的な繰り返しは、元素の周期律表の並び方 と一致する結果になります。 pH pH pH pH Theory is necessary, Experience is essential. 6.3 元素の周期律表 By Hamilton, pH Electrode system R&D team 1869 年、ロシアの化学者ディミトリ・イワノビッチ・メンデレーエフ(1834-1907)は当時知り得たすべ ての元素の質量を表にまとめました。そうすることで、彼はある元素の性質はそれ自身周期的に繰り返す ことを発見しました。それによりメンデレーエフは似たような化学的挙動を持つ元素をまとめ、それぞれ 順番に行の下にまとめていきました。この元素の配置表は周期律表と呼ばれています。 年数を重ねることでより多くの元素が発見され、周期律表は数度にわたり配置し直されていきました。そ の表、今日私たちが知っているものは次のページに示されている通りです。元素は昇順の原子番号(原子 核にある陽子の数の順番)で水平に 7 行にわたり配置されています。それぞれの行は原子の 7 つの電子殻 1 つを表します。1 番目の行の一番目にある水素は最も軽い元素になります。目下の一番重い元素、表の最 後の元素はウンウンビウム(放射性超ウラン元素)で、原子質量 277、112 番目になります。なお表には 全部で 118 の元素が載っています。列のグループは、元素を化学的な動きにより 18 に分けています。こ れは最も外側の殻の電子の数に従っています。 76 3 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical pH pH pH pH pH Theory is necessary, Experience is essential. By Hamilton, pH Electrode system R&D team 76 4 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical pH 6.4 分子 分子とは、ある物質が同じ化学的性質を持つ化合物の最も小さい単位です。 水の分子は 1 つの酸素原子と二つの水素原子を電気的な力、いわゆる化学結 合により構成されています。 フランスの化学者アントワーヌ=ローラン・ド・ラヴォアジエ(1743-1794) pH pH により、古代の哲学者が考えていた「水は基本的な元素」、がそうではないこ とを証明しました。それは酸素と水素の化合物であり、現代の表現では H2O という式で表されます。 分子はともに 2 つの電子を分かち合うことで保たれています(共有結合)。この結合を最大化するために、 原子は相互に特定の位置を取ります。例えばそれぞれの分子はそれ自身定められた形を持ち、単純に見る と、水分子内の水素原子は酸素原子に対し 104.5 の角度でくっついています。そのため水は水素の電子が pH 大きな酸素原子の原子核に向かってわずかに囚われている状態、双極子モーメントを持っています。それ に対し、二酸化炭素 CO2 は酸素原子が直線に配置され、それゆえ双極子モーメントを持ちません。 pH Theory is necessary, Experience is essential. またある元素は他の元素と化合することはありません。 これらは希ガス のグループとなります。これらの原子は 2 つの電子(He)あるいは外 By Hamilton, pH Electrode system R&D team 殻に 8 つの電子(Ne、Ar、 Kr、Xe、Rn)を持ちます。8 個以下の電子 を外殻に持つ元素は、他の活性元素と結合することになります。 原子が作れる化合の数は原子価と呼ばれます。酸素は原子価が 2 となりますが、それは 6 つの電子をその 外殻に持ち、そこを 8 つにするために 2 つの電子を必要としていることを意味します。水素は 1 つの電子 をその唯一の外殻に持っています。そのため原子価は 1 となります。電子殻を充足させるため他の電子を 獲得するか、もしくは 1 つ電子が足りない原子に 1 つ電子を与えることができます。このため 2 つの電子 を必要とする酸素原子の場合、2 つの水素原子は酸素原子の足りない電子の需要を満たし、一緒になるこ とで水分子を作ることになります。 76 5 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical pH ナトリウム元素と塩素元素がくっついて塩(NaCl:食卓塩)を作る場 合、この結合はイオン結合となります。中世のナトリウム原子、それ は 1 つの電子を外殻に持ちますが、それを 7 つの電子を外殻に持つ塩 素原子と共有します。再びですが、この結合の場合、外殻は 8 つの電 子で満たされます。 pH pH この 1 つの電子の変化の過程は両方の原子の電気的平衡状態を放り出 すことになります。ナトリウム原子は正の電荷を持つようになり(1 つの電子を失う)Na となり、そして塩素は 1 つの電子を受け取り負の + 電荷を持つようになります(Cl )。新しいナトリウムイオンの殻の構造 - は、ネオン原子のものに似ており、塩素イオンの新しい殻構造はアル ゴン原子のものに似ています。両方のイオンは互いにそれらのクーロ pH ン引力により保たれます。 6.5 イオン pH 1 つのイオンは中性の原子が 1 つあるいは複数の電子を獲得するあるいは失うときに作られる 1 つの粒子 Theory is necessary, Experience is essential. です。1 つの電子を失った 1 つの原子は、カチオンと呼ばれる正の電荷を持つイオンになります。電子を 1 By Hamilton, pH Electrode system R&D team つ得た 1 つの原子は、アニオンと呼ばれる負の電荷を持つイオンになります。イオンという単語はギリシ ャ語から派生し、 「旅人」を意味します。電場(電界)の影響の元、イオンはそれとは逆の極性を持つもの へと移動(旅)します。そしてその結果、気体や液体に導電率を発生させます。 もし NaCl のイオン結合が高温あるいは水に溶けることで壊されると、塩素原子は獲得した電子を保ち、そ して負の電荷を持つイオンとして存在します。ナトリウムイオンは逆に正の電荷(電子を失う)を持つイ オンとして存在します。 6.6 解離 水は 3 つの化合物のグループ、塩、酸、そして塩基にとって優れた 1 つの溶剤です。水にこれらの化合物 を入れると、これらの化合物はイオンへと解離します。NaCl を水に入れると、水分子の極の作用は負と正 の電荷を持つイオンの間で静電引力が減少し、それらのイオンを引き離すようになります。このためイオ ンの再結合を妨げるように働き、それらのイオンは水の分子に囲まれた状態になります(水和される)。 HCl(塩酸)も同じように水の中で解離し、H と Cl イオンになります。NaOH(水酸化ナトリウム)も同 + - 76 6 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical じように解離し Na と、OH イオンになります。 + - pH 水の中の塩の解離、酸と塩基の解離は水に対し優れた導電性を生じます。このように発生した溶液を電解 液と呼びます。 pH pH pH もし二つの電極が電解液に差し込まれ電解液と電位の違いがこれらの電極に発生すると、負の電荷を持つ pH イオンは正の電荷を持つ電極(アノード)により捕らわれ、そしてアノードに達することでそれらの電荷 Theory is necessary, Experience is essential. を失います (電子を失う) 。これによりアニオンと名付けられています。 同じように負の電荷を持つ電極(カ ソード)はその電荷は電子を受け取ることで失います(カチオン)。 By Hamilton, pH Electrode system R&D team 6.7 酸̶塩基̶塩 化学において、私たちは三つの電解液、酸、塩基そして塩に分類してきました。 酸: 水の中に溶かすと水素イオン濃度 H+(プロトン)を発生し、それは純水の持つ量よりも高く なります。従って酸はプロトンドナー(陽子供与体)です(プロトン=正の電荷を持つ水素イ 76 7 pH 測定と電極の選び方 pH 付属参考文書 150322_E-J by T&C Technical オン)。 酸はその味は酸っぱくそしてリトマス試験紙を赤くします。リトマス試験紙は溶液が酸なのか 塩基なのかを判別するために用いられる最も古い、そして最も使われるものです。それは地衣 類、ある真菌の作り出した植物性有機物、そしてそれと共生しているある藻から分離された桃 色の色素です。 最も知られている酸: pH pH 塩酸 HCl 消化液を構成している成分 硝酸 HNO3 色素と爆薬で使用されます。 酢酸 CH3COOH 酢 蟻酸 HCOOH 染色と製革で使用されます。 硫酸 pH H2SO4 電池 リン酸 H3PO4 歯科用セメント、化学肥料 pH 塩基: 水の中に溶かすとヒドロキシルイオン濃度 OH-(プロトンアクセプター)を発生し、プロトン Theory is necessary, Experience is essential. を受け止めます。塩基は苦味を持ちリトマス試験紙を青に変色させます。塩基はぬるぬるした 感じになります。最も一般的な塩基は以下の通りです。 By Hamilton, pH Electrode system R&D team 塩: 水酸化ナトリウム NaOH 排水、オーブンクリーナ 水酸化カルシウム Ca(OH)2 建築用モルタルに使われています。消石灰 水酸化アルミニウム Al(OH)3 アルミニウム化合物を作る原料 水酸化カリウム KOH 軟石鹸 水酸化マグネシウム Mg(OH)2 マグネシア乳(緩下剤) アンモニア NH3 家庭用クリーナー 酸と塩基の水溶液が混ざった時、中性化する反応が発生します。この反応は瞬く間に進行し、 一般的には水と塩を作ります。例えば硫酸と水酸化ナトリウムは水そして硫酸ナトリウムにな ります。 例: 76 8 pH 測定と電極の選び方 pH 6.8 硝酸ナトリウム NaNO3 = Na + NO3 硫化アルミニウム Al2(SO4)3 = 2Al リン酸カルシウム Ca3(PO4)2 = + 3+ 3Ca - + 3SO4 2+ 付属参考文書 150322_E-J by T&C Technical 2- + 2PO4 3- モル(mol) pH モルは、0.012Kg(12g)の炭素pH中に存在する原子の数と等しい要素粒子を含む系の物質量の SI 単位で 12 す。構成する要素粒子は原子、分子、イオン、電子、あるいは他の微粒子と特定されなくてはならない。 本単位は国際的に用いられるために 1971 年に定められた。(Oxford Dictionary) 炭素 12(他の物質を測定する際の対象としての原子)の 12g に含まれる元素の粒子の数は 6.0221367 x 10 23 pH です。この数字はアボガドロ数として知られ、イタリアの物理学者アメデオ・アボガドロ(1776-1856) の功績を称え名付けられています。1811 年、アボガドロは、温度、圧力の等しい体積のガスは、同数の分 子含むと仮定しました。1 モルはある物質の原子量が等しいもので、グラム重量で表されます。 pH 1mol の H2 Theory is necessary, Experience is essential. 1mol の H2O = 2g = 18g By Hamilton, pH Electrode system R&D team 1mol の Cl2 = 71g 1mol の Rn = 222g 1mol の HCl = 36.5g 1mol の NaOH 6.9 = 40g 水の中の水素イオン濃度 水は電解質を溶解、分離するだけでなく、それ自身、わずかな量の水分子は正の電荷を持つ水素イオン H+ と、負の電荷を持つ水酸化イオン OH-に分離しています。 + H2O = H + OH - H+ =酸の反応を持つ正の電荷を持つ水素イオン OH- =アルカリ反応を持つ負の電荷を持つ水酸化イオン 76 9 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical もし水素イオンの量が水酸化イオンの量と同じ場合、水は中性と呼ばれます。清浄な中性の水は 10 の水 7 pH 分子の 1 つだけが解離した状態です。 現実には水素イオンは 1 つのプロトンであり、1 つの電子を持つ水素イオンが水素イオン同士結合してい ます。そのため一般化された酸と塩基の理論においては、最も単純な酸と塩基の中性反応としてみること になります。 + 2H + 2e =H2 pH pH しかし実際は水素イオン、あるいはプロトンは水の中で自由に存在せず、水分子を伴っています。水のイ オン化はそのため正確には以下のように書かれる必要があります。 + 2HOH = H3O + OH - 水の中にある H3O はヒドロニウムイオンと呼ばれ、そのイオンは酸の性質を持ちます。この解離状態を単 + pH 純化にするため、この等式は通常 H を用いて書かれていることになります。 + pH 1 つの酸を中性の水に入れることで、H イオン濃度は酸の解離により作られる H イオンによりその濃度は + + 上昇します。その水はその性質を変化させ、あたかも酢あるいはレモンジュースのように酸っぱくなりま Theory is necessary, Experience is essential. す。そしてそれは腐食性となり金属を分解します。 By Hamilton, pH Electrode system R&D team 1 つの塩基を中性の水に入れると、OH イオン濃度は塩基の解離により作られる OH イオンによりその濃度 - - は上昇します。それに対応する量の H イオンがありますが、それは減ることになります。水はその性質が + 変わり、苦味があり濡れた石鹸のようにぬるぬるした感触を持ちます。 両方の場合において、私たちはそれを水と呼ぶことはできなくなり、それを「水溶液」として扱うことに なります。 76 10 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical pH すべての酸と塩基の水溶液は、加えられた酸、塩基の化学的反応により、それらに対応した水素イオン H + と水酸化イオン OH の濃度に変化を与えます。水溶液内の水素イオン濃度は 1 つの水素イオンに対応する - 解離していない水分子量により表現されます。 例 もし 1 つの水素イオン H が 100 の水分子のなかに存在した場合、それは 1:100 あるいは 1/10 + pH 2 pH あるいは 10 と書かれます。 -2 もし 1 つの水素イオンが 10,000,000 の水分子のなかに存在した場合、それは 1:10,000,000 あ るいは 1/10 あるいは 10 と書かれます。 7 -7 さらにもし 1 つの水素イオンが 1000,000,000 の水分子のなかに存在した場合、それは 1:1000,000,000 あるいは 1/10 あるいは 10 と書かれます。 9 -9 pH 水の中では H イオンと OH イオンに解離したイオン物質が 22℃、10 (mol/L)では一定の状態で存在しま + - -14 す。そのため、純水中の H イオンと OH イオンの濃度が等しい時、H イオン濃度は 10 に必ずなり、OH pH + - + -7 - イオン濃度は同じように 10 になります。この状況は自動的に pH の定義へと導かれ、その結果 pH は以下 -7 Theory is necessary, Experience is essential. のように表現されます。 By Hamilton, pH Electrode system R&D team 水溶液中における活性水素イオン濃度のマイナスの常用対数 あるいは数学的表現では 1 pH = log 水素イオン濃度(mol/L) 76 11 pH 測定と電極の選び方 付属参考文書 150322_E-J by T&C Technical pH 最後に pH 測定において大切な点は以上です。 「理論は必要です ‒ 経験は欠かせません」 Erich L.Springer pH pH pH pH Theory is necessary, Experience is essential. By Hamilton, pH Electrode system R&D team 76 12