3 INSTALLATION CONDUCTIVITY 8225 3.2.5 Electrical wiring, compact version, 12-30 VDC, with cable glands, with relays Before wiring the device, please read § 3.2.1 and 3.2.3 carefully. Lift the transparent lid after having unfastened the screw. Remove the cover of the device by unscrewing the 4 screws, pass the cables through the cable glands and connect according to the pin assignment below. Always seal the unused cable gland using the supplied obstructor to ensure the tightness of the device. Unscrew the cable gland nut, insert the obstructor and screw the nut back onto the cable gland. 4-20 mA output L+ (12-30 VDC) L- (0V) Connect the power supply cable shielding Switch 1: see Fig. 3.9 CURRENT SOURCE SINK Strap to be removed if 4-20 mA output is used Relays PE Without NC L+ With Iout L+ 1 2 3 4 SENSOR 3 4 Relay cables must be fitted here to the board using the sup plied cable clips (see § 3.2.3) CELL PE REL1 2 PE PT1000 REL2 1 PE LL- Supply 12..30Vdc SENSOR 3A/230VAC Connect the conductivity sensor + - + - Connection relay 2 to type 121 Connection relay 1 to type 142 Fig. 3.8 Wiring, compact transmitter, 12-30 VDC, with cable glands, with relays and connection examples for the relays E-22-