Temperature Dry Well Calibrators CTD9300 3. Moduses and Functions GB During the program run, you can toggle between the main menu and the data logging menu to display temperature levels already processed. Once the program run of the logging program has been completed, the ramp program and the data logger are automatically set to OFF. The calibration data recorded can either be displayed in the data logging menu or downloaded via the RS 232 interface. 3.1.2.2 Switch Test This menu is used for configuring the temperature switch test. It is possible to determine switch points and the hysteresis of a temperature switch. The switch test program and the temperature ramp program are automatic program runs that exclude one another. When the program is enabled, all the functions not used for testing temperature switches are disabled. Fig. 16: Switch test setup menu The switch test is started by pressing ENTER and can be aborted at any time with CONTROL ON/OFF. An interrupted switch test is identified with the status bar message SWITCH TEST. HEAT UP or COOL DOWN will be displayed in the status bar during the switch test. 9037926 07/2015 Rem. temp, Tmax or Tmin are displayed to the left-hand side of the set temperature. After the test has been completed and the hysteresis has been calculated, the removal temperature will be controlled such that a new switch test can be started. If the temperature fluctuations are < 0.1 K, TEMP.STABLE will be displayed in the status bar. 32 WIKA Operating Instructions CTD9300 Rev. 1