Download Eignungsprüfbericht für Gaskomponenten

Transcript
TÜV Rheinland Energie und Umwelt GmbH
Air Pollution Control
Report on performance testing of the Fidas® 200 S measuring system manufactured by PALAS GmbH for the components suspended particulate matter
PM10 and PM2.5, Report no.: 936/21218896/A
Page 61 of 476
The following value pairs were discarded:
Table 8:
Discarded reference PM10 value pairs according to Grubbs
Test site
Date
Reference 1 [µg/m³]
Reference 2 [µg/m³]
Cologne, summer
21.05.2012
45.7
41.6
Bornheim, summer
12.07.2013
28.7
33.5
Table 9:
Discarded reference PM2.5 value pairs according to Grubbs
Test site
Date
Reference 1 [µg/m³]
Reference 2 [µg/m³]
Cologne, summer
18.05.2012
7.1
16.0
Cologne, summer
23.05.2012
27.3
35.0
1_3185580_2013_936_21218896A_EN.doc