Download PSD-30 - TechnoMad
Transcript
GB 7. Starting, operation Zero offset compensation Zero offset compensation When putting the device into operation, check the zero point indicated in the display. If, due to the mounting, an offset is displayed, a reset can be made in the programming mode using the parameter 0SET. For relative or vacuum measuring ranges, make sure that the device is depressurized before carrying out a zero offset compensation. For absolute pressure measuring ranges, the zero offset compensation must be carried out in a vacuum, that means at 0 bar absolute. Due to the fact that special references are required for this, we recommend to have this work carried out by the manufacturer. Circular connector M12x1, 4-pin Circular connector M12x1, 5-pin 2 switching outputs or 1 switching output + 1 analogue output 2 switching outputs + 1 analogue output U+ = 1 Ingress Protection per IEC 60 529 U- = 3 IP 65 and IP 67 SP 1 = 4 / SP2 = 2 / C=4 S+ = 2 U+ = 1 U- = 3 SP1 = 4 / SP2 = 2 C=4 S+ = 5 IP 65 and IP 67 The ingress protection classes specified only apply while the pressure transmitter is connected with female connectors that provide the corresponding ingress protection. 8 WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio PSD-30 11430486.04 GB/D/F/E 03/2011 Wiring details