Download USER MANUAL WEIGHT INDICATORS
Transcript
DGT series indicators DGT _07_12.10_EN_U 6.3 LIMITATION OF THE TARE FUNCTIONS With approved instrument, it is possible to limit the tare functions, by setting “yES” in the step SEtuP >> d.SALE (TECH.MAN.REF.). The tare operations will have the following specifications: SCALE CAPACITY < 100kg ≥ 100kg FUNCTIONING SEtuP >> d.SALE >> rEM.dSP >> no SEtuP >> d.SALE >> rEM.dSP >> yES (no remote display for the visualization of the (remote display for the visualization of the tare) tare) All the tare functions are disabled. - The SEMIAUTOMATIC TARE value can not be modified with a manual tare. - The manual tare can be entered or modified only with an UNLOADED scale and tare equal to zero. - It’s possible to cancel the tare value only with an UNLOADED scale, by pressing the ZERO key or by entering a manual tare equal to zero. - The SEMIAUTOMATIC TARE value can not be - The SEMIAUTOMATIC TARE value can not be modified with a manual tare. modified with a manual tare. - The manual tare can be entered or modified only - The manual tare can be entered or modified with an UNLOADED scale and tare equal to only with an UNLOADED scale and tare equal zero. to zero. - It’s possible to cancel the tare value only with an - It’s possible to cancel the tare value only with UNLOADED scale, by pressing the ZERO key an UNLOADED scale, by pressing the ZERO key or by entering a manual tare equal to zero. key or by entering a manual tare equal to zero. With approved instrument, the d.SALE and rEM.dSP steps are read-only. 6.4 AUTO POWER OFF FUNCTION It is possible to automatically turn off the indicator (from 1 to 255 minutes), or disable it; the auto power off takes place when, with unloaded scale, the weight has not been moved or a key has not been pressed for the time set: the display shows the "- oFF – " blinking message, after this the indicator turns off. For the setting, follow the procedures below: - Turn on the scale, press the TARE key while the firmware version is displayed (the display shows the "tYPE" menu). - Select "F.ModE" (by pressing once the ZERO key) and press PRINT to enter in the menu. - Scroll until one finds the "AutoFF" parameter and select it. - With the ZERO or TARE keys select the possible options: "diSAb" (auto switch-off disabled), "EnAb" (auto switch-off enabled). - Confirm with PRINT; if "EnAb" has been selected, one will be asked to enter the number of minutes after which the indicator should turn off: enter a number between 1 and 255 (using the MODE key to select the digit to be modified and ZERO/TARE to decrease/increase it) and confirm with PRINT. - Press many times the C – ON/OFF key until the display shows "SAVE?". - Press PRINT to confirm the changes made or another key for not saving. 6.5 MULTI RANGE FUNCTIONING (for legal for trade approved instruments) The multirange functioning allows to subdivide the scale capacity in two or three ranges, each which is up to 3000 divisions, improving in this way the first range division in the dual range. For example, with a 10 kg cell platform it is possible to approve the weighing system with: - A single range: 6 kg capacity and 2 g division (3000 div.). - Dual range: 6 / 3 kg capacity and 2/1 g division (3000 + 3000 div.). NOTES: - For the approval of the weighing system in dual range the cell must have better technical features in comparison to the 21