Technical specifications
5.1
General






5.2






5.3


5.4

5.5




Ambient temperature range: 10 - 40C
Ambient operating humidity: 5 - 90%
Compatible with FOCUS (CombiPAL) autosampler. The autosampler should
have a slit in the injection unit.
Compatible with any standard OPTIC liner (5 mm OD)
Liner tray capacity: 28- or 98-positions
Conforms to safety standards:
The European Machine Safety Directive 98/37/EEC,
The Electromagnetic Compatibility Directive 89/336/EEC
Head
Pneumatically operated. Requires 5 bars stable, oil-free, compressed air supply
Compatible with any standard OPTIC injector
Maximum operating injector pressure: up to 500 kPa
Integrated septum purge line
Compatible with Merlin MicroSeal septum
Allows liquid injection with the FOCUS autosampler. The 70 mm syringe needle is
required
Gripper
Mounted on FOCUS Injection Unit
Pneumatically operated. Requires 5 bars stable, oil-free, compressed air supply
Software
Based on CTC PAL Cycle Composer
Capping-De-Capping Station (option)
Mounted on FOCUS X-rail
Pneumatically operated. Requires 5 bars compressed air supply
Compatible with OPTIC liners capped with ATAS GL PTFE liner caps
Liner tray capacity: 28- or 98-positions for capped liners
_______________________________________________
Technical Specifications
Page 5.1