Download Operating Instructions Electronic crane scale
Transcript
8.2 Access service menu Parameters which can be changed in Service Mode are protected through verification. On a verified crane scale, the verification seal must first be destroyed in order to activate Service Mode. It is then necessary to re-verify the crane scale. Entering the service menu, small construction 1. Remove the sealing screw (1) *. 2. Switch on the scale. 3. Press the switch in the hole. The display shows F1. The scale is ready for modifying parameters. Entering the service menu, large construction 1. Remove the sealing screw (1) *. 2. Switch on the scale. 3. Press the switch in the hole. The display shows F1. The scale is ready for modifying parameters. Operation in service menu Navigation in the service menu is the same as in the user menu, see chapter 7.3. * To do this on verified scales the lead seal must be destroyed; the verification status is then no longer valid, see chapter 8.4. HTS-BA-e-1024 51