Chapter 4 75 4.7 – ESPRIT menu. menu. The ESPRITmenu contains commands to control the hardware functions of the instrument. Figure Fig ure 4 .11 – ESPRIT menu items. items . 4.7.1 – Manual Control of the Autolab SPR. SPR. Opens the pump control window used for manual control of a measurement. Sampler position controls and shows the position of the needles. Pump 1 and 2 are syringe pumps Aspirate/Dispense section to aspirate or dispense sample in or out of the cuvette, the sample can be positioned in a microtiter plate or a vial which is held under the needles. Valve options control a three-way valve: valve to needle connects channel from pump to cuvette. Valve to buffer connects channel from buffer flask to pump. Flow control enables to select the speed of the pump in µl/s. Corresponds to the aspirating, dispensing and mixing speed. Mix section for mixing the solution in the cuvette. Fill in the mixing volume in µl and check the mix box to start mixing. This option should only be used when the needles are inside the cuvette and the mix volume should not exceed the cuvette volume. Synchronized Pumps (Channels 1 & 2) This section is meant for synchronized mixing of pump1 and pump2. Set volume and Flow, check the mix box and both pumps