MI 2094 CE MultiTester
General presentation
1. General presentation ............................................................................................................. 4
Safety in use ............................................................................................................................. 4
1.2 Warranty.............................................................................................................................. 5
1.3. List of measurements carried out by the instrument................................................ 7
1.4 List of applicable standards ........................................................................................... 7
2. Description of the instrument.............................................................................................. 8
3. Technical specifications.....................................................................................................10
3.1. Withstanding test (PROG.HV and HV position)........................................................10
3.2. Burn out (HV position)...................................................................................................11
3.3. Low resistance (Continuity position)..........................................................................11
3.4. Voltage drop scaled to 10 A~ (option in Continuity position).................................12
3.5. Insulation resistance.....................................................................................................14
3.6. Leakage current .............................................................................................................14
3.7. Substitute leakage current ...........................................................................................15
3.8. Touch leakage current ..................................................................................................15
3.9. Functional test ................................................................................................................16
3.10. Discharge time (DISC.TIME position)......................................................................17
3.11. General..........................................................................................................................18
4. Measurements ......................................................................................................................19
4.1. Withstanding test: ..........................................................................................................19
4.2. Withstanding test with preset voltage/time diagram:..............................................25
4.3. Low resistance test with currents of 0.1A / 0.2A / 10A / 25A~ ...............................27
4.4. Voltage drop scaled to test current of 10 A~ .............................................................30
4.5. Insulation resistance.....................................................................................................31
4.6. Discharge time - external (input MAINS TEST).........................................................36
4.7. Discharge time-internal................................................................................................38
4.8. Leakage current .............................................................................................................40
4.9 Subtitute leakage current...............................................................................................42
4.10 Touch leakage current.................................................................................................43
4.11. Functional test..............................................................................................................45
4.12. Autotest..........................................................................................................................48
5. Operation ................................................................................................................................53
5.1. Warnings .........................................................................................................................53
5.2. Memorizing results ........................................................................................................54
5.3. Recalling of stored results...........................................................................................57
5.4. RS 232 Communication...............................................................................................58
5.5. System configuration.....................................................................................................59
5.6. Contrast of the display ..................................................................................................62
5.7. Use of remote control pedal ........................................................................................63
5.8. Use of WARNING LAMP................................................................................................65
5.9. Use of barcode reader..................................................................................................66
5.10. Use of EXT/DOOR input .............................................................................................67
5.10.1. DOOR input ...............................................................................................................67
5.10.2 EXTERNAL input........................................................................................................67
6. Maintenance...........................................................................................................................70
6.1. Metrological check .........................................................................................................70
6.2. Service..............................................................................................................................70
3
MI 2094 CE MultiTester
General presentation
6.3. Cleaning ..........................................................................................................................70
6.4. Replacing the fuses ( properly trained service personnel only!) ..........................70
7. PC Software - CE Link..........................................................................................................72
7.1. Installing CE Link...........................................................................................................72
7.2. Introduction notes ..........................................................................................................72
7.3. Download data ...............................................................................................................74
7.4. Open data file..................................................................................................................77
7.5. Printing documents .......................................................................................................80
7.5.1 Printing selected rows................................................................................................80
7.5.2 Print separately ............................................................................................................82
7.6. Header programming ...................................................................................................83
7.7. Sequence editor.............................................................................................................84
1. General presentation
Safety in use
• Use the instrument in accordance with the manual, otherwise the instrument
•
•
•
•
•
•
•
may be dangerous for the operator!
Read this instruction manual carefully, otherwise use of the instrument may be
dangerous for the operator, for the instrument or for equipment under test!
Use only grounded mains outlets to supply the instrument!
Do not use any damaged mains outlet or damaged mains connection cable!
A competent authorized person must only carry out Service or calibration
procedure!
Only a skilled person, who is familiar with hazardous voltage operations, can
handle CE MultiTester!
Disconnect all unused test leads before starting measurement; otherwise the
instrument can be damaged!
In case of failure of the instrument test results may be impaired. It is a good
practice to regularly confirm/check the accuracy of the tester (by simple field
calibration, comparison to known reference values etc).
Meaning of
,
signs on front panel:
Insulation section
Dangerous voltage may be present;
measurements shall be carried out only
on de-energized object
Continuity section
Measurements shall be carried out only
on de-energized object. Read
instructions on how to replace blown-up
fuse., measurements shall be carried
4out only on de-energized object.
MI 2094 CE MultiTester
General presentation
instructions on how to replace blown-up
fuse., measurements shall be carried
out only on de-energized object.
Withstanding section
Dangerous voltage may be present.,
Sswitch off the instrument immediately if
the red warning lamp (fig. 1., pos. 12,
fig.1.) does not light after switching on
HV generator, and service the
instrument. Always handle as the test
leads are energized.
Touch
Leakage
Current
TEST
SOCKET
230V / 16A
50Hz
Dangerous voltage is present on the
TEST SOCKET (fig.1., pos.10, fig.1.)
after pressing the START/STOP push
button, .
Main plug
230V~ / 50Hz
16A
MAIN
F1
F2,5A/250V
MAIN
F2
F2,5A/250V
TEST
SOCKET
F3
T16A/250V
TEST
SOCKET
F4
T16A/250V
Dangerous voltage is present on the
FUSES fuses. - sSwitch off the
instrument and disconnect all test
cables and mains cord before replacing
the fuses or opening the instrument.
1.2 Warranty
Unless stated differently, our instruments are guaranteed against any manufacturing
or material defect. They do not bear the specification known as the safety
specification. Our guarantee, which may not under any circumstances exceed the
amount of the invoiced price, is for the repair of our faulty equipment only, carriage
paid to our workshops. It is applicable for normal use of our instruments and does not
5
MI 2094 CE MultiTester
General presentation
apply to any damage or destruction caused, notably by error in mounting, mechanical
accident, faulty maintenance, defective use, overload or excess voltage.
Our responsibility is strictly limited to the pure and simple replacement of the faulty
parts of our equipment; the buyer expressly renounces any attempt to find u s
responsible for damages or losses caused directly or indirectly.Our guarantee is
applicable for twelve (12) months after the date at which the equipment is made
available. The repair, modification or replacement of a part during the guarantee
period will not result in this guarantee being extended.
6
MI 2094 CE MultiTester
General presentation
1.3. List of measurements carried out by the instrument
•
•
•
•
•
•
•
•
•
Withstanding programmed voltage - time sequence test
Withstanding voltage test
High voltage burn-out test
Continuity test
Voltage drop test
Insulation resistance test
Leakage current tests (leakage, substitute, touch)
Functional tests (power, voltage, current, cos _, frequency)
Discharge time measurement
1.4 List of applicable standards
CE MultiTester is designed in accordance with the following standards:
•
•
EN 61010-1 ............ safety
EN 61326................ electromagnetic compatibility
Measurements in compliance with:
•
•
•
•
•
•
•
•
•
•
•
•
•
EN 60204-1 ............ Electrical equipment of machines
EN 60335-1 ............ Household and similar electrical appliances
EN 60439-1 ............ Switch-gear and control-gear assemblies
EN 60598-1 ............ Luminaries
IEC 60745............... Hand-held motor-operated tools
IEC 60755............... Residual current operated protect. devices
EN 60950................ Safety of information technology equipment
EN 61010-1 ............ Safety requirements for electrical equipment
IEC 61029............... Transportable motor-operated tools
EN 61558-1 ............ Transformers and power supply units
EN 60065................ Audio, video, and similar electronic apparatus
VDE 701 T1............. Repair and modification inspections
VDE 702 T1............. Repeat tests of electrical appliances
The CE MultiTester instrument supply upper standards up to 16 A according to its
maximum power capability. Measurements according to standards above are
possible only up to 16 A ( maximum power capability of the CE MultiTester).
7
MI 2094 CE MultiTester
Description of the instrument
2. Description of the instrument
RS232C
CE MultiTester
ISO
+
MI2094
1000V=
WARNING LAMP
12V~ / 0,3A
EXT / DOOR IN
REMOTE
BAR-CODE READER
600 Vmax
CAT III 300V
DISCHARGING
TIME
Touch
Leakage
Current
TEST
SOCKET
230V~ / 50Hz
16A
230V / 16A
50Hz
MAIN
F1
F2,5A/250V
MAIN
F2
F2,5A/250V
TEST
SOCKET
F3
T16A/250V
TEST
SOCKET
F4
T16A/250V
I
F1
F2
F3
F4
MEM
HELP
RS232
SET
CONTINUITY
P1
10V
25A~
C2
O
P2
LEAK.CURRENT
ISO
TEST ON
C1
CONTINUITY
FUNCTIONALITY
HV
DISC.TIME
PROG.HV
WITHSTANDING
AUTOTEST
5KV~
Fig.1. Front panel layout
Legend:
1 ...... REMOTE connector to connect REMOTE CONTROL PEDAL
2 ...... WARNING LAMP connector to connect WARNING LAMP
3 ...... RS 232 connector to connect external printer or PC
4 ...... EXT/DOOR IN connector
5 ...... BAR CODE READER connector
6 ...... LCD dot matrix display with continuous backlight
7 ...... Positive INSULATION test terminal
8 ...... Grounded INSULATION test terminal
9 ...... DISCHARGE TIME-TEST test terminals
10.... TEST SOCKET 230V /16A
11.... TOUCH LEAKAGE test terminal
12 CONTINUITY test terminals (current test terminals C1, C2, and potential test
terminals P1, P2)
13.... WITHSTANDING test terminals
8
MI 2094 CE MultiTester
Description of the instrument
14.... SET key (press when the instrument is switched ON):
• Set date / time
• Set serial port baud rate
• Set barcode reader baud rate
• Clear records memory
• Clear devices / records
• Clear program memory
• Load default setting
• INPUT DOOR IN disabled / enabled
• EARTH CONTROL disabled / enabled
15.... RS232 key:
• Select RS 232 communication mode
• Transmit memorized data to PC
16.... ROTARY SWITCH to select desired function
17.... HELP key:
18.... MEM key:
• Memorize results
• Recall memorized results
19 General keys F1 to F4 (function of each key is shown on display)
20.... START/STOP push button
21.... TEST ON warning lamp
22.... POWER ON/OFF switch with indication lamp
23.... T16A 250V 6.3x 32 fuses protect test socket from overloading
24.... F 2.5A 250V fuses protect instrument’s power supply
25.... Mains connector
9
MI 2094 CE MultiTester
Technical specifications
3. Technical specifications
3.1. Withstanding test (PROG.HV and HV position)
Nominal test voltage:
adjustable (100 V÷ 5000) V / (50 Hz, 60) Hz at
Umains = 230 V, Pload = 500 VA
Open circuit test voltage:............Un (nominal test voltage) (-1 % / +10 %) at Umains = 230 V
Output dDifferential output:........2 HV plugs
Voltage shape: ............................sinusoidal
Test voltage readout
Range (kV)
0.100 ÷ 0.999
1.000 ÷ 5.000
Resolution (kV)
0.001
0.001
Accuracy
± (2 % of reading + 5 dig.)
± (3 % of reading + 5 dig.)
Two different voltage procedure modes:
• Standard voltage mode
• Programmable voltage mode (parameters t1, t2, t3, U1, U2)
Trip-out out test current for nominal test voltages up to 1000 V adjustable to 0.5*, 1.0*,
1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0, 9.5, 10, 20, 30, 40,
50, 60, 70, 80, 90, 100, 150, 200, 250, 300, 350, 400, 450, 500 mA.
* except inat Autotest
For test voltages greater than 1000 V, maximum current limit depends on max high
voltage power characteristic (500 VA = max. power).
Accuracy of trip-out test current:.........±10 % of set value.
Test current readout (sin wave)
*0.0 ÷ 500.0
Resolution
(mA)
0.1
0.0 ÷ 500.0
0.1
Range (mA)
Accuracy
± (5 % of reading + 5 digit) absolute value
± (30 % of reading +10 digit) resistive or capacitive
value**
* (1.5 -– 500) mA inat Autotest
** not displayed at ‘Trip out ’ STOP
IIIARC=+22
Absolute value of test current is always displayed (
selected resistive ( ) or capacitive ( ) component.
) together with
Maximum Ttrip out time: < 30 ms after break through.
Timer: adjustable 1 s - 9 min 59 s with resolution of 1 s. Timer OFF function available.
10
MI 2094 CE MultiTester
Technical specifications
3.2. Burn out (HV position)
Selectable voltage ..................................................(100 V÷ 5000) V
Minimum burn out time before overheating: ......10 s
I max.:.........................................................................(50 mA ÷ 60) mA
3.3. Low resistance (Continuity position)
Resistance readout for currents 10 A and 25 A
Range R (Ω)*
0.000 ÷ 0.999
1.000 ÷ 2.000
2.001 ÷ 9.999
* autoranging
Resolution (Ω)
Accuracy
0.001
0.001
0.001
± (3 % of reading + 3 dig.)
± (3 % of reading + 10 dig.)
indicator only
Resolution (Ω)
Accuracy
0.01
0.1
± (5 % of reading + 12 dig.)
± (5 % of reading + 6 dig.)
Resistance readout for current 0.10 A
Range R (Ω)*
0.00 ÷ 9.99
10.0 ÷ 100.0
* autoranging
Resistance readout for current 0.20 A
Range R (Ω)*
0.00 ÷ 9.99
10.0 ÷ 100.0
* autoranging
Resolution (Ω)
Accuracy
0.01
0.1
± (5 % of reading + 6 dig.)
± (5 % of reading + 6 dig.)
Max. output voltage:........................................<6 V∼
Measuring current (electronically stabilized)
selectable 100 mA, 200 mA, 10 A, 25 A
- 100 mA
at R < 50 Ω (Umains : 230 V, original test leads)
- 200 mA
at R < 8 Ω (Umains : 230 V, original test leads)
- 10 A ............ at R < 0.5 Ω (Umains : 230 V, original test leads)
- 25 A ............ at R < 0.2 Ω (Umains : 230 V, original test leads)
Current shape:................................................sinusoidal
Threshold Limit value adjustable:...............10 mΩ ÷ 1.0 Ω (in steps of 10 mΩ ),
1.0 Ω ÷ 2.0 Ω (in steps of 100 mΩ),
or ignored (*** Ω sign is selected)
Timer:................................................................adjustable 1 s ÷ 59 s, resolution 1 s
Connectionsystem:.......................................4 wire, safety connectors, electrically
separated
11
MI 2094 CE MultiTester
Technical specifications
Test voltage readout with currents 10 A and 25 A
Range (V)
0.000 – 10.000
Resolution (V)
0.001
Accuracy
± (3 % of reading + 0.05 V)
Test voltage readout with currents 0.1 A and 0.2 A
Range (V)
0.000 – 10.000
Resolution (V)
0.001
Accuracy
± (5 % of reading + 0.1 V)
Test current readout 10 A and 25 A
Range (A)
0.0 – 30.0
Resolution (A)
0.1
Accuracy
± (3 % of reading + 5 dig.)
Test current readout 0.1A and 0.2A
Range (A)
0.000 – 1.000
Resolution (A)
0.001
Accuracy
± (5 % of reading + 5 dig.)
3.4. Voltage drop scaled to 10 A~ (option in Continuity
position)
Voltage drop readout (scaled to 10A~)
Range ΔU (V)
0.00 – 10.00
10.00 – 99.99
Resolution (V)
0.01
0.01
Accuracy
± (3 % of reading + 3 dig.)
indicator only
Resolution (A)
0.1
Accuracy
± (3 % of reading + 3 dig.)
Test current readout
Range (A)
0.0 – 30.0
ThresholdLimit value of voltage drop versus wire section:
2
ThresholdLimit voltage drop (V)
5.0
5.0
3.3
2.6
1.9
1.4
1.0
Wire section (mm )
0.5
0.75
1
1.5
2.5
4
≥6
Any of the wire sections in the previous table can be selected in order to evaluate
voltage drop result.
12
MI 2094 CE MultiTester
Technical specifications
13
MI 2094 CE MultiTester
Technical specifications
Max. output voltage:.....................10 V∼
Electronically stabilized current
Current shape:.............................sinusoidal
Measuring current (external resistance. of (0.0 Ω - 0.5) Ω connected to original test
cable):..........................................>10 A∼
Timer:.............................................adjustable (1 s - 59) s, resolution 1 s
Connection system:....................4 wire, safety connectors, electrically separated
3.5. Insulation resistance
• Nominal voltage 250 V, 500 V, 1000 V
Insulation resistance readout:
Range* (MΩ)
0.000 - 1.999
2.000 – 199.9
Resolution ** (MΩ)
0.001
0.001, 0.01, 0.1
200 - 999
1
Accuracy
± (5 % of reading + 10 dig.)
± (3 % of reading + 3 dig.)
± (10 % of reading + 10
dig.)
* autoranging, depends on test voltage
** depends on test voltage
Measuring range (stable and accurate result even with capacitive load): .. (0 MΩ - 1)
MΩ
Nominal voltage:............(250 V, 500 V, 1000) V (+30 % / - 0 %)
Short circuit current: ......3.5 mA max.
Measuring current: ........1 mA min. at (250 kΩ / 250 V, 500 kΩ / 500 V, 1000) kΩ / 1000
V
Settable limits: ...............0.2 MΩ - 200.0 MΩ (resolution 0.1 MΩ), no limit
(*** MΩ sign is selected)
Timer:...............................adjustable 1 s - 9 min 59 s with resolution of 1 s
Timer OFF function available.
Subresult:........................measuring voltage
Output: .............................2 safety plugs, grounded
Auto-discharge after test.
3.6. PE leakage current
Leakage current readout:
Range (mA)
0.00 – 3.99
4.0 – 20.0
Resolution (mA)
0.01
0.1
Accuracy
± (5 % of reading + 3 dig.)
± (5 % of reading + 3 dig.)
Settable limits: .............................(0.1 mA – 20.0) mA (in step of 0.1 mA)
Output: ...........................................16 A power test socket
14
MI 2094 CE MultiTester
Technical specifications
Timer:.............................................adjustable 1 s - 9 min 59 s with resolution 1 s
Timer OFF function available.
Frequency response………….complies to EN61010-Figure A1
3.7. Substitute leakage current
Substitute leakage current readout:
Range (mA)
0.00 – 20.0
Resolution (mA)
0.01
Accuracy
± (5 % of reading + 3 dig.)
Settable limits: .............................no, (0.1 mA – 20.0) mA (in step of 0.1 mA)
Short circuit current: ....................< 30 mA
Open circuit voltage:....................40 V
Output: ...........................................16 A power test socket
Displayed current is calculated to 230 V.
Frequency response………….complies to EN61010-Figure A1
3.8. Touch leakage current
Touch leakage current readout:
Range (mA)
0.00 – 2.00
Resolution (mA)
0.01
Accuracy
± (5 % of reading + 3 dig.)
Settable limits: .............................no, (0.1 mA – 2.0) mA (in step of 0.1 mA)
Output: ...........................................16 A power test socket + TOUCH safety plug
RAmeter: 2 k_
Frequency response………….complies to EN61010-Figure A1
15
MI 2094 CE MultiTester
Technical specifications
3.9. Functional test
Active Power, Apparent Power, Voltage, Current and Frequency monitoring on the test
socket:
Active Power, Apparent Power
Range (W)
0 – 199.9
200 – 3500
Resolution (W)
0.1
1
Accuracy
± (5 % of reading + 10 digit)
± (5 % of reading + 3 digit)
Resolution (V)
1
Accuracy
± (2 % of reading + 2 digit)
Resolution (A)
0.001
0.01
Accuracy
± (3 % of reading + 5 dig.)
± (5 % of reading + 5 dig.)
Resolution
0.01
Accuracy
± (3 % of reading + 3 dig.)
Resolution (Hz)
0.01
Accuracy
± (0.1% of reading +3 dig.)
Test voltage readout
Range (V)
0 – 400 V
Test current readout
Range (A)
0 – 0.999
1.00 – 15.99
Cos _ readout
Range
0 – 1.00
Frequency readout
Range (Hz)
45.00 – 65.00
ThresholdLimit Apparent Power:.......adjustable
(10 VA ÷ 3500) VA
(10 VA ÷ 100) VA (resolution 1 VA)
(100 VA ÷ 3500) VA (resolution 10 VA)
Output: ....................................................16 A power test socket
Timer:
adjustable 1 s - 9 min 59 s with resolution 1 s
Timer OFF function available.
16
MI 2094 CE MultiTester
Technical specifications
3.10. Discharge time (DISC.TIME position)
• Discharge time on power plug (external)
Max. working voltage......................................750V p
Min. working voltage.......................................(70 Vp, 140) V p
Measuring range ............................................(0 s - 10) s
Resolution .......................................................0.1 s
ThresholdLimit of discharging time............1 s
Accuracy ...........................................................±(2 % of reading + 0.2 s)
Safe voltage level............................................60 V, 120 V
Internal resistance of input...........................96 MΩ
• Discharge time on internal electronic components (internal)
Max. working voltage......................................750 V p
Min. working voltage.......................................(70 Vp, 140) V p
Measuring range ............................................(0 s - 10) s
Resolution ......................................................0.1 s
ThresholdLimit of discharging time............5 s
Accuracy ...........................................................±(2% of reading + 0.2 s)
Safe voltage level............................................60 V, 120 V
Internal resistance of input...........................96 MΩ
17
MI 2094 CE MultiTester
Technical specifications
3.11. General
Mains voltage: ..............................230 V (-10 % ÷ +6 %) / (50 Hz, 60) Hz
Max. power consumption:..........660 VA (without load on TEST SOCKET)
Max. power consumption:..........3.85 kVA (with load on TEST SOCKET)
Display:..........................................LCD dot matrix, (160 x 116) dots with cont. backlight
RS232 interface:..........................1 start bit, 8 data bits, 1 stop bit
RS232 baud rate adjustable:....9600, 19200, 38400 Baud
Memories:.....................................1638 memory locations
Remote control signals:.............START / STOP, SAVE
EXT/ Door in signals:..................Next test, Pass / Fail result, External input, Door in
Bar code reader:..........................EAN13
Bar code baud rate adjustable: 2400, 4800 or 9600 Baud
Measurement circuitry protection:
F1.......F 2.5 A / 250 V (5 mm × 20) mm) (general protection of the instrument)
F2.......F 2.5 A / 250 V (5 mm × 20) mm) (general protection of the instrument)
F3.......T 16 A / 250 V (6.3 mm × 32 mm) (test socket protection)
F4.......T 16 A / 250 V (6.3 mm× 32 mm) (test socket protection)
Note !
For correct operation of the instrument fuses F3 and F4 must be in good condition
because the test socket is the point from which the instrument observes the input
voltage (zero crossing for soft start of internal generator).
Case: ................................................................shock proof plastic / portable
Dimensions (w × h × d):...............................(410 m m× 175 mm× 370) mm
Mass (without accessories):........................13.5 kg
Pollution degree: ............................................2
Degree of protection (at closed cover): ......IP 50
Overvoltage category:.....................................Cat III / 300V, Cat II / 600V
Protection classification:...............................I
Working temp. range: ....................................(0 0C ÷ +40) 0C
Ref. temp. range:............................................(+105 0C ÷ +3530) 0C
Ref. humidity range:.......................................(+40 RH ÷ +70) % RH
Storage temp. range:.....................................(-10 0C ÷+ 60) 0C
Max. working humidity:...................................85% RH (0 0C ÷ +40) 0C)
Max. storage humidity:...................................90% RH (-10 0C ÷ +40) 0C)
80% RH (+40 0C ÷ +60) 0C)
Accuracies apply for 1 year in reference conditions. Temperature coefficient outside
this limits is 0.1% of measured value per 0C° and 1 digit.
Withstanding proof test:
• Between mains and withstanding terminals 7500 Veff / 1 min
• Between mains and other terminals or accessible metallic parts 2200 Veff / 1
min
18
MI 2094 CE MultiTester
Measurements
4. Measurements
4.1. Withstanding test:
Warning !
•
•
•
•
•
•
•
Disconnect all unused test leads before starting measurement; otherwise
the instrument can be damaged!
Only a skilled person, who is familiar with hazardous voltage operations, can
perform this measurement!
Check instrument and test leads for any sign of damage or abnormality
before connecting them to the instrument. DO NOT use test probes in case
of any damage or abnormality!
Always handle the instrument and connected accessories as the
Withstanding test sockets and leads are under the hazardous voltage!
Never touch exposed probe tip, connected equipment under test, or any
other energized part during the measurements. Make sure that NOBODY
else can contact them either!
Connect test probes only for withstanding measurement, and disconnect
them immediately after the test!
DO NOT touch any part of test probe in front of the barrier (keep your fingers
behind the finger guards on of the probe) – possible danger of electric
shock!
Always set the lowest possible trip-out current.
HV (high voltage) position
OBJECT UNDER TEST
CE MultiTester MI2094
iC
(0 - 230) V/ (50, 60)Hz
regulated
iR
100V - 5 kV
V
C LOAD
A
Fig. 2. Test circuitry
19
iA
RLOAD
MI 2094 CE MultiTester
Measurements
iR
0
Re
iA
i=
i +C i
A
R
i.....absolute
value of test current
A
iC
i.....capacitive
current
C
i.....resistive
current
R
-jIm
Fig. 3. Test current diagram
How to carry out the measurement
STEP 1.
Set rotary switch to HV (high voltage) position. The following heading
is displayed:
Function name
Tripping out current
Nominal test voltage
Leakage current
(main result)
Character of selected part of leakage current
Selected (resistive or capacitive)
part of leakage current
Set timer value
Test voltage
Timer key to set timer
value or to switch it on/off
Un key to select
nominal test voltage
Burn key to set burn mode
Ilim key to select tripping out
current or to select character
of displayed leakage current
Fig. 4. Main menu in HV function
STEP 2.
•
Select test parameters as follows:
Test voltage
-
Use U N key to select appropriate test voltage that can be set using ↑ ↓ keys
from 100 V to 1000 V in 10 V steps and from 1000 V to 5000 V in 50 V steps.
20
MI 2094 CE MultiTester
Measurements
Selected test voltage
Fig. 5. Test voltage selection menu
- Press Exit key to exit Test voltage selection menu.
•
Trip-out current / character of displayed part of leakage current
-
Press Ilim key in order to reach menu for trip-out current selection and
character of displayed part of leakage current (resistive or capacitive). See the
following figure.
Selected tripping out current
Absolute value of test current
Character of selected part of leakage current
Fig. 6. Trip-out current selection menu
-
Use ↑ and ↓ keys to select appropriate trip-out current.
Press Chr I in order to change the character of displayed part of leakage
current. If the selected part is resistive,
sign is displayed behind mA unit.
For capacitive part
sign is displayed behind mA unit.
Press Exit key to exit Trip-out current selection menu.
NOTE !
The set thresholdLimit value of test current is always compared with the absolute
value of test current.
21
MI 2094 CE MultiTester
•
Measurements
Timer value / timer OFF/ON
-
Press Timer key and the menu for timer value selection is displayed.
Use ↑ and ↓ keys to select appropriate test time value.
To deactivate the timer press Toff key or Ton key to activate it. See the
following figure.
After pressing Toff
key
After presing Toff
key test time is not displayed
To activate timer press Ton
key
Fig. 7. Timer value selection menu
-
•
Press Exit key to exit Timer value selection menu.
BURN mode
-
Press Burn key to select Burn out mode. In this mode current is limited only
by the internal generator characteristics.
Instead of tripping out current value
BURN message is displayed
Press Ilim
key to cancel Burn mode
Fig. 8. Main menu in HV function when the BURN mode is selected
NOTE !
The result of BURN test cannot be saved
22
MI 2094 CE MultiTester
STEP 3.
Measurements
Connect test probes (pistols) to the instrument as shown in figure below.
Fig. 9. Connection of test probes
STEP 4.
Close DOOR IN