Download Operating Instructions MB 11 Classic
Transcript
Operating Instructions MB 11 Classic 5.2 Placement of the Electrodes For electrode placement, the skin should be cleaned with cotton pad dipped in alcohol or acetone to remove any oil and dirt from skin surface to minimize any potential skin impedance. Using the finger, rub the skin lightly until it turns slightly red, this will help to reduce skin impedance and ensure accurate BERA result. If the patient has very rough or dry Figure 15 Placement of the Electrodes skin, rubbing the mastoid area with cotton pad soaked in salted water will help to improve the BERA result. After skin preparation, apply one disposable electrode sticking pad on each of the three electrode placement points, the vertex, the ground, and the mastoid. An optimal BERA measurement can be achieved by having good electrodeskin contact and accurate positioning of the electrodes. 5.3 Placement of the eartips As human grow, so do the ears and the ear canals. Hence it is important to choose the right ear tip according to the patient’s age and the size of the ear canal. 5.3.1 Placement of eartips for infants As the ear canals for newborn infants are very small, it is important to choose suitable ear tips especially designed for babies (Figure 16). Before inserting the ear tips, connect them to the insert-phone, and rub the soft rubber tip between the thumb and the index finger to warm it up slightly before putting them into the ears of the babies. This will help to avoid discomfort and thus facilitates the screening process. MB11_Classic_E_12a_2.1.1.doc 21 Figure 16 Moulding of the earphones 851 882/2 01/12