Download HMT330 User's Guide
Transcript
Chapter 6 ___________________________________________________Calibration and Adjustment NOTE Fixed pressure compensation value of 1013.25 hPa is used when in adjustment mode Relative Humidity Adjustment Using Push-Buttons A simple push-button adjustment is carried out by using two relative humidity references: 11 % RH (LiCl) and 75 % RH (NaCl). 1. Carry out the chemical purge (if available). LiCl reference 2. Press the ADJ button (see Figure 68 on page 140) on the motherboard to open the adjustment mode. The indicator led starts flashing. 3. Remove the filter from the probe and insert the probe into a measurement hole of the 11 % RH (LiCl) in the humidity calibrator HMK15. Use the adapter fitting for the probes of HMT334, HMT335, HMT337 and HMT338. 4. Wait at least 30 minutes for the sensor to stabilize (the indicator led is lit continuously). Adjustment cannot be done if the conditions are not stabilized (indicator led is flashing). 5. When the indicator led is lit continuously, press the button LiCl~11% to adjust the 11 % RH condition. After adjustment transmitter returns to normal operation mode (indicator LED is unlit). NaCl reference 6. When adjusting in the second reference 75 % RH, press the ADJ button to open the adjustment mode. The indicator led starts flashing. 7. Insert the probe into a measurement hole of the 75 % RH (NaCl) reference chamber of the humidity calibrator HMK15. Use the adapter fitting for the probes of HMT334, HMT335, HMT337 and HMT338. VAISALA ______________________________________________________________________ 141