Mounting
1
2
3
100%
0%
Fig.
1
2
3
Materials, wetted parts
3: Measuring range (operating range) and max. measuring distance
full
empty (max. measuring distance)
Measuring range
Make sure that the wetted parts of VEGAPULS 62, especially
the seal and process fitting, are suitable for the existing
process conditions such as pressure, temperature etc. as well
as the chemical properties of the medium.
You will find specification in chapter "Technical data" in the
"Supplement".
4.2 Mounting instructions
Mounting
By using the standpipe version, the influence of turbulence and
vessel installations, such as e.g. heating spirals or agitators, is
excluded. If turbulence or vigorous product movement occurs
in the vessel, long standpipe antennas should be fastened to
the vessel wall.
28740-EN-070130
The standpipe antenna must extend all the way down to the
requested min. level, as measurement is only possible within
the tube. If good mixing of the product is important, you should
use the version with perforated surge pipe.
Warning:
Remember that with the version with ¾" thread, a max. torque
of 80 Nm (59 lbf ft) must not be exceeded.
VEGAPULS 62 standpipe version - Foundation Fieldbus
15