Innovision Innocor™ Software User Manual inspiration. Default / recommended value is 50%. Solubility coef. for tissue should for N2O be set to 0.412. Solubility coef. for blood should for N2O be set to 0.407. Valve dead space Bag dead space 2.5.2.3 Breath-by-Breath setup The Breath-by-Breath data can be averaged during on-line and off-line display by the BBB Filter Time. The filter is a moving average filter. O2 flow to gas delay is the delay time of the O2 gas signal compared to the flow found during the flow gas delay calibration, see section 2.5.9.3. Similar with CO2 flow to gas delay, which is the delay time of the CO2 gas signal compared to the flow. Both delays are typically in the range 1400 to 1700 ms. The values should not be entered manually, but measured by the flow gas delay calibration. Valve to Mouth dead space is the volume of the respiratory valve unit from the inlet to the mouth. The default value is 120 ml for a valve connected with mouthpiece and bacterial filter (PALL). If a flexible tube is used the value 172 ml shall be used instead. The Valve to Mouth dead space is March 2007 COR-SUM-0000-001-IN /UK, A/7 48